N67xxA: Readback will not
return any information and
errors, such as -410,“Query
INTERRUPTED” and -420,
“Query UNTERMINATED”
will be reported.
Multiple, Semicolon-
Separated Queries
The 662xA overwrites any
previous unread query
responses. The N6700A,
however, can respond to
multiple queries.
Example:
VOUT? 1; VOUT? 2; VOUT? 3
<readback>
662xA: Readback returns only
the information for the last
(VOUT? 3) query.
N6700A: Readback returns
values for all the queries,
separated by semicolons.
Numeric Data
662xA: The power supply will
return numeric data (ASCII
characters) in a format that is
dependent on the type of data
requested. Please refer to the
Numeric Data section in the
Remote Operation chapter of
the 662xA Operating Manuals.
N67xxA Module: Floating-point
numbers returned by the
instrument may not have
exactly the same syntax or
number of digits.
Initial Conditions
If the power supply is not
programmed to recall the
values in the 0 (zero) register
(RCL0), immediately after
power on, the power supply
sets all parameters to their
initial, or “CLR”, values. These
are the values that result from
the “CLR” command. Table 4
below lists these parameters
and their initial values.
10
Parameter 40 W/80 W High/Low 25 W/50 W N67xxA
Voltage Outputs Precision Outputs Modules
Voltage 0 V 0 V High Range Minimum Voltage
Current Minimum current limit 10 mA High Range Minimum Current
Reprogramming Delay 20 ms 20 ms 20 ms
Store/Recall 0 volts and min. 0 V High Range and Minimum voltage and
Registers current limit 10 mA High Range minimum current
Overvoltage (OV) 23 V on low voltage 55 V (High Range) 55 V
outputs and 55 V on
high voltage outputs
Output Channels On On Off
OCP Enabled Off Off 0 (off)
UNMASK Register 0 (cleared) 0 (cleared) 0 (cleared)
SRQ 0 (off) 0 (off) 0 (n/a)
Front Panel Metering Output #1 Output #1 Output #1
Power Supply Last stored value Last stored value Last stored value
Address (Factory set to 5) (Factory set to 5) (Factory set to 5)
Local Control On (enabled) On (enabled) On (enabled)
PON Bit On On 0 (n/a)
PON SRQ Last stored value 0 (off) 0 (n/a)
(Factory set to 0)
Cal Mode Off Off 0 (off)
Table 4. Initial Conditions