3 Press the Vrms softkey.
The oscilloscope automatically measures the rms voltage and displays the
result on the display.
The oscilloscope makes automatic measurements on the first pulse or period
in the display. If a cycle of the waveform cannot be found as shown in the
delayed window in figure 2-10, the measurement is made using the delayed
window as the cycle. Figure 2-10 shows how to use delayed sweep to isolate
a pulse for an rms measurement.
Delayed sweep isolates an area of interest for an rms voltage measurement
Figure 2-10
Operating Your Oscilloscope
To make voltage measurements automatically
2–21