4-6
Calibrating and Adjusting
To adjust the power supply
4
Adjust R40 if necessary to bring the +3.4 V voltage into tolerance. The
adjustment location is on the power supply as shown below.
Figure 4-2
Low Voltage Power Supply Adjustments (on the top of the oscilloscope)
5 If adjusting the power supply does not bring all the voltages within
tolerance, go to the “Troubleshooting” chapter.
Fan Connector
+3.4V Adjustment
service.book Page 6 Wednesday, August 28, 2002 3:25 PM