Agilent EXA Signal Analyzer
Inputs/Outputs
Description Specifications Supplemental Information
Trigger Outputs
(Trigger 1 Out, Trigger 2 Out)
Connector BNC female
Impedance 50Ω (nominal)
Level 0 to 5 V (CMOS)
Description Specifications Supplemental Information
Monitor Output
Connector
Format
Resolution
VGA compatible,
15-pin mini D-SUB
1024 × 768
XGA (60 Hz vertical sync rates,
non-interlaced)
Analog RGB
Description Specifications Supplemental Information
Analog Out Refer to Chapter 15 , “Option YAS -
Y-Axis Screen Video Output,” on page
151 for more details.
Connector BNC female
Impedance <140Ω (nominal)
Description Specifications Supplemental Information
Noise Source Drive +28 V (Pulsed)
Connector BNC female
Output voltage on 28.0 ± 0.1 V 60 mA maximum current
Output voltage off < 1.0 V
Description Specs Supplemental Information
SNS Series Noise Source For use with Agilent Technologies SNS Series noise sources
Description Specifications Supplemental Information
Digital Bus This port is intended for use with the Agilent N5105 and N5106
products only. It is not available for general purpose use.