Agilent 8757D
Scalar Network Analyzer
10 MHz to 110 GHz
Data Sheet
Accurate measurement of transmission and reflec-
tion characteristics is a key requirement in your
selection of a scalar network analyzer. To help you
achieve these goals, Agilent Technologies offers you
a choice of microwave scalar measurement solutions
which provide an excellent balance of cost, system
versatility, and measurement precision.
Agilent 8757D overview of features
Four display channels
Three detector inputs
An optional fourth detector input
An optional internal power calibrator
Accurate power measurements with
Agilent 85037 series precision detectors
High resolution color display
Internal plotter/printer buffer
+16 to –60 dBm dynamic range
AC/DC detection modes
101 to 1601 measurement points/trace
Noise figure measurement display capability
Compatible with the Agilent 85025 and 85026
series detectors and the Agilent 85027 series
directional bridges
Limit line testing (channels 1 and 2)
Adaptive normalization
Cursor search functions (max, min, n dB, BW)
Specifications describe the instrument’s warranted performance over the
temperature range 0 to 55°C (except where noted). Supplemental characteristics
are intended to provide information useful in applying the instrument, by giving
typical but non-warranted performance parameters. These are denoted as “typical,”
“nominal,” or “approximate.”
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - 10 MHz to 110 GHz

Agilent 8757DScalar Network Analyzer 10 MHz to 110 GHzData SheetAccurate measurement of transmission and reflec-tion characteristics is a key requirem

Page 2

10Directional bridgesAgilent 85027 series directional bridges (AC/DC)The 85027 series directional bridges are designed tooperate with either the 8757

Page 3 - Modulation requirements

11Power splittersThe Agilent 11667 series power splitters are two-resistor splitters recommended for external sourceleveling or for ratio measurement

Page 4 - System interface

12Agilent 11679A/B extension cablesThese cables extend the distance between thescalar network analyzer and the detector or bridgeto a maximum of 200 f

Page 5 - Ordering information

13Literature guide Pub. numberAgilent 8757D/E scalar network 5091-2469Eanalyzers, BrochureAgilent 8757D scalar network 5967-6177Eanalyzer, Configurati

Page 6 - System accuracy

Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, while minimiz

Page 7

2The Agilent 8757D has four independent displaychannels that process the signals from the Agilent85037 series precision detectors, 85025 and 85026dete

Page 8 - System accessories

3Internal save/recall registersUp to 9 complete front panel states may be savedor recalled. If the source (Agilent PSG, 8360 or 8370series) is connect

Page 9

4GPIBInterfaceGPIB operates according to IEEE 488-1978 andIEC-625 interface standards.Interface function codesSH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP

Page 10 - Directional bridges

5General informationTemperature rangeOperating: 0 to 55 °CStorage: –40 to 75 °CPower requirements48 to 66 Hz, 100/120/220/240 V ±10%, typically 155 VA

Page 11 - Power dividers

6Transmission measurement accuracyTransmission loss or gain measurements are maderelative to a 0 dB reference point established atcalibration.Transmis

Page 12 - System accessories, continued

7Precision detector vs. power sensor absolutepower measurement accuracyAccuracy terms differ depending on the test equip-ment used to make absolute po

Page 13 - Literature guide Pub. number

8DetectorsAgilent 85037 series precision detectors (AC/DC)The 85037 series precision detectors are designedspecifically for operation with the Agilent

Page 14

9Coaxial Detector Summary, Agilent 85025 Series For use with the 8757 in either AC or DC detection modesFrequency Connector Dynamic Return Frequency P