FOLD FOLD
- turns the 34401A on and off. A self-
test occurs during the power-on
sequence.
- selects the front or rear panel
terminals as the input to the
34401A.
- transfers instrument control from the
computer (remote) to the front panel
(local).
Example: display indicating remote operation and rear panel
terminals selected.
2. Power, Input Terminal Selection,
and Local Control
The measurement function is set using the
following keys:
- when preceded by the “Shift” key the
functions shown above the keys are
selected.
Example: Period function selected
3. Setting the Function
Resolution is the number of digits the 34401A
can measure or display. Resolution is also
unique to the function selected and is set using
the following keys:
- selects 4 digit resolution
- selects 5 digit resolution
- selects 6 digit resolution
For increased measurement speed, select 4
digits. For increased accuracy, select 6 digits.
6. Triggering
Measurements are triggered from the front panel
using the following keys:
- disables auto-triggering (default) and
issues a single trigger to the 34401A
each time the key is pressed.
- Auto: enables auto-triggering if
the 34401A is in single trigger
mode (above).
- Hold: displays the measurement
after three consecutive readings
within the sensitivity band
(Trigger menu:
READ HOLD).
Front Panel Menu Reference
A: MEAS MENU
1: AC FILTER > 2: CONTINUITY > 3: INPUT R > 4: RATIO FUNC > 5: RESOLUTION
B: MATH MENU
1: MIN-MAX > 2: NULL VALUE > 3: dB REL > 4: dBm REF R > 5: LIMIT TEST > 6: HIGH LIMIT > 7: LOW LIMIT > 8: SCALE >
9: SCALE GAIN > 10: SCALE OFST
C: TRIG MENU
1: READ HOLD > 2: TRIG DELAY > 3: N SAMPLES
D: SYS MENU
1: RDGS STORE > 2: SAVED RDGS > 3: ERROR > 4: TEST > 5: DISPLAY > 6: BEEP > 7: COMMA > 8: REVISION >
9: STORE STATE > 10: RCL STATE > 11: POWER ON
E: I/O MENU
1: GPIB ADDR > 2: INTERFACE > 3: BAUD RATE > 4: PARITY > 5: LANGUAGE
F: CAL MENU
1: SECURED > [1: UNSECURED > 2: CALIBRATE] > 3: CAL COUNT > 4: MESSAGE
G: TEMP MENU
1: TEMP FUNC > 2: UNITS > 3: RTD TYPE 4: RTD Ro > 5: T/C TYPE > 6: THERM TYPE > 7: JUNCT TEMP
NOTE TO PRINTER:
This is FACE “B”
(See folding diagram on page 1)
NOTE TO PRINTER:
This is FACE “C”
(See folding diagram on page 1)
NOTE TO PRINTER:
This is FACE “D” (short face, folds inside)
(See folding diagram on page 1)
- dB measurement relative to value
in ‘dB Relative’ register (register
value is first db measurement
after dB function is enabled).
- dBm measurement relative to a
resistance referenced to 1 mW
(Math menu:
dBm REF R).
reading = 10Log
10
(measurement
2
/ reference resistance/ 1mW)
Example: dBm measurement @ ~ 3 VDC - 600 reference
resistance
Selected math operations are available from the
34401A front panel using the keys:
- value subtracted from the measurement
(Math menu:
NULL VALUE)
reading = measurement – null value
- displays minimum and maximum values
for the triggered set of readings.
Example: maximum value indicated with Math operation
‘Min/Max’ enabled
Menu Navigation
The front panel menu provides access to a
subset of the 34401A functionality. The menu is
navigated using the following keys:
Sequence:
- enables the front panel menu
- scrolls through menus ‘A’ - ‘G’
- selects menu command list and
scrolls through commands
- selects parameter list and scrolls
through parameters
- selects (saves) parameter and exits
the menu
(Enter is not a shifted function).
- enables menu (if off) and recalls
the last command displayed
when the menu was exited.
Note: for numeric parameters (0-9):
- selects and increments/
decrements each digit
individually
Example: number of reading samples increased to 5
0000 SMPLS
04.09495 VDC
Adrs Rmt Rear
05.02287 MAX
MaTh
200.001 uSEC
11.7842 DBm
MaTh
Measurement Configuration
Function DCV
AC Filter 20 Hz
Range Autorange
Resolution 5 digits, slow mode
Integration Time 10 PLCs
Autozero On
Input Resistance 10 M
Continuity Threshold 10
Math Operations
State Off
Function Null
dBm Reference Resistance 600
Registers Cleared
Triggering
Trigger Source Auto-trigger
Trigger Delay Automatic
Samples Per Trigger 1
Reading Hold Sensitivity 0.10% of range
Other
Reading Memory Cleared
34401A Power-on and Reset States
4. Setting the Range and Resolution
5. Math Operations
The measurement range is unique to the
selected function. The range is set using the
following keys:
- selects between auto-range and manual
range
- manually decreases the range
- manually increases the range
Example: display indicating 10 volt range, 4 digit resolution
03.987 VDC
MAN