User Manual: Agilent Acqiris 10-bit Digitizers Page 21 of 27
Direction
Low Level
High Level
Input
<0.8 V
>2.0 to 3.45 V
Output
0 to 0.8 V
1.7 to 3.3 V
An unconnected input will be pulled high. The high level output will typically give 0.8 V into 50 Ω.
3.8.2. TRIGOUT
The TRIGOUT Signal occurs after an accepted TRIGGER. It can be chosen to be either synchronous to the Trigger
IN or synchronous to the acquisition Clock and, in this case, can be used to trigger events synchronously to the
acquisition clock.
3.9. External Trigger Output
When the digitizer is ready to be triggered and a valid trigger signal occurs, a trigger output is generated for external
use. It is always available on the Front Panel Trigger Out MMCX connector. The pulse ends when the data
acquisition for the trigger in question is complete.
NOTE: The External Trigger Output functionality is implemented in the hardware. No Trigger Out signal occurs for
software-generated triggers such as those of the AUTO mode of AcqirisLive or through the use of the function
AcqrsD1_forceTrigger. However, AcqrsD1_forceTriggerEx does generate the signal.
Trigger Output Block diagram:
The output swing is 1.6 V (± 0.8 V) when
unloaded and 0.8 V when terminated on 50
. The rise and fall times are 2.5 ns typical.
The offset can be adjusted, by software
control in the range [2.3 V, +2.3 V]
unloaded, or [-1.15 V, +1.15 V] into 50 .
The maximum output current capability is ±
15 mA. As the output is retro-terminated, it
is possible to drive a 50 line unterminated
(HiZ) without loss of performance.
For a TTL compatible signal, set the offset
to 1.0 V and the swing at destination will be
+0.2 to +1.8 V.
For an ECL compatible signal, terminated
on 50 to 1.2 V, set the offset to 1.2 V
and the output will be in the range [0.8 V,
1.6 V]).
Alternatively, to reduce the current drawn from the digitizer, the terminations shown here can be used:
A standardized trigger out signal can also be routed to the PXI Bus Star Trigger line.
Signal: 1.6V swing
0v centered
Offset: +/-2.5 V
G=1
50 Ohm
Trigger Out
R
68
R
220
-5V
GND
Trigger
Out
With
Offset
-1.18V
R
68
R
220
GND
+5V
ECL
Signal
TTL
signal
Trigger Out
With
Offset
+1.18V