Agilent Technologies 3000 X-Series Service Manual Page 140

  • Download
  • Add to my manuals
  • Print
  • Page
    / 156
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 139
8Service
140 Agilent 3000 Series Oscilloscopes User’s Guide
9 Press Acquire
.
10 Select the Mode menu item until “Average” appears.
11 Select the Average menu item until “8” appears.
12 Press Measure
.
13 Select the Voltage menu item.
14 Select the Vpp menu item.
15 Set the signal generator to a 1 MHz sine wave with a peak-to- peak
amplitude of about 6 divisions as it appears on the oscilloscope screen.
Figure 60 Averages Menu Item
Page view 139