H
Effective Impedance Measurement
Using OPEN/SHORT/LOAD
Correction
Application Note 346-3
This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard
and describes products or services now available through Agilent. It may also refer to products/services no longer supported by Agilent. We
regret any inconvenience caused by obsolete information. For the latest information on Agilent’s test and measurement products go to:
www.agilent.com/find/products
Or in the U.S., call Agilent Technologies at 1-800-452-4844 (8am8pm EST)
Page view 0
1 2 3 4 5 6 ... 11 12

Summary of Contents

Page 1 - Application Note 346-3

HEffective Impedance MeasurementUsing OPEN/SHORT/LOADCorrectionApplication Note 346-3This literature was published years prior to the establishment of

Page 3

10Model No. Correction Capability Cable Length CorrectionHP 4192A OPEN/SHORT Correction 0m/1m HP 4194A OPEN/SHORT Correction 0m/1m HP 4195A(*

Page 4 - Correction Model

HFor more information about Hewlett-Packard test & measurement products,applications, services, and for acurrent sales office listing, visitour we

Page 5

1Figure 1. OPEN/SHORT Correction ModelSince Zs <<1/Yo, stray admittance Yo can be measured when the test termi-nals are open. Similarly residual

Page 6 - Figure 3. Cable Extension

2 Zm-ZsZdut= ---------------------------------- l-(Zm-Zs)Yowhere,Zdut: True value of DUTZm: Measurement val

Page 7 - Figure 4. Comparison

3Figure 2. OPEN/SHORT/LOAD Correction ModelWith the assumption that a DUT with an impedance of Z2 is connected to the front panel terminals, the instr

Page 8 - Conclusion

4Otherwise, when using other LCR meters or analyzers, a computer can be programmed to make the OPEN/SHORT/LOAD corrections through HP-IB.Consideration

Page 9

5Figure 3. Cable Extension Using an HP 16048E (4m)impedance value. However, when the DUT's impedance value is very low or very high, it is recomm

Page 10

6Figure 4. ComparisonBetween OPEN/SHORT Correction and OPEN/SHORT/LOAD CorrectionOPEN/SHORT/LOAD correction with an external controllerThe OPEN/SHORT/

Page 11 - (*) Obsolete

7ConclusionThis application note shows the principle of the OPEN/SHORT/LOAD correction and some points to be considered in the execution. With the pro

Page 12

8Figure 5. Sample Program for HP 4194A OPEN/SHORT/LOAD Correction