Agilent Technologies 22D User's Guide Page 176

  • Download
  • Add to my manuals
  • Print
  • Page
    / 276
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 175
5-6
Making Measurements
To reduce the random noise on a signal
To reduce the random noise on a signal
If the signal you are applying to the oscilloscope is noisy, you can set up the
oscilloscope to reduce the noise on the displayed waveform. First, you stabilize
the displayed waveform by removing the noise from the trigger path. Second,
you reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Remove the noise from the trigger path by turning on either
high-frequency reject or noise reject.
High-frequency reject (HF reject) adds a low-pass filter with the 3-dB point at
50 kHz. You use HF reject to remove high-frequency noise, such as AM or FM
broadcast stations, from the trigger path.
HF Reject
3 dB down point
Pass
Band
0 dB
50 kHz
dc
Page view 175
1 2 ... 171 172 173 174 175 176 177 178 179 180 181 ... 275 276

Comments to this Manuals

No comments