Chapter 2 7
Option Installation
Procedure
Saving System calibration Data from Removable Hard Disk to USB
Pen Drive
The following is a description of the procedures to save the system calibration data stored
in the removable hard disk to a USB Pen drive.
Step 1. Exit the measurement window.
System > Service Menu > Administorator Menu > Exit E5052B > OK
Step 2. Insert a USB Pen drive into the USB port in front panel.
Step 3. Open D:\Recovery drive.
Step 4. Click .D:\syscal from the explorer.
Step 5. Select Edit - Copy from menu bar of the explorer.
Step 6. Click the USB Pen drive of the explorer.
Step 7. Select Edit - Paste from menu bar of the explorer.
Step 8. Click .D:\limit from the explorer.
Step 9. Select Edit - Copy from menu bar of the explorer.
Step 10. Click the USB Pen drive of the explorer.
Step 11. Select Edit - Paste from menu bar of the explorer.
Step 12. Turn off the E5052B, and then remove the cord from the outlet before removing the
Removable Hard Disk.