1100 LC Modules Control Manual - 10 - 03-914947-30:Rev.6 Configuring the Driver The driver software configuration is done in the Galaxie Configurati
1100 LC Modules Control Manual - 100 - 03-914947-30:Rev.6 Overview Window Click on the button: the Status Overview window is displayed and contain
1100 LC Modules Control Manual - 101 - 03-914947-30:Rev.6 General Window This window allows direct control of the autosampler. Information about t
1100 LC Modules Control Manual - 102 - 03-914947-30:Rev.6 Times Window This is a read-only window. The elapsed run time and the elapsed post run ti
1100 LC Modules Control Manual - 103 - 03-914947-30:Rev.6 Thermostat Window This is a read-only window. The actual temperature and the setpoint te
1100 LC Modules Control Manual - 104 - 03-914947-30:Rev.6 Overview Window Click on the button: the Status Overview window is displayed and contain
1100 LC Modules Control Manual - 105 - 03-914947-30:Rev.6 Overview. Click on the required sub-section to display the corresponding screen. General
1100 LC Modules Control Manual - 106 - 03-914947-30:Rev.6 Times Window This is a read-only window. The elapsed run time and the elapsed post run ti
1100 LC Modules Control Manual - 107 - 03-914947-30:Rev.6 Thermostat Window This is a read-only window. The actual temperature and the setpoint te
1100 LC Modules Control Manual - 108 - 03-914947-30:Rev.6 • Not ready: contains the not ready information sent by the instrument. • Events: contai
1100 LC Modules Control Manual - 109 - 03-914947-30:Rev.6 The button turns off all the modules that compose the system. The button turns on the
1100 LC Modules Control Manual - 11 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HP
1100 LC Modules Control Manual - 110 - 03-914947-30:Rev.6 Elution Window This is a read-only window. The actual flow rate and the pressure evolutio
1100 LC Modules Control Manual - 111 - 03-914947-30:Rev.6 Bottles Window Remaining in Bottle Parameter Group The remaining volume of the bottle is
1100 LC Modules Control Manual - 112 - 03-914947-30:Rev.6 The button turns on the pump module only. The button turns off the pump module only. •
1100 LC Modules Control Manual - 113 - 03-914947-30:Rev.6 General Window This window allows direct control of the pump. Information about the pump
1100 LC Modules Control Manual - 114 - 03-914947-30:Rev.6 By clicking on Set Flow, Set Composition, Set Flow Ramp, Turn Seal wash Off or Active Sing
1100 LC Modules Control Manual - 115 - 03-914947-30:Rev.6 The actual flow rate and the percentages of each solvent in the mobile phase are displaye
1100 LC Modules Control Manual - 116 - 03-914947-30:Rev.6 Remaining in Bottle Parameter Group The remaining volume of each bottle is indicated in th
1100 LC Modules Control Manual - 117 - 03-914947-30:Rev.6 • Not ready: contains the not ready information sent by the instrument. • Events: cont
1100 LC Modules Control Manual - 118 - 03-914947-30:Rev.6 General Window This window allows direct control of the pump. Information about the pump
1100 LC Modules Control Manual - 119 - 03-914947-30:Rev.6 The button activates the B1 channel of the Solvent Selection Valve B. The button a
1100 LC Modules Control Manual - 12 - 03-914947-30:Rev.6 NOTE: For more details about the Bus configuration, refer to the Configuring the Communica
1100 LC Modules Control Manual - 120 - 03-914947-30:Rev.6 Elution Window This is a read-only window. The actual flow rate and the percentages of ea
1100 LC Modules Control Manual - 121 - 03-914947-30:Rev.6 Bottles Window Remaining in Bottle Parameter Group The remaining volume of each bottle i
1100 LC Modules Control Manual - 122 - 03-914947-30:Rev.6 • General: displays the synopsis of the instrument, where some current values can be read
1100 LC Modules Control Manual - 123 - 03-914947-30:Rev.6 General Window This window allows direct control of the column compartments. Information
1100 LC Modules Control Manual - 124 - 03-914947-30:Rev.6 Select Set Left (or Right) temperature, then enter the desired temperature in the Left (or
1100 LC Modules Control Manual - 125 - 03-914947-30:Rev.6 The actual and setpoint temperatures for the right and left compartments are displayed V
1100 LC Modules Control Manual - 126 - 03-914947-30:Rev.6 Please note that the Injections field is automatically incremented by one after each inje
1100 LC Modules Control Manual - 127 - 03-914947-30:Rev.6 • Not ready: contains the not ready information sent by the instrument. • Events: cont
1100 LC Modules Control Manual - 128 - 03-914947-30:Rev.6 Information about the detector status is displayed: Faults, not ready parameters, etc. The
1100 LC Modules Control Manual - 129 - 03-914947-30:Rev.6 Times Window This is a read-only window. The elapsed run time and the elapsed post run t
1100 LC Modules Control Manual - 13 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 130 - 03-914947-30:Rev.6 Lamps Window This is a read-only window. The UV Lamp Ignition and On-Time are displayed.
1100 LC Modules Control Manual - 131 - 03-914947-30:Rev.6 • Not ready: contains the not ready information sent by the instrument. • Events: con
1100 LC Modules Control Manual - 132 - 03-914947-30:Rev.6 This window allows direct control of the detector. Information about the detector status i
1100 LC Modules Control Manual - 133 - 03-914947-30:Rev.6 The elapsed run time and the elapsed post run time (if any) are displayed. Signals Window
1100 LC Modules Control Manual - 134 - 03-914947-30:Rev.6 The UV and VIS Lamp Ignition and burn are displayed. When a lamp is changed, reset the cou
1100 LC Modules Control Manual - 135 - 03-914947-30:Rev.6 Status of the Agilent LC 1100 Fluorescence Detector (G1321) Click on the button to disp
1100 LC Modules Control Manual - 136 - 03-914947-30:Rev.6 The button reset the detector module. The button and button enables to take spectra.
1100 LC Modules Control Manual - 137 - 03-914947-30:Rev.6 Signal Window This is a read-only window. The current Excitation and Emission wavelength
1100 LC Modules Control Manual - 138 - 03-914947-30:Rev.6 • Not ready: contains the not ready information sent by the instrument. • Events: conta
1100 LC Modules Control Manual - 139 - 03-914947-30:Rev.6 General Window This is a read-only window. Information about the detector is displayed:
1100 LC Modules Control Manual - 14 - 03-914947-30:Rev.6 In the Plates X scrolling lists, select the type of plates installed in the Well Plate tray
1100 LC Modules Control Manual - 140 - 03-914947-30:Rev.6 The button displays the reference valve control parameters. The available valve position
1100 LC Modules Control Manual - 141 - 03-914947-30:Rev.6 Actuals Window This is a read-only window. The current Signal parameters, the current Va
1100 LC Modules Control Manual - 142 - 03-914947-30:Rev.6 The button allows displaying the recycle valve control parameters. The available valve p
1100 LC Modules Control Manual - 143 - 03-914947-30:Rev.6 Running Instrument Maintenance Software Performing Maintenance and Diagnosis Run the sep
1100 LC Modules Control Manual - 144 - 03-914947-30:Rev.6 Diagnostic Section for the Agilent Technologies LC 1100 Modules Maintenance of the Agile
1100 LC Modules Control Manual - 145 - 03-914947-30:Rev.6 The Limit values correspond to the maximum allowed values for the counters before the
1100 LC Modules Control Manual - 146 - 03-914947-30:Rev.6 Each counter modification is automatically listed in the maintenance audit trail. Press t
1100 LC Modules Control Manual - 147 - 03-914947-30:Rev.6 Info event history Limit event history Error logbook of last analysis State change histor
1100 LC Modules Control Manual - 148 - 03-914947-30:Rev.6 To finish a test, click on the button. The autosampler is then automatically re-initiali
1100 LC Modules Control Manual - 149 - 03-914947-30:Rev.6 Gripper Verification test Gripper Verification Run the Gripper Verification procedure t
1100 LC Modules Control Manual - 15 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 150 - 03-914947-30:Rev.6 Press to close the gripper and move the fingers as close as possible to the top of the v
1100 LC Modules Control Manual - 151 - 03-914947-30:Rev.6 Actions Window This window is used to list the autosampler maintenance actions. Select t
1100 LC Modules Control Manual - 152 - 03-914947-30:Rev.6 Maintenance of the Agilent LC 1100 Quaternary pump (G1311) Click on the button to displa
1100 LC Modules Control Manual - 153 - 03-914947-30:Rev.6 Select the MCGV in the module schema. The following counters limits, actual and accumula
1100 LC Modules Control Manual - 154 - 03-914947-30:Rev.6 The Limit values correspond to the maximum allowed values for the counters before the mai
1100 LC Modules Control Manual - 155 - 03-914947-30:Rev.6 Each counter modification is automatically listed in the maintenance audit trail. Press
1100 LC Modules Control Manual - 156 - 03-914947-30:Rev.6 MCGV switches A: Total number of channel A valve switches. MCGV switches B: Total number o
1100 LC Modules Control Manual - 157 - 03-914947-30:Rev.6 Limit event history of last analysis Parameters changes during the last analysis Run hist
1100 LC Modules Control Manual - 158 - 03-914947-30:Rev.6 Perform the Pressure Test to demonstrate pressure–tightness up to 400 bars. The test cons
1100 LC Modules Control Manual - 159 - 03-914947-30:Rev.6 For analyzing the test results, refer to the pump operating manual. Leak test Run the Le
1100 LC Modules Control Manual - 16 - 03-914947-30:Rev.6 For the G1312 module, the following screen is displayed:
1100 LC Modules Control Manual - 160 - 03-914947-30:Rev.6 Leak test pressure profile05010015020025030035040001234567Time (min)Pressure (bar) Ramp 1:
1100 LC Modules Control Manual - 161 - 03-914947-30:Rev.6 For analyzing the test results, refer to the pump operating manual.
1100 LC Modules Control Manual - 162 - 03-914947-30:Rev.6 Actions Window This window is used to list the pump module maintenance actions. Select th
1100 LC Modules Control Manual - 163 - 03-914947-30:Rev.6 Pump pistons replaced: no outcome on counters. Purge valve frit replaced: no outcome on c
1100 LC Modules Control Manual - 164 - 03-914947-30:Rev.6 The Limit values correspond to the maximum allowed values for the counters before the ma
1100 LC Modules Control Manual - 165 - 03-914947-30:Rev.6 Each counter modification is automatically listed in the maintenance audit trail. Press
1100 LC Modules Control Manual - 166 - 03-914947-30:Rev.6 Logbook Window Select in the scrolling list the logbook to display. The available logbook
1100 LC Modules Control Manual - 167 - 03-914947-30:Rev.6 Tests Window Select a test in the scrolling list and press to start the test. The avail
1100 LC Modules Control Manual - 168 - 03-914947-30:Rev.6 Self-Test Test The Self-Test test consists in running the following individual tests con
1100 LC Modules Control Manual - 169 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Dark Current Tes
1100 LC Modules Control Manual - 17 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 170 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Intensity Test Pe
1100 LC Modules Control Manual - 171 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Holmium Test Per
1100 LC Modules Control Manual - 172 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Cell Test Perfor
1100 LC Modules Control Manual - 173 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Filter Test Perf
1100 LC Modules Control Manual - 174 - 03-914947-30:Rev.6 For analyzing the test results, refer to the detector operating manual. Wavelength Calibra
1100 LC Modules Control Manual - 175 - 03-914947-30:Rev.6 During the test, a theoretical analog signal of approximately 52 mV with a square wave of
1100 LC Modules Control Manual - 176 - 03-914947-30:Rev.6 Actions Window This window is used to list the detector module maintenance actions. Selec
1100 LC Modules Control Manual - 177 - 03-914947-30:Rev.6 UV lamp replaced: UV lamp ignition, UV lamp Burn and UV lamp On Time counters can be rese
1100 LC Modules Control Manual - 178 - 03-914947-30:Rev.6 Appendix GPIB Addresses The 1100 Master module connected to Galaxie with the GPIB cable
1100 LC Modules Control Manual - 179 - 03-914947-30:Rev.6 10 Down Up Down Up Down 11 Down Up Down Up Up 12 Down Up Up Down Down 13 Down Up
1100 LC Modules Control Manual - 18 - 03-914947-30:Rev.6 For the G1311 module, the following screen is displayed:
1100 LC Modules Control Manual - 180 - 03-914947-30:Rev.6 Configuring a Fixed IP Address This appendix describes how to fix the IP address in the A
1100 LC Modules Control Manual - 181 - 03-914947-30:Rev.6 3. Click on the button on the left of the screen. 4. Click on the Configuration tab.
1100 LC Modules Control Manual - 182 - 03-914947-30:Rev.6 5. Select Manual in the TCP Configuration Type dropdown list.
1100 LC Modules Control Manual - 183 - 03-914947-30:Rev.6 6. Type the desired IP address, subnet mask and Gateway in the corresponding fields. 7
1100 LC Modules Control Manual - 184 - 03-914947-30:Rev.6 8. Reboot the HPLC chain.
1100 LC Modules Control Manual - 19 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 2 - 03-914947-30:Rev.6 Table of Contents Introduction...
1100 LC Modules Control Manual - 20 - 03-914947-30:Rev.6 Configuring the Agilent Technologies LC 1100 Column Compartment Click on the button to di
1100 LC Modules Control Manual - 21 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 22 - 03-914947-30:Rev.6 Configuring the Agilent Technologies LC 1100 Detectors The Agilent Technologies 1100 Variab
1100 LC Modules Control Manual - 23 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 24 - 03-914947-30:Rev.6 Miscellaneous Parameter Group Check or uncheck the Galaxie run started by this module (No c
1100 LC Modules Control Manual - 25 - 03-914947-30:Rev.6 Communications Parameter Group Master module HPIB Address: Select this option if the HPLC
1100 LC Modules Control Manual - 26 - 03-914947-30:Rev.6 Configuring the Synopsis In the Overview part of the system configuration, a synopsis can b
1100 LC Modules Control Manual - 27 - 03-914947-30:Rev.6 (temperature, flow, signal, etc.) depending on the installed devices. For example, the fol
1100 LC Modules Control Manual - 28 - 03-914947-30:Rev.6 Agilent 1100 detectors Galaxie Configuration Manager Screen Agilent 1100 detectors Galaxi
1100 LC Modules Control Manual - 29 - 03-914947-30:Rev.6 The method created on this system will have the following structure: However, if some of
1100 LC Modules Control Manual - 3 - 03-914947-30:Rev.6 Status of the Agilent LC 1100 Binary Pump (G1312)...
1100 LC Modules Control Manual - 30 - 03-914947-30:Rev.6 The icon indicates that the corresponding channel is activated. The icon indicates that
1100 LC Modules Control Manual - 31 - 03-914947-30:Rev.6 Building a Method Creating a New Method The method contains parameters for instrument cont
1100 LC Modules Control Manual - 32 - 03-914947-30:Rev.6 2. Choose the system (chromatograph) associated with the method. A method is created for
1100 LC Modules Control Manual - 33 - 03-914947-30:Rev.6 4. Enter the New Method Name in the first field. In the Description field, enter informa
1100 LC Modules Control Manual - 34 - 03-914947-30:Rev.6 Control Method Section for the Agilent Technologies LC 1100 Modules Once the new method is
1100 LC Modules Control Manual - 35 - 03-914947-30:Rev.6 Press the button to display the System Control Method Advanced Tools window: to send t
1100 LC Modules Control Manual - 36 - 03-914947-30:Rev.6 Control Section for the Agilent LC 1100 AutoSampler (G1313) Click on the button to displa
1100 LC Modules Control Manual - 37 - 03-914947-30:Rev.6 Others Parameter Group In the Needle Offset field, enter the vertical needle offset value.
1100 LC Modules Control Manual - 38 - 03-914947-30:Rev.6 • Injection user program: when selecting this injection mode, another tab is displayed all
1100 LC Modules Control Manual - 39 - 03-914947-30:Rev.6 After a new line has been added, double-click on it to edit. A new window is displayed: U
1100 LC Modules Control Manual - 4 - 03-914947-30:Rev.6 Introduction General This document describes the configuration of the Agilent Technologies
1100 LC Modules Control Manual - 40 - 03-914947-30:Rev.6 Move Syringe back and forth Move Syringe home Move Syringe service Move vial to new locatio
1100 LC Modules Control Manual - 41 - 03-914947-30:Rev.6 Speeds Parameter Group In the Sample Drawing Speed field, enter the sample aspiration spe
1100 LC Modules Control Manual - 42 - 03-914947-30:Rev.6 • Injection user program: when selecting this injection mode, another tab is displayed
1100 LC Modules Control Manual - 43 - 03-914947-30:Rev.6 Use the following buttons to edit the table: to add a line to the table. to delete a
1100 LC Modules Control Manual - 44 - 03-914947-30:Rev.6 Use the scrolling list to choose the action to program. The available actions are: Draw fr
1100 LC Modules Control Manual - 45 - 03-914947-30:Rev.6 Depending on the selected action, the parameters to fill in will be different. Thermost
1100 LC Modules Control Manual - 46 - 03-914947-30:Rev.6 Speeds Parameter Group In the Sample Drawing Speed field, enter the sample aspiration spee
1100 LC Modules Control Manual - 47 - 03-914947-30:Rev.6 Others Parameter Group In the Needle Offset field, enter the vertical needle offset value.
1100 LC Modules Control Manual - 48 - 03-914947-30:Rev.6 • Injection user program: when selecting this injection mode, another tab is displayed a
1100 LC Modules Control Manual - 49 - 03-914947-30:Rev.6 Use the following buttons to edit the table: to add a line to the table. to delete a
1100 LC Modules Control Manual - 5 - 03-914947-30:Rev.6 Installation and Configuration Installing the Driver Software The Galaxie software must al
1100 LC Modules Control Manual - 50 - 03-914947-30:Rev.6 Use the scrolling list to choose the action to program. The available actions are: Draw fr
1100 LC Modules Control Manual - 51 - 03-914947-30:Rev.6 Control Section for the Agilent LC 1100 Isocratic Pump (G1310) Click on the button to di
1100 LC Modules Control Manual - 52 - 03-914947-30:Rev.6 Settings Control Prerun Elution Parameter Group In the Flow field, enter the total flow ra
1100 LC Modules Control Manual - 53 - 03-914947-30:Rev.6 Use the following buttons to edit the table: to add a line to the table. to delete a lin
1100 LC Modules Control Manual - 54 - 03-914947-30:Rev.6 Pressure Alarms Parameter Group Enter the Minimum and Maximum pressure limits of the pump.
1100 LC Modules Control Manual - 55 - 03-914947-30:Rev.6 Compressibility Parameter Group To take into account the compressibility of the solvents,
1100 LC Modules Control Manual - 56 - 03-914947-30:Rev.6 General Control Miscellaneous Parameter Group Check Automatic Pump Stop to turn off the pu
1100 LC Modules Control Manual - 57 - 03-914947-30:Rev.6 Settings Control Prerun Elution Parameter Group In the Flow field, enter the total flow r
1100 LC Modules Control Manual - 58 - 03-914947-30:Rev.6 composition. The time is the elapsed time since the beginning of the run. The Pump Gradient
1100 LC Modules Control Manual - 59 - 03-914947-30:Rev.6 Pressure Alarms Parameter Group Enter the Minimum and Maximum pressure limits of the pump.
1100 LC Modules Control Manual - 6 - 03-914947-30:Rev.6 NOTE: For more details about the configuration of a GPIB_Interface or an NI488 bus, refer t
1100 LC Modules Control Manual - 60 - 03-914947-30:Rev.6 Compressibility Parameter Group To take into account the compressibility of the solvents, s
1100 LC Modules Control Manual - 61 - 03-914947-30:Rev.6 General Control Check Automatic Pump Stop to turn off the pump at the end of a sequence
1100 LC Modules Control Manual - 62 - 03-914947-30:Rev.6 Settings Control Prerun Elution Parameter Group In the Flow field, enter the total flow ra
1100 LC Modules Control Manual - 63 - 03-914947-30:Rev.6 Program Parameter Group Program an elution flow and/or a composition gradient. Each step o
1100 LC Modules Control Manual - 64 - 03-914947-30:Rev.6 Limits Control Pressure Alarms Parameter Group Enter the Minimum and Maximum pressure limi
1100 LC Modules Control Manual - 65 - 03-914947-30:Rev.6 Control Control Left and Right Drive Compressibility Parameter Group For each drive, to t
1100 LC Modules Control Manual - 66 - 03-914947-30:Rev.6 Control Section for the Agilent LC 1100 Column Compartment (G1316) Click on the button to
1100 LC Modules Control Manual - 67 - 03-914947-30:Rev.6 Left Compartment sub-section (the Right Compartment sub-section will not be active). Check
1100 LC Modules Control Manual - 68 - 03-914947-30:Rev.6 Left Compartment Prerun Setpoints Parameter Group Check the On box to activate the column c
1100 LC Modules Control Manual - 69 - 03-914947-30:Rev.6 Valves Control Valve Prerun Position Parameter Group Select the position of the valve bef
1100 LC Modules Control Manual - 7 - 03-914947-30:Rev.6 Ethernet Communications The Agilent Technologies 1100 modules can use Ethernet ® as the met
1100 LC Modules Control Manual - 70 - 03-914947-30:Rev.6 Valve Program Parameter Group Program valve position changes during the run. The time is th
1100 LC Modules Control Manual - 71 - 03-914947-30:Rev.6 Settings Control Signal Parameter Group In the Wavelength field, enter the value of the i
1100 LC Modules Control Manual - 72 - 03-914947-30:Rev.6 <0.005 min (RiseTime<0.12 sec) (DataRate=13.74 Hz) >0.005 min (RiseTime>0.12 se
1100 LC Modules Control Manual - 73 - 03-914947-30:Rev.6 In the Range scrolling list, enter the output voltage range value. The available values ar
1100 LC Modules Control Manual - 74 - 03-914947-30:Rev.6 Wavelength Parameter Group In the Wavelength Start field, enter the minimum spectra wavele
1100 LC Modules Control Manual - 75 - 03-914947-30:Rev.6 Time Table Control Program Parameter Group Program detector changes during the run. In t
1100 LC Modules Control Manual - 76 - 03-914947-30:Rev.6 to delete a line in the table. to clear the entire table. Control Section for the Agilent
1100 LC Modules Control Manual - 77 - 03-914947-30:Rev.6 Signals Control Sample Diode Parameter Group Define the center wavelength in the Wl field
1100 LC Modules Control Manual - 78 - 03-914947-30:Rev.6 Define the center wavelength in the Wl field and the bandwidth in the Bw field. The signal
1100 LC Modules Control Manual - 79 - 03-914947-30:Rev.6 >0.85 min (Response Time =16 sec) (DataRate=0.31 Hz) Analogs Control This detector ha
1100 LC Modules Control Manual - 8 - 03-914947-30:Rev.6 GPIB cable81-839700-00GPIB boardGalaxie Acquisition ServerLANGPIB cable Extender81-839704-00
1100 LC Modules Control Manual - 80 - 03-914947-30:Rev.6 Spectra Control Wavelength Parameter Group In the Wavelength Start field, enter the minimu
1100 LC Modules Control Manual - 81 - 03-914947-30:Rev.6 Peak Detector Parameter Group In the Peak Detector Threshold field, enter the peak detecto
1100 LC Modules Control Manual - 82 - 03-914947-30:Rev.6 to add a line to the table. to delete a line in the table. to clear the entire table. C
1100 LC Modules Control Manual - 83 - 03-914947-30:Rev.6 Settings Control Mode Parameter Group Choose the Acquisition Mode between Fluorescence, P
1100 LC Modules Control Manual - 84 - 03-914947-30:Rev.6 Excitation Parameter Group Check Wavelength A to Wavelength D to activate the corresponding
1100 LC Modules Control Manual - 85 - 03-914947-30:Rev.6 Time Table Control Program Parameter Group Program detector changes during the run. Chan
1100 LC Modules Control Manual - 86 - 03-914947-30:Rev.6 to clear the entire table. Baseline Behavior Parameter Group Select in the scrolling list
1100 LC Modules Control Manual - 87 - 03-914947-30:Rev.6 In the Attenuation scrolling list, select the attenuation value. The available values are:
1100 LC Modules Control Manual - 88 - 03-914947-30:Rev.6 Settings Parameter Group Enter the Delay in the corresponding field. Enter the Gate in the
1100 LC Modules Control Manual - 89 - 03-914947-30:Rev.6 Photo Multiplier Tube Parameter Group Select the value of the Gain parameter (amplificatio
1100 LC Modules Control Manual - 9 - 03-914947-30:Rev.6 Galaxie Acquisition ServerGPIB cable81-839700-00National Instrument GPIB boardCAN Cablespro
1100 LC Modules Control Manual - 90 - 03-914947-30:Rev.6 Settings Control General Parameter Group Check Heater On to turn the heater on. In the Tem
1100 LC Modules Control Manual - 91 - 03-914947-30:Rev.6 Automatic Purge Parameter Group In the Purge time field, enter the time during which the i
1100 LC Modules Control Manual - 92 - 03-914947-30:Rev.6 Analogs Control In the Offset field, enter the offset value to adjust the baseline. In the
1100 LC Modules Control Manual - 93 - 03-914947-30:Rev.6 Time Table Control Program Parameter Group Program peakwidth and/or the polarity changes
1100 LC Modules Control Manual - 94 - 03-914947-30:Rev.6 Starting and Viewing an Acquisition Quick Start and Sequence Once the instrument control m
1100 LC Modules Control Manual - 95 - 03-914947-30:Rev.6 In the Acquisition parameters tab, required fields like the Run Name, the Run ID and the
1100 LC Modules Control Manual - 96 - 03-914947-30:Rev.6 Sequence 1. To create a sequence, select the FILE / NEW SEQUENCE menu, and complete the di
1100 LC Modules Control Manual - 97 - 03-914947-30:Rev.6 Running Acquisition Status Viewing a Running Acquisition To view a running acquisition, cl
1100 LC Modules Control Manual - 98 - 03-914947-30:Rev.6 The button turns on all the modules that compose the system. The button turns off all th
1100 LC Modules Control Manual - 99 - 03-914947-30:Rev.6 Injection Window This is a read-only window. The vial and the injection volume are displa