Page view 39
37
N3970A UM-1
Model N3970A Optical Power Meter
Total Uncertainty/Accuracy Test
Test Wavelengths = 1300 nm and 1310 nm
Source Wavelength = nm
Report No. Date
PERFORMANCE TEST
Setting
Number
1
2
3
4
-3.00 dBm
-20.00 dBm
-30.00 dBm
-50.00 dBm
-3.30 dBm
-20.30 dBm
-30.30 dBm
-50.30 dBm
-2.70 dBm
-19.70 dBm
-29.70 dBm
-49.70 dBm
Measurement Uncertainty dB
dB
dB
dB
dB
dBm
dBm
dBm
dBm
Reference
(8163A Display)
Attenuator
Setting
Minimum Spec.
(-0.3 dB of Ref.)
N3970A, DUT
Measurement results
Maximum Spec.
(+0.3 dB of Ref.)