6 In the Tree View Panel, navigate to the test category you want to execute. Select the
category via IQ/OQSW/HWPC name/Bioanalyzer nameTest Category.
NOTE
To execute hardware tests (HW branch) the bioanalyzer must be properly connected
and switched on.
The Configuration tab now lets you select qualification tests to be executed in the
validation run:
To select tests, check the Apply check box next to the test(s).
7 To start the selected tests, click on Start button in the toolbar.