Agilent Technologies U3401A User Manual Page 3

  • Download
  • Add to my manuals
  • Print
  • Page
    / 7
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 2
triggering point is set to 1V and the triggering slope is set to be negative, then the waveform
displayed is as shown in the right figure below.
The simple scheme works for a large number of waveforms. The triggering level can be
changed accordingly. The duration of the display frame can be computed from the time base
setting of the scope. There are 10 divisions in total across the screen, and multiply the time
base setting (in time-unit per division) by 10 gives the duration of the triggering frame.
3) Why do the waveforms sometimes appear to be unstable on the screen?
A major cause is that the input signal is not periodic and the waveforms in the triggering
frames are not identical, resulting in an unstable waveform which is the overlap of many
non-identical waveforms.
A second possible cause is that the signal is periodic but has a complex shape. In this case,
the triggering circuitry may find more than one triggering point that satisfies the triggering
slope and triggering level requirement, within any one cycle of the input signal. The situation
is shown below.
A third possible cause is that the input signal is contaminated by external noise, which again
leads to a complex waveform. The situation is illustrated below. The solution is to remove
the source of noise or select the triggering point with a steepest slope, which can help to
decrease the noise sensitivity.
Page 3 of 9
Page view 2
1 2 3 4 5 6 7

Comments to this Manuals

No comments