Agilent TechnologiesAgilent 1260 Infinity High Performance AutosamplerUser Manual
10 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionFeaturesFeaturesThe 1260 Infinity High Performance Autosampler features
100 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesMetering Home FailedError ID: 4054, 4704The
Agilent 1260 Infinity High Performance Autosampler User Manual 101Error Information7Module Error MessagesMotor TemperatureError ID: 4027, 4040, 4261,
102 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesInvalid Vial PositionError ID: 4042The vial
Agilent 1260 Infinity High Performance Autosampler User Manual 103Error Information7Module Error MessagesPeristaltic Pump ErrorError ID: 4514The peris
104 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesVessel Stuck to NeedleError ID: 4453The vess
105Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies8Test FunctionsIntroduction 106System Pressure Test 107System Pre
106 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsIntroductionIntroductionAll tests are described based on the Agilent
Agilent 1260 Infinity High Performance Autosampler User Manual 107Test Functions8System Pressure TestSystem Pressure TestThe test determines the leak
108 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsSystem Pressure Test1 Run the System pressure test with the Agilent
Agilent 1260 Infinity High Performance Autosampler User Manual 109Test Functions8System Pressure TestSystem Pressure Test EvaluationSystem Pressure Te
Agilent 1260 Infinity High Performance Autosampler User Manual 11Introduction1Overview of the ModuleOverview of the ModuleThe Autosampler transport me
110 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsSample Transport Self AlignmentSample Transport Self AlignmentThe sa
Agilent 1260 Infinity High Performance Autosampler User Manual 111Test Functions8Sample Transport Self Alignment1 Run the Transport Alignment with the
112 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsMaintenance PositionsMaintenance PositionsSome maintenance procedure
Agilent 1260 Infinity High Performance Autosampler User Manual 113Test Functions8Maintenance PositionsChange NeedleThe position is positioning the nee
114 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsMaintenance PositionsArm PositionThe home position of the autosample
Agilent 1260 Infinity High Performance Autosampler User Manual 115Test Functions8Maintenance PositionsChange Metering DeviceWhen removing the metering
116 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsInjector StepsInjector StepsEach movement of the sampling sequence c
Agilent 1260 Infinity High Performance Autosampler User Manual 117Test Functions8Injector StepsStep CommandsTabl e 5 Step CommandsStep Action Comments
118 Agilent 1260 Infinity High Performance Autosampler User Manual8 Test FunctionsInjector Steps
119Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies9MaintenanceIntroduction to Maintenance 120Warnings and Cautions
12 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionOverview of the ModuleControl of the vial/plate temperature in the ther
120 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceIntroduction to MaintenanceIntroduction to MaintenanceFigure 29 on page
Agilent 1260 Infinity High Performance Autosampler User Manual 121Maintenance9Warnings and CautionsWarnings and CautionsWARNINGToxic, flammable and ha
122 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceOverview of MaintenanceOverview of MaintenanceThe following pages descr
Agilent 1260 Infinity High Performance Autosampler User Manual 123Maintenance9Cleaning the moduleCleaning the moduleTo keep the module case clean, use
124 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceRemoving the needle assemblyRemoving the needle assembly When When the
Agilent 1260 Infinity High Performance Autosampler User Manual 125Maintenance9Removing the needle assembly1In the user interface start the maintenance
126 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceRemoving the needle assembly5Attach a 5/16 inch wrench to hold the posi
Agilent 1260 Infinity High Performance Autosampler User Manual 127Maintenance9Installing the needle assemblyInstalling the needle assembly When When t
128 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceInstalling the needle assembly1Push the silicon safety tube delivered w
Agilent 1260 Infinity High Performance Autosampler User Manual 129Maintenance9Installing the needle assembly5Close the leak guideNext Steps:6 Check th
Agilent 1260 Infinity High Performance Autosampler User Manual 13Introduction1Autosampler PrincipleAutosampler PrincipleThe movements of the autosampl
130 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceExchanging the Needle SeatExchanging the Needle SeatWhen When seat is v
Agilent 1260 Infinity High Performance Autosampler User Manual 131Maintenance9Exchanging the Needle Seat3Disconnect the seat capillary from the Inject
132 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceReplacing the Rotor sealReplacing the Rotor seal When When poor injecti
Agilent 1260 Infinity High Performance Autosampler User Manual 133Maintenance9Replacing the Rotor seal3Unscrew and remove the three stator screws from
134 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceReplacing the Rotor seal7Install new rotor seal and isolation seal. 8 R
Agilent 1260 Infinity High Performance Autosampler User Manual 135Maintenance9Removing the Metering SealRemoving the Metering Seal When When poor inje
136 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceRemoving the Metering Seal3Remove the two attached capillaries with a ¼
Agilent 1260 Infinity High Performance Autosampler User Manual 137Maintenance9Removing the Metering Seal7Remove the head body. 8 Carefully remove the
138 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceInstalling the Metering SealInstalling the Metering Seal When When poor
Agilent 1260 Infinity High Performance Autosampler User Manual 139Maintenance9Installing the Metering Seal3Press the piston into the seal. 4 Reinstall
14 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionAutosampler PrincipleInjection SequenceBefore the start of the injectio
140 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceReplacing Peristaltic Pump CartridgeReplacing Peristaltic Pump Cartridg
Agilent 1260 Infinity High Performance Autosampler User Manual 141Maintenance9Replacing Peristaltic Pump Cartridge3Pull the cartridge forward off the
142 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceReplacing Peristaltic Pump Cartridge7Push the cartridge onto the motor
Agilent 1260 Infinity High Performance Autosampler User Manual 143Maintenance9Installing the Interface BoardInstalling the Interface Board1 Switch OFF
144 Agilent 1260 Infinity High Performance Autosampler User Manual9 MaintenanceReplacing the Module FirmwareReplacing the Module FirmwareTo upgrade/do
145Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies10Parts for MaintenanceOverview of Maintenance Parts 146Vial Tray
146 Agilent 1260 Infinity High Performance Autosampler User Manual10 Parts for MaintenanceOverview of Maintenance PartsOverview of Maintenance Parts I
Agilent 1260 Infinity High Performance Autosampler User Manual 147Parts for Maintenance10Vial TraysVial Trays &'()*+,Item p/n Description1 G2
148 Agilent 1260 Infinity High Performance Autosampler User Manual10 Parts for MaintenanceRecommended Plates and Closing MatsRecommended Plates and Cl
Agilent 1260 Infinity High Performance Autosampler User Manual 149Parts for Maintenance10Recommended Vial PlatesRecommended Vial Platesp/n Description
Agilent 1260 Infinity High Performance Autosampler User Manual 15Introduction1Autosampler PrincipleWhen the sample sequence begins, the valve unit swi
150 Agilent 1260 Infinity High Performance Autosampler User Manual10 Parts for MaintenanceKitsKitsAccessory Kit Injection Ugrade KitUpgrade Kit for hi
Agilent 1260 Infinity High Performance Autosampler User Manual 151Parts for Maintenance10Analytical Head AssemblyAnalytical Head Assembly Figure 30 An
152 Agilent 1260 Infinity High Performance Autosampler User Manual10 Parts for MaintenanceInjection Valve AssemblyInjection Valve Assembly &
Agilent 1260 Infinity High Performance Autosampler User Manual 153Parts for Maintenance10Cover PartsCover Parts&Item p/n Description1 5067-4662 Ca
154 Agilent 1260 Infinity High Performance Autosampler User Manual10 Parts for MaintenanceLeak System PartsLeak System Parts Figure 31 Leak system par
155Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies11Identifying CablesCable Overview 156Analog Cables 158Remote Cab
156 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesCable OverviewCable OverviewAnalog cablesRemote cablesBCD cable
Agilent 1260 Infinity High Performance Autosampler User Manual 157Identifying Cables11Cable OverviewCAN cablesLAN cablesRS-232 cablesp/n Description51
158 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesAnalog CablesAnalog CablesOne end of these cables provides a BN
Agilent 1260 Infinity High Performance Autosampler User Manual 159Identifying Cables11Analog CablesAgilent Module to General Purposep/n 01046-60105 Pi
16 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionAutosampler PrincipleThe standard injection starts with draw sample fro
160 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesRemote CablesRemote CablesOne end of these cables provides a Ag
Agilent 1260 Infinity High Performance Autosampler User Manual 161Identifying Cables11Remote CablesAgilent Module to 3396 Series III / 3395B Integrato
162 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesRemote CablesAgilent Module to General Purposep/n 01046-60201 W
Agilent 1260 Infinity High Performance Autosampler User Manual 163Identifying Cables11BCD CablesBCD CablesOne end of these cables provides a 15- pin B
164 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesBCD CablesAgilent Module to 3396 Integratorsp/n 03396-60560 Pin
Agilent 1260 Infinity High Performance Autosampler User Manual 165Identifying Cables11CAN/LAN CablesCAN/LAN CablesBoth ends of this cable provide a mo
166 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesExternal Contact CableExternal Contact CableOne end of this cab
Agilent 1260 Infinity High Performance Autosampler User Manual 167Identifying Cables11Agilent Module to PCAgilent Module to PCp/n DescriptionG1530-606
168 Agilent 1260 Infinity High Performance Autosampler User Manual11 Identifying CablesAgilent 1200 Module to PrinterAgilent 1200 Module to Printerp/n
169Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies12Hardware InformationFirmware Description 170Boot-up and Initial
Agilent 1260 Infinity High Performance Autosampler User Manual 17Introduction1Autosampler PrincipleFlush the Needle Before injection and to reduce the
170 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationFirmware DescriptionFirmware DescriptionThe firmware of the i
Agilent 1260 Infinity High Performance Autosampler User Manual 171Hardware Information12Firmware DescriptionFirmware UpdatesFirmware updates can be do
172 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationFirmware DescriptionThe firmware update tools, firmware and d
Agilent 1260 Infinity High Performance Autosampler User Manual 173Hardware Information12Boot-up and Initialization ProcessBoot-up and Initialization P
174 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationElectrical ConnectionsElectrical Connections• The CAN bus is
Agilent 1260 Infinity High Performance Autosampler User Manual 175Hardware Information12Electrical ConnectionsRear view of the moduleFigure 33 Rear vi
176 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationInterfacesInterfacesThe Agilent 1200 Infinity Series modules
Agilent 1260 Infinity High Performance Autosampler User Manual 177Hardware Information12InterfacesDetectorsG1314B VWD VLG1314C VWD VL+2 Yes N o Ye s 1
178 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationInterfaces• CAN connectors as interface to other modules• LAN
Agilent 1260 Infinity High Performance Autosampler User Manual 179Hardware Information12Interfaces Overview Interfaces CANThe CAN is inter- module com
18 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionAutosampler PrincipleInject-and-Run The final step is the inject- and-
180 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationInterfacesThe RS- 232C is designed as DCE (data communication
Agilent 1260 Infinity High Performance Autosampler User Manual 181Hardware Information12InterfacesAPG RemoteThe APG Remote connector may be used in co
182 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationInterfacesSpecial InterfacesThe CAN DC- out provides 24 VDC p
Agilent 1260 Infinity High Performance Autosampler User Manual 183Hardware Information12Setting the 8-bit Configuration SwitchSetting the 8-bit Config
184 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationSetting the 8-bit Configuration SwitchCommunication Settings
Agilent 1260 Infinity High Performance Autosampler User Manual 185Hardware Information12Setting the 8-bit Configuration SwitchUse the following tables
186 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationSetting the 8-bit Configuration SwitchSpecial SettingsThe spe
Agilent 1260 Infinity High Performance Autosampler User Manual 187Hardware Information12Instrument LayoutInstrument LayoutThe industrial design of the
188 Agilent 1260 Infinity High Performance Autosampler User Manual12 Hardware InformationEarly Maintenance FeedbackEarly Maintenance FeedbackMaintenan
189Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies13LAN ConfigurationSetting up the module in a LAN environment 190
Agilent 1260 Infinity High Performance Autosampler User Manual 19Introduction1System OverviewSystem OverviewLeak and Waste HandlingThe 1200 Infinity S
190 Agilent 1260 Infinity High Performance Autosampler User Manual13 LAN ConfigurationSetting up the module in a LAN environmentSetting up the module
Agilent 1260 Infinity High Performance Autosampler User Manual 191LAN Configuration13Connecting the module via LANConnecting the module via LANIf the
192 Agilent 1260 Infinity High Performance Autosampler User Manual13 LAN ConfigurationConnecting the module via LAN
193Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies14AppendixGeneral Safety Information 194Lithium Batteries Informa
194 Agilent 1260 Infinity High Performance Autosampler User Manual14 AppendixGeneral Safety InformationGeneral Safety InformationGeneral Safety Inform
Agilent 1260 Infinity High Performance Autosampler User Manual 195Appendix14General Safety InformationOperationBefore applying power, comply with the
196 Agilent 1260 Infinity High Performance Autosampler User Manual14 AppendixGeneral Safety InformationSafety SymbolsTabl e 18 Safety SymbolsSymbol De
Agilent 1260 Infinity High Performance Autosampler User Manual 197Appendix14Lithium Batteries InformationLithium Batteries InformationWARNINGLithium b
198 Agilent 1260 Infinity High Performance Autosampler User Manual14 AppendixThe Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/E
Agilent 1260 Infinity High Performance Autosampler User Manual 199Appendix14Radio InterferenceRadio InterferenceNever use cables other than the ones s
Agilent 1260 Infinity High Performance Autosampler User ManualNotices© Agilent Technologies, Inc. 2010-2013, 2014No part of this manual may be reprodu
20 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionSystem OverviewFigure 6 Leak and waste handling concept (overview - typ
200 Agilent 1260 Infinity High Performance Autosampler User Manual14 AppendixSound EmissionSound EmissionManufacturer’s DeclarationThis statement is p
Agilent 1260 Infinity High Performance Autosampler User Manual 201Appendix14Use of SolventsUse of Solvents Observe the following recommendations on th
202 Agilent 1260 Infinity High Performance Autosampler User Manual14 AppendixAgilent Technologies on InternetAgilent Technologies on InternetFor the l
Agilent 1260 Infinity High Performance Autosampler User Manual 203IndexIndex88-bit configuration switchwithout On-Board LAN 183Aaccessory kit 34Agilen
204 Agilent 1260 Infinity High Performance Autosampler User ManualIndexexternal contactcable 166extra-column volume 66Ffan failed 92firmwaredescriptio
Agilent 1260 Infinity High Performance Autosampler User Manual 205IndexSsafety class I 194safety informationlithium batteries 197safetygeneral informa
© Agilent Technologies 2010-2013, 2014Printed in Germany 02/2014*G1367-90014**G1367-90014*G1367-90014Rev. Bwww.agilent.comAgilent TechnologiesIn This
Agilent 1260 Infinity High Performance Autosampler User Manual 21Introduction1System OverviewThe solvent cabinet (1) is designed to store a maximum vo
22 Agilent 1260 Infinity High Performance Autosampler User Manual1 IntroductionSystem Overview
23Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies2Site Requirements and SpecificationsSite Requirements 24Physical
24 Agilent 1260 Infinity High Performance Autosampler User Manual2 Site Requirements and SpecificationsSite RequirementsSite RequirementsA suitable en
Agilent 1260 Infinity High Performance Autosampler User Manual 25Site Requirements and Specifications2Site RequirementsPower CordsDifferent power cord
26 Agilent 1260 Infinity High Performance Autosampler User Manual2 Site Requirements and SpecificationsSite RequirementsBench SpaceThe module dimensio
Agilent 1260 Infinity High Performance Autosampler User Manual 27Site Requirements and Specifications2Physical SpecificationsPhysical SpecificationsTa
28 Agilent 1260 Infinity High Performance Autosampler User Manual2 Site Requirements and SpecificationsSpecificationsSpecificationsTabl e 2 Performanc
Agilent 1260 Infinity High Performance Autosampler User Manual 29Site Requirements and Specifications2SpecificationsLocal Control Agilent Instant Pilo
Agilent 1260 Infinity High Performance Autosampler User Manual 3In This GuideIn This GuideThis manual covers the Agilent 1260 Infinity High Performanc
30 Agilent 1260 Infinity High Performance Autosampler User Manual2 Site Requirements and SpecificationsSpecifications
31Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies3Installing the AutosamplerUnpacking the Autosampler 32Damaged Pac
32 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerUnpacking the AutosamplerUnpacking the AutosamplerDamaged
Agilent 1260 Infinity High Performance Autosampler User Manual 33Installing the Autosampler3Unpacking the AutosamplerDelivery Checklist Ensure all par
34 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerUnpacking the AutosamplerAutosampler Accessory Kit Conten
Agilent 1260 Infinity High Performance Autosampler User Manual 35Installing the Autosampler3Optimizing the Stack ConfigurationOptimizing the Stack Con
36 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerOptimizing the Stack ConfigurationOne Stack Configuration
Agilent 1260 Infinity High Performance Autosampler User Manual 37Installing the Autosampler3Optimizing the Stack ConfigurationFigure 8 Recommended Sta
38 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerOptimizing the Stack ConfigurationTwo Stack Configuration
Agilent 1260 Infinity High Performance Autosampler User Manual 39Installing the Autosampler3Optimizing the Stack ConfigurationFigure 10 Recommended Tw
4 Agilent 1260 Infinity High Performance Autosampler User ManualIn This Guide8Test FunctionsThis chapter describes the tests for the module.9 Maintena
40 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerInstallation Information on Leak and Waste HandlingInstal
Agilent 1260 Infinity High Performance Autosampler User Manual 41Installing the Autosampler3Installation Information on Leak and Waste HandlingFigure
42 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerInstallation Information on Leak and Waste Handling1 Stac
Agilent 1260 Infinity High Performance Autosampler User Manual 43Installing the Autosampler3Installation Information on Leak and Waste HandlingFigure
44 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerInstalling the AutosamplerInstalling the Autosampler1 Pla
Agilent 1260 Infinity High Performance Autosampler User Manual 45Installing the Autosampler3Installing the Autosampler3 Connect the power cable to the
46 Agilent 1260 Infinity High Performance Autosampler User Manual3 Installing the AutosamplerFlow Connections to the AutosamplerFlow Connections to th
47Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies4Using the ModuleLeak and Waste Handling 48Preparing the Autosampl
48 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleLeak and Waste HandlingLeak and Waste HandlingFor details on correc
Agilent 1260 Infinity High Performance Autosampler User Manual 49Using the Module4Preparing the AutosamplerPreparing the AutosamplerFor best performan
Agilent 1260 Infinity High Performance Autosampler User Manual 5ContentsContents1 Introduction 9Features 10Overview of the Module 11Autosampler Princi
50 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModulePreparing the Autosampler• Priming and Purging the System - When t
Agilent 1260 Infinity High Performance Autosampler User Manual 51Using the Module4Setting up the Autosampler with Agilent ChemStationSetting up the Au
52 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleSetting up the Autosampler with Agilent ChemStationFigure 14 ChemSt
Agilent 1260 Infinity High Performance Autosampler User Manual 53Using the Module4Setting up the Autosampler with Agilent ChemStationThe Autosampler U
54 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleSetting up the Autosampler with Agilent ChemStationModule Status sh
Agilent 1260 Infinity High Performance Autosampler User Manual 55Using the Module4Setting up the Autosampler with Agilent ChemStationControl SettingsT
56 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleSetting up the Autosampler with Agilent ChemStationMethod Parameter
Agilent 1260 Infinity High Performance Autosampler User Manual 57Using the Module4Setting up the Autosampler with Agilent ChemStationInjection ModeThe
58 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleSetting up the Autosampler with Agilent ChemStationInjection Cleani
Agilent 1260 Infinity High Performance Autosampler User Manual 59Using the Module4Setting up the Autosampler with Agilent ChemStationInjection Program
6 Agilent 1260 Infinity High Performance Autosampler User ManualContents6 Troubleshooting and Diagnostics 79Overview of the Module’s Indicators and Te
60 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleSetting up the Autosampler with Agilent ChemStationModule Configura
Agilent 1260 Infinity High Performance Autosampler User Manual 61Using the Module4Main Screens of the Autosampler with Agilent Instant Pilot (G4208A)M
62 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleMain Screens of the Autosampler with Agilent Instant Pilot (G4208A)
Agilent 1260 Infinity High Performance Autosampler User Manual 63Using the Module4Main Screens of the Autosampler with Agilent Instant Pilot (G4208A)T
64 Agilent 1260 Infinity High Performance Autosampler User Manual4 Using the ModuleMain Screens of the Autosampler with Agilent Instant Pilot (G4208A)
65Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies5Optimizing PerformanceDelay Volume and Extra-Column Volume 66Dela
66 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceDelay Volume and Extra-Column VolumeDelay Volume and Extra-Co
Agilent 1260 Infinity High Performance Autosampler User Manual 67Optimizing Performance5How to Configure the Optimum Delay VolumeHow to Configure the
68 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Configure the Optimum Delay VolumeFigure 16 Schematic
Agilent 1260 Infinity High Performance Autosampler User Manual 69Optimizing Performance5How to Configure the Optimum Delay VolumeThe ADVR function is
Agilent 1260 Infinity High Performance Autosampler User Manual 7Contents10 Parts for Maintenance 145Overview of Maintenance Parts 146Vial Trays 147Rec
70 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Achieve Higher Injection VolumesHow to Achieve Higher
Agilent 1260 Infinity High Performance Autosampler User Manual 71Optimizing Performance5How to Achieve Higher Injection Volumesbut it is very dependen
72 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Achieve High ThroughputHow to Achieve High ThroughputT
Agilent 1260 Infinity High Performance Autosampler User Manual 73Optimizing Performance5How to Achieve Higher ResolutionHow to Achieve Higher Resoluti
74 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Achieve Higher ResolutionThe resolution equation shows
Agilent 1260 Infinity High Performance Autosampler User Manual 75Optimizing Performance5How to Achieve Higher ResolutionIn isocratic separations, incr
76 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Achieve Higher SensitivityHow to Achieve Higher Sensit
Agilent 1260 Infinity High Performance Autosampler User Manual 77Optimizing Performance5How to Achieve Lowest Carry OverHow to Achieve Lowest Carry Ov
78 Agilent 1260 Infinity High Performance Autosampler User Manual5 Optimizing PerformanceHow to Achieve Lowest Carry OverThe flush port is located abo
79Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies6Troubleshooting and DiagnosticsOverview of the Module’s Indicator
8 Agilent 1260 Infinity High Performance Autosampler User ManualContents14 Appendix 193General Safety Information 194Lithium Batteries Information 197
80 Agilent 1260 Infinity High Performance Autosampler User Manual6 Troubleshooting and DiagnosticsOverview of the Module’s Indicators and Test Functio
Agilent 1260 Infinity High Performance Autosampler User Manual 81Troubleshooting and Diagnostics6Status IndicatorsStatus IndicatorsTwo status indicato
82 Agilent 1260 Infinity High Performance Autosampler User Manual6 Troubleshooting and DiagnosticsStatus IndicatorsModule Status IndicatorThe module s
Agilent 1260 Infinity High Performance Autosampler User Manual 83Troubleshooting and Diagnostics6User InterfacesUser Interfaces• Depending on the user
84 Agilent 1260 Infinity High Performance Autosampler User Manual6 Troubleshooting and DiagnosticsAgilent Lab Advisor SoftwareAgilent Lab Advisor Soft
85Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies7Error InformationWhat are Error Messages 87General Error Messages
86 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationAgilent Lab Advisor SoftwareVessel or Wash Vessel Error 103Vessel
Agilent 1260 Infinity High Performance Autosampler User Manual 87Error Information7What are Error MessagesWhat are Error MessagesError messages are di
88 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationGeneral Error MessagesGeneral Error MessagesGeneral error messages
Agilent 1260 Infinity High Performance Autosampler User Manual 89Error Information7General Error Messages Remote Timeout Error ID: 0070A not- ready co
9Agilent 1260 Infinity High Performance Autosampler User ManualAgilent Technologies1IntroductionFeatures 10Overview of the Module 11Autosampler Princi
90 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationGeneral Error MessagesLeakError ID: 0064A leak was detected in the
Agilent 1260 Infinity High Performance Autosampler User Manual 91Error Information7General Error MessagesLeak Sensor ShortError ID: 0082The leak senso
92 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationGeneral Error MessagesCompensation Sensor ShortError ID: 0080The a
Agilent 1260 Infinity High Performance Autosampler User Manual 93Error Information7Module Error MessagesModule Error MessagesThese errors are autosamp
94 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesSide Door ErrorError ID: 4355, 4459The side d
Agilent 1260 Infinity High Performance Autosampler User Manual 95Error Information7Module Error MessagesValve to Bypass FailedError ID: 4014, 4701The
96 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesNeedle Lock FailedError ID: 4702, 4703The loc
Agilent 1260 Infinity High Performance Autosampler User Manual 97Error Information7Module Error MessagesNeedle to Needle Seat PositionError ID: 4510,
98 Agilent 1260 Infinity High Performance Autosampler User Manual7 Error InformationModule Error MessagesMissing Vial or Missing Wash VialError ID: 40
Agilent 1260 Infinity High Performance Autosampler User Manual 99Error Information7Module Error MessagesInitialization FailedError ID: 4020The autosam
Comments to this Manuals