Agilent Technologies 6626A User Manual

Browse online or download User Manual for Power supply units Agilent Technologies 6626A. to view the Agilent HP 6624A Datasheet.

  • Download
  • Add to my manuals
  • Print
  • Page
    / 5
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Two, three, or four isolated outputs
are integrated into one package,
conserving rack space and GPIB
addresses. Most of the outputs
also provide dual ranges, for more
current at lower voltage levels. The
outputs can be connected in parallel
or series to further increase the
flexibility that these products
offer the system designer.
Programming is done using indus-
try standard SCPI commands.
Test system integration can be
further simplified be using the
VXIPlug&Play drivers. These power
supplies help reduce test time with
fast up and down programming,
which is enhanced by an active
downprogrammer which can sink
the full rated current.
Agilent Performance dc Power Supplies speed and accuracy for test optimization
Specifications
(at 0
˚
to 55
˚
C unless
otherwise specified)
40 W 40 W 80 W 80 W 105 W
output output output output output
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
55
Fast up and down programming
Proven reliability keeps test systems running
Easy to integrate into a system
Extensive protection for DUTs
6621A-6624A, 6627A
Multiple-Output:
40 W-105 W GPIB
Output power Low-range volts, amps 0 to 7 V, 0 to 20 V, 0 to 7 V, 0 to 20 V, 0-35 V,
0 to 5 A 0 to 2 A 0 to 10 A 0 to 4 A 0-3 A
High range volts, amps 0 to 20 V, 0 to 50 V, 0 to 20 V, 0 to 50 V,
0 to 2 A 0 to 0.8 A 0 to 4 A 0 to 2 A
Output combinations
for each model
(total number of outputs) 6621A (2) 2 —
6622A (2) — — 2
6623A (3) 1 1 1
6624A (4) 2 2
6627A (4) — 4
6623A(3) 2 — — 1
Special Order Option J03
Programming accuracy Voltage 19 mV 50 mV 19 mV 50 mV 35 mV
+ 0.06% + 0.06% + 0.06% + 0.06% + 0.06%
Current 50 mA 20 mA 100 mA 40 mA 30 mA
+ 0.16% + 0.16% + 0.16% + 0.16% + 0.16%
Readback accuracy Voltage 20 mV 50 mV 20 mV 50 mV 35 mV
(at 25°C ±5°C) + 0.05% + 0.05% + 0.05% + 0.05% + 0.05%
+Current 10 mA 4 mA 20 mA 8 mA 6 mA
+ 0.1% + 0.1% + 0.1% + 0.1% + 0.1%
-Current 25 mA 8 mA 50 mA 20 mA 15 mA
+ 0.2% + 0.2% + 0.2% + 0.2% + 0.2%
Ripple and noise
(peak-to-peak, 20 Hz to 20 MHz;
rms, 20 Hz to 10 MHz)
Constant voltage rms 500 µV 500 µV 500 µV 500 µV 500 µV
peak-to-peak 3 mV 3 mV 3 mV 3 mV 3 mV
Constant current rms 1 mA 1 mA 1 mA 1 mA 1 mA
Load regulation Voltage 2 mV 2 mV 2 mV 2 mV 2 mV
Current 1 mA 0.5 mA 2 mA 1 mA 2 mA
Load cross regulation Voltage 1 mV 2.5 mV 1 mV 2.5 mV N/A
Current 1 mA 0.5 mA 2 mA 1 mA N/A
Line regulation Voltage 0.01% 0.01% 0.01% 0.01% 0.01%
+ 1 mV + 1 mV + 1 mV + 1 mV + 1 mV
Current 0.06% 0.06% 0.06% 0.06% 0.06%
+ 1 mA + 1 mA + 1 mA + 1 mA + 1 mA
Transient response time Less than 75 µs for the output to recover to within 75 mV of nominal value following
a load change within specfications
Page view 0
1 2 3 4 5

Summary of Contents

Page 1 - 40 W-105 W GPIB

Two, three, or four isolated outputsare integrated into one package,conserving rack space and GPIBaddresses. Most of the outputs also provide dual ran

Page 2 - (Continued)

56Supplemental Characteristicsfor all model numbersdc Floating Voltage: All outputs can be floated up to ±240 Vdc from chassisgroundRemote Sensing: Up

Page 3 - 25 W-50 W GPIB

Two or four isolated outputs are integrated into one package,conserving rack space and GPIBaddresses. Dual ranges allow formore current at lower volta

Page 4

58Supplemental Characteristicsfor all model numbersdc Floating Voltage: All outputs can befloated up to ±240 Vdc from chassisgroundRemote Sensing: Up

Page 5 - Agilent Technologies

To see a copy of the user’s guide, please visit our Web site at www.agilent.com/find/manualsBy internet, phone, or fax, get assistance with all your t

Comments to this Manuals