Agilent Technologies 4352B User Manual Page 11

  • Download
  • Add to my manuals
  • Print
  • Page
    / 16
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 10
Source characteristics
Dc power supply
Dc voltage level :
0 to +15.5 V (50 mA max.)
Dc voltage resolution:
•1 mV
Dc voltage accuracy:
±(0.2 % + 2 mV)
Noise density:
< 10 nVHz (@ 10 kHz, typical)
Connector type:
• BNC (F)
Dc control source
Dc voltage level:
0 to +20 V (20 mA max.)
Option 001:
-15 to +35 V (20 mA max.)
Dc voltage resolution:
• 100 µV
Dc voltage accuracy:
±(0.1% + 2 mV)
Noise density:
< 1 nVHz (@ 10 kHz, typical)
Settling time:
< 20 msec at 0.1 % error (typical)
Output resistance (dc):
• < 10 (typical)
Connector type:
• BNC (F)
1 kHz signal source
Frequency:
• 1 kHz
Level:
0 to 1 Vrms (@ open load)
Level resolution:
• 1 mVrms
Level accuracy:
• 1 mVrms
Output impedance:
• 50 nominal (typical)
Connector type:
• BNC (F)
Receiver
characteristics
Frequency range:
10 MHz to 3 GHz
Input level:
-10 to +20 dBm
Input Impedance:
• 50
SWR:
< 1.2 (@ < 2 GHz),
< 1.3 (@ 2 to 3 GHz)
Connector type:
• N-type (F)
Measurement parameters
RF power
Input level:
-10 to +20 dBm
Resolution:
• 0.01 dB
Accuracy:
±0.6 dB (@ 2 GHz, 15 dBm)
±1 dB (@ other conditions)
±0.2 dB (@ 1 GHz, -5 dBm, typical)
Number of measurement points per sweep
(analyzer mode):
2 to 801 points
Note: cable loss compensation
The 4352B can compensate for the RF power
level loss of the cable connecting the DUT output
terminal and the 4352B RF IN connector when
measuring RF power.
Frequency
Resolution:
• 1 kHz
Accuracy:
±(1 kHz + Time base accuracy of the external
signal generator)
Number of measurement points per sweep
(analyzer mode):
2 to 801 points
Phase noise (carrier-to-noise ratio)
Offset frequency range:
100 Hz to 10 MHz
Noise floor:
Offset
frequency Spec Typical
100 Hz -85 dBc/Hz -90 dBc/Hz
1 kHz -110 dBc/Hz -117 dBc/Hz
10 kHz -130 dBc/Hz -137 dBc/Hz
100 kHz -140 dBc/Hz -147 dBc/Hz
1 MHz -150 dBc/Hz -157 dBc/Hz
Note: The phase noise of signal generator isn’t
included in these values.
Accuracy:
±4 dB @ 100 Hz - 1 kHz (typical)
±2 dB @ 1 kHz - 1 MHz
±4 dB @ 1 MHz - 10 MHz
Spectrum
Span:
10 MHz (Max.)
Resolution band width:
1 Hz to 3 kHz (1, 3 step)
Noise floor:
< -95 dBm (@ RBW = 30 Hz, typical)
< -75 dBm (@ RBW = 3 kHz, typical)
Absolute level accuracy:
±2 dBm (@ -5 dBm, typical)
Relative level accuracy:
• ±1.5 dB
±0.5 dB (typical)
Agilent 4352B VCO/PLL Signal Analyzer Specification Summary
11
Page view 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16