Error MessagesAgilent TechnologiesE8257D/67D PSG Signal GeneratorsThis guide applies to the following signal generator models:E8257D PSG Analog Signal
lpErrorDoc.txt fragmentation). Report this error to Agilent Technologies. The instrument is still functional.-321 Out of memory;
lpErrorDoc.txt Cannot restore specified cal from backup because of a file error.-313 Calibration memory lost; Cannot store cal to backup: {
lpErrorDoc.txt valid state.-310 System error; Critical digital module error. Report this error to Agilent Technologies. This
lpErrorDoc.txt The command issued expected a given personality to be loaded on the given board. Report this error to Agilent Te
lpErrorDoc.txt File system volumes were verified. Informational warning. Powering down system (turning the power off) durin
lpErrorDoc.txt The calibration will take an initial coarse YO pretune calibration that consists of a gain and offset term, and r
lpErrorDoc.txt-271 Macro syntax error A syntactically legal macro program data sequence, written in accordance with IEEE 488.2,
lpErrorDoc.txt-259 LAN setup error; IP address and default gateway must be on the same subnet. Offending entries are changed. The IP ad
lpErrorDoc.txt-256 File name not found; {name} The burst shape type has been defaulted. The specified file does not exist, so the burst sha
lpErrorDoc.txt properly. Report this error to Agilent Technologies.-230 Data corrupt or stale; Either the file is not a bit file, or the
ii NoticeThe material contained in this document is provided “as is”, and is subject to being changed, without notice, in future editions.Further, to
lpErrorDoc.txt-224 Illegal parameter value; The OSR must be an even value. RADio:ARB:GENerate:SINE generates a valid sine waveform and the
lpErrorDoc.txt Due to hardware limitations, only a certain number of unique data points can be created for the IQ map. IQ rotati
lpErrorDoc.txt command.-222 Data out of range; A burst rise shape must have at least two points. Two points are needed to specif
lpErrorDoc.txt The sweep/list manual point has been reassigned to a smaller number value due to the longest list decreasing in
lpErrorDoc.txt source frequency; number of points and start/stop values for list mode; start/stop values for internal I/Q calibr
lpErrorDoc.txt turn the mode on. HSPA has been turned off. To output a valid RF signal, at least one channel must be turned
lpErrorDoc.txt / NumberOfBlks. But block size does not match the BlockSetSize exactly. Try another value. Block size has been adjus
lpErrorDoc.txt-221 Settings conflict; Cannot turn AWGN on, option 403 is required for any AWGN functionality Option 403 (AWGN) is requi
lpErrorDoc.txt The clipping function is not available for the CDMA2000 or 1xEV-DV reverse link radio configurations 1 and 2. The
lpErrorDoc.txt format is off: {name}. This error is reported when a file header command is tried when the format is not on.
Error MessagesIntroductionIntroductionIf an error condition occurs in the signal generator, it is reported to both the front panel display error queu
lpErrorDoc.txt This error is generated if the module is not connected when digital data is turned on.-221 Settings conflict; Th
lpErrorDoc.txt E-TFCI value should be 0 to 125 when E-TFCI table is 1 and TTI is 2msec. E-TFCI value shoudl be 0 to 1
lpErrorDoc.txt match the FM1/PM1 deviation.-221 Settings conflict; FM2/PM2 value set greater than FM1/PM1 value. FM1/PM1 changed to mat
lpErrorDoc.txt-221 Settings conflict; The frequency spacing has been limited by the number of carriers. The frequency spacing and the n
lpErrorDoc.txt-221 Settings conflict; In fixed position the block size must be either 0 or equal to block set size. The block size has been set to
lpErrorDoc.txt-221 Settings conflict; Invalid Channel code. Channel code has been changed to the highest Channel code for this symbol rate
lpErrorDoc.txt-221 Settings conflict; IR buffer size is too small. The 3GPP standard defines the transport block set size as 0 to
lpErrorDoc.txt GAP2 is a single frame method gap that is greater than 7 slots long. DTX slots per frame should be less than 8.-2
lpErrorDoc.txt data source is external. Single is not a valid repeat setting for an external data source.-221 Settings conflict; PHS or
lpErrorDoc.txt-221 Settings conflict; QPCH and FFCH/OCNS cannot be on at the same time. FFCH/OCNS has been turned off. QPCH is generate
Error MessagesError Message FormatError Message FormatWhen accessing error messages through the front panel display error queue, the error numbers, me
lpErrorDoc.txt An attempt was made to setup the R-DCCH while in a radio configuration that does not support this channel. This c
lpErrorDoc.txt channel, you must change to the proper radio configuration before continuing.-221 Settings conflict; RSCH is only
lpErrorDoc.txt firmware release is not available, then an alternate solution is to load a compatible FPGA image by pressing the Update SSIO In
lpErrorDoc.txt 15 >= TGSN + TGL2 + TGD TGD parameter is set to Undefined or 0. TGPL1 * 15 >= TGSN + TGL1 TGPL1 is the duratio
lpErrorDoc.txt increased. To avoid the error, increase the symbol rate of physical channel or decrease the number of bits of Tr
lpErrorDoc.txt is defaulted back to EDGE modulation.-221 Settings conflict; When Skew Mode is enabled, the maximum number of filter sym
lpErrorDoc.txt The data rate is not contained in the defined data rate table.-220 Parameter error; Spread factor exceeds maximum allowable
lpErrorDoc.txt An arming signal was received and recognized by the device but was ignored.-211 Trigger ignored A GET,
lpErrorDoc.txt or type of parameters. -183 Invalid inside macro definition The program message unit sequence, sent with a *DDT
lpErrorDoc.txt The specified filename exceeds the maximum length.-151 Invalid string data; Header Description String is too long and was
Error MessagesError Message TypesError Message TypesEvents do not generate more than one type of error. For example, an event that generates a query e
lpErrorDoc.txt 7.7.2.4.1).-123 Exponent too large The magnitude of an exponent was greater than 32000 (see IEEE 488.
lpErrorDoc.txt-104 Data type error The parser recognized a data element that is not allowed. For example, numeric or string dat
lpErrorDoc.txt In 8340/8757 compatibility mode, the OA query was received but it is not supported for the current active functio
lpErrorDoc.txt211 Not owner; {name} Permission has been denied to perform an operation on the specified file.219 Bad file numbe
lpErrorDoc.txt mode, the user has requested a power setting that is below the ALC range for the attenuator setting. The power ha
lpErrorDoc.txt517 Calibration failure; Cal aborted by user. While executing the internal I/Q calibration, either a DCAS was sent
lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred while attempting to access the IQ CAL CONSTS Arr
lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred. An internal software error has occurred. A meas
lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred. Please check the hardware.520 Sampler Loop unlo
lpErrorDoc.txt One or more OCNS channels must be enabled for channel balancing to occur.600 RPP has tripped. The reve
-440 Query UNTERMINATED after indefinite response A query was received in the same program message in which a query requesting a
lpErrorDoc.txt to respond within the appropriate amount of time. If cycling power does not correct the problem, report this erro
lpErrorDoc.txt During the system configuration scan of parallel bus boards, a board with a bad configuration header was detected
lpErrorDoc.txt file conflict or a hardware configuration error. Report this error to Agilent Technologies.617 Configuration Erro
lpErrorDoc.txt errors below 3.2 GHz. The lowband coupler detector has failed. Report this error to Agilent Technologies.625
lpErrorDoc.txt provide a challenge value; this value cannot be zero.627 Licensing error; The waveform file already has or does not require
lpErrorDoc.txt First {name} is file name, second is file revision and third is supported revision.629 File format invalid; Seque
lpErrorDoc.txt begin on an even sample pair boundary.630 Arb Sequence Error; Unable to reset sequence on baseband board after change in
lpErrorDoc.txt time. This error is reported when the user attempts to activate one type of modulation while the other is activa
lpErrorDoc.txt messages may clarify why this error was reported.700 State Save Recall Error; Failed to save the instrument state. {name}
lpErrorDoc.txt Burst shape files must have two or more points for both the rise and the fall shape specification.702 Load/Store
lpErrorDoc.txt-321 Out of memory; ARB is not functional. There is insufficient memory for the ARB to generate the requested wave
lpErrorDoc.txt Digital module: {name}. Report this error to Agilent Technologies. This error is reported when the FPGAs on the digital
lpErrorDoc.txt valid, but old version of the interface FPGA.808 Digital module unlock error; The output clock is not locked. The clock
lpErrorDoc.txt A file system operation has failed due to a lack of memory in the system. Reduce memory usage and try again.-321
lpErrorDoc.txt There is not enough memory to read the specified data file. Try a smaller file. Reduce the size of any list/sweep