Agilent PXT Wireless Communications Test Set
User’s Guide
166
To load a scenario on Cell B:
Select the B-Cell after pressing BSE > Config > Cell Setup > Cell Selection.
Press Mode Setup > Call Scenario. Select a scenario file and load it into the setup.
Note that the “B-Cell Scenario File Name” window, now displays the scenario loaded, as shown
below.
Figure 5 - 2: B-Cell Scenario File Loaded
General Parameters
Most PHY Parameters are independent between Cell A and Cell B. For simplicity, the settings may be
copied from one to the other via the page as shown below: Press BSE > Config > Cell Setup > Cell
Selection.
Figure 5 - 3: Copying PHY Parameters
For a UE to differentiate between the two cells, the Physical Cell ID should be
different for each.
Power
The relative power allocated to each of the two cells is configured as a percentage of the total. Set this by
pressing: Config > Cell Setup.