U1251B Frequency Sensitivity During Voltage Measurement
Frequency Sensitivity And Trigger Level
Input Range Minimum Sensitivity
(R.M.S. Sine Wave)
Trigger Level For Dc Coupling
(Maximum input for specied
accuracy = 10 x Range or 1000 V)
20 Hz - 100 kHz >100 kHz ~ 200 kHz < 100 kHz >100 kHz ~ 200 kHz
50.000 mV 10 mV 15 mV 10 mV 15 mV
500.00 mV 25 mV 35 mV 60 mV 70 mV
1000.0 mV 40 mV 50 mV 100 mV 150 mV
5.0000 V 0.25 V 0.5 V 0.5 V / 1.25 V (< 100 Hz) 0.6 V
50.000 V 2.5 V 5 V 5 V 6 V
500.00 V 25 V N/A 50 V N/A
1000.0 V 50 V N/A 300 V N/A
U1252B/U1253B Frequency Sensitivity During Voltage Measurement
Frequency Sensitivity And Trigger Level
Input Range Minimum Sensitivity
(R.M.S. Sine Wave)
Trigger Level For Dc Coupling
(Maximum input for specied
accuracy = 10 x Range or 1000 V)
20 Hz ~ 200 kHz >200 kHz ~ 500 kHz < 100 kHz >100 kHz ~ 500 kHz
50.000 mV 10 mV 25 mV 10 mV 25 mV
500.00 mV 70 mV 150 mV 70 mV 150 mV
1000.0 mV 120 mV 300 mV 120 mV 300 mV
5.0000 V 0.3 V 1.2 V 0.6 V 1.5 V
50.000 V 3 V 5 V 6 V 15 V
500.00 V 30 V < 100 kHz N/A 60 V N/A
1000.0 V 50 V < 100 kHz N/A 120 V N/A
[1] Input impedance: Refer to Table A on page 9.
[2] Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000
V ranges where it is 1.5 at full scale, 3.0 at half scale. For non-sinusoidal
waveform, add 0.1% of reading ± 0.3% of range.
[3] The additional error to be added as frequency >20 kHz and signal input
<10% of range: 3 counts of LSD per kHz.
[4] The additional error to be added as frequency >15 kHz and signal input
<10% of range: 3 counts of LSD per kHz.
[6] Input current >35 µArms.
[7] Current can be measured from 2.5 A up to 10 A continuously. An additional
0.5% needs to be added to the specified accuracy if the signal measured is in
the range of 10 A ~ 20 A for 30 seconds maximum. After measuring a current
of >10 A, leave the meter to cool down for twice the measuring time used
before application of low current measurement.
[8] Input current < 3 Arms.
[9] For non-square wave signals, add 5 counts.
[10] AC voltage and AC current specifications are true RMS AC coupled, valid
from 5% to 100% of range.
Frequency Sensitivity During Current Measurement
Input Range Minimum Sensitivity (R.m.s. Sine Wave)
20 Hz ~ 20 Khz
500.00 µA 100 µA
5000.0 µA 250 µA
50.000 mA 10 mA
440.00 mA 25 mA
5.0000 A 1 A
10.000 A 2.5 A
Frequency Specications
[5]
Range Resolution Accuracy Min. Input Frequency
U1251B/2B U1253B
99.999 Hz 0.001 Hz 0.02%+3
[9]
0.02%+3
[9]
1 Hz
999.99 Hz 0.01 Hz 0.02%+3, <600 kHz 0.02%+3, <600 kHz
9.9999 kHz 0.0001 kHz
99.999 kHz 0.001 kHz
999.99 kHz 0.01 kHz