Technical SpecificationsAgilent TechnologiesPNA Series Network AnalyzersE8362B/C, E8363B/C, and E8364B/CManufacturing Part Number: E8364-90031Printed
6 the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic range is required (i.e. device
96 Table 32. X11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
97NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
98 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of
99Table 33. Uncorrected System Performancea Specifications apply over environmental temperature of 23° ±3 °C, with < 1 °C deviation from the calib
100 Table 33 (Continued). Uncorrected System Performancea Reflection Tracking Typical: 10 MHz to 45 MHzb -- ±1.5 dB 45 MHz to 20 GHz -- ±
101Table 33 (Continued). Uncorrected System Performancea Crosstalk - Option 080 enabledb,e Typical: 10 MHz to 45 MHz -- -65 45 MHz to 1 GHz
102 Table 34. Test Port Output Description Specification Supplemental Frequency Range Standard Opt 014 Opt UNL Opt 014/UNL E8362B/C 10 MHz
103Table 34 (Continued). Test Port Output Power Rangea, e, f 10 MHz to 45 MHzb -25 to +2 dBm (typical) -25 to +2 dBm (typical)-87 to +2 dBm (typi
104 Table 34 (Continued). Test Port Output 10 kHz offset from center frequency, nominal power at test port Typical: 10 MHz to 45 MHz -
105Table 34 (Continued). Test Port Output 1 MHz offset from center frequency, nominal power at test port - Option 080 enabled Typical:
7Table 2. Receiver Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Dynamic Range (in a 1
106 Table 35: Test Port Input Description Specification Supplemental Standard Opt 014 Opt UNL Opt 014/UNL Test Port Noise Floora 10 Hz IF Ba
107Table 35 (Continued). Test Port Input Test Port Noise Floora,b Option 080 enabledd 10 Hz IF Bandwidth 10 MHz to 45 MHzb <-77 dBm (typical) <
108 Table 35 (Continued). Test Port Input Direct Receiver Access Input Noise Floora,b 10 Hz IF Bandwidth 10 MHz to 45 MHz -- <-127 dBm -- &l
109Table 35 (Continued). Test Port Input Direct Receiver Access Input Noise Floora,b - Option 080 enabledd 10 Hz IF Bandwidth 10 MHz to 45 MHz -- &
110 Table 35 (Continued). Test Port Input Third Order Intercept - Tone spacing from 100 kHz - 5 MHz Typical 10 MHz to 150 MHz -- +33 dBm 150
111Table 35 (Continued). Test Port Input Trace Noise Magnitude - Option 080 enabledb,d 1 kHz IF bandwidth. Ratio measurement, nominal power at test p
112 Table 35 (Continued). Test Port Input Stability Magnituded Typical ratio measurement, made at the test port. 10 MHz to 45 MHz -- -- -- --
113Table 36. Dynamic Accuracy (Specificationa) Accuracy of the test port input power reading relative to the reference input power level. NOTE: The
114
115
8 Table 2 (Continued). Receiver Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Dynamic
116 a Dynamic accuracy is verified with the following measurements: Compression over frequency. IF linearity at a single frequency of 1.
117Table 37. Test Port Input (Group Delay)a Description Specification Supplemental Information (typ.) Aperture (selectable) (frequency span)/(num
118 General Information Table 38. Miscellaneous Information Description Specification Supplemental Information System IF Bandwidth Range -- 1 Hz
119Table 40. Rear Panel Information Description Supplemental Information 10 MHz Reference In Connector BNC, female Input Frequency 10 MHz ± 10 ppm
120 Table 40 (Continued). Rear Panel Information Test Set Drivers Function Used for driving remote mixers Connectors SMA (f) for RF and LO outputs
121Table 40 (Continued). Rear Panel Information USB Port One port on front panel and five ports on rear panel. Universal Serial Bus jack, Type A c
122 Table 41. Analyzer Dimensions and Weight Description Supplemental Information Cabinet Dimensions Height Width Depth Excluding front and rear
123Measurement Throughput Summary Table 42 Typical Cycle Timea,b (ms) for Measurement Completion Number of Points 201 401 1601 16,001 Start 28 G
124 Table 43. Cycle Time vs IF Bandwidtha Applies to the Preset condition (201 points, correction off) except for the following changes: CF = 28 G
125Table 45. Frequency Converter Application (option 083) Cycle Time for Fixed-IF Measurements (ms)1 Number of Points 101 201 401 Stimulus star
9E8363B/C AND E8364B/C Corrected System Performance with 2.4mm Connectors ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Table 3. 85056A Calibration Ki
126 Specifications: Front-Panel Jumpers ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Models E8362B/C, E8363B/C, and E8364B/C Option 014 ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
127 E8363B/C: 10 Hz IF Bandwidth 45 MHz to 500 MHz < -127 dBm -- 500 MHz to 2 GHz < -133 dBm -- 2 GHz to 10 GHz <
128 Table 48: Reference Receiver Inputs (Rcvr R1, Rcvr R2) Description Specification Supplemental Information Minimum Input Level E8362B/C, E8363
129 Table 49: Reference Outputs (Reference 1 Source Out, Reference 2 Source Out) Description Specification Supplemental Information Maximum Output
130 Table 50: Source Outputs (Port 1 Source Out, Port 2 Source Out) Description Specification Supplemental Information Maximum Output Level E8362
131 E8364B/C, Option 014 and UNL: 45 MHz to 500 MHz -- 3.5 dBm, typical 500 MHz to 2 GHz -- 5 dBm, typical 2 GHz to 10 GHz --
132 Table 51: Coupler Inputs (Port 1 Cplr Thru, Port 2 Cplr Thru) Description Specification Supplemental Information Insertion Loss to Test Port
133 E8364B/C, Option 014 and UNL: 45 MHz to 500 MHz -- 0.5 dB, typical 500 MHz to 2 GHz -- 1 dB, typical 2 GHz to 10 GHz -- 2
134 Test Set Block Diagrams E836xB/C - Standard Configuration and Standard Power Range E836xB/C - Option UNL Standard Configuration with Extende
135E836xB/C - Option UNL Standard Configuration with Extended Power Range and Bias - Tees, and Option 016, Receiver Attenuators
10 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
136 Test Set with Option 014 Block Diagrams E836xB/C - Option 014 Configurable Test Set and Standard Power Range Item Description Item Descript
137E836xB/C - Option 014 Configurable Test Set and Standard Power Range, and Option 081 Reference Channel Transfer Switch Item Description Item
138 E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees Item Description Item Description a SOURCE
139E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees, and Option 081 Reference Channel Transfer Switch
140 E836xB/C - Option 014 Configurable Test Set and Option UNL, Extended Power Range and Bias - Tees and Option 016 Receiver Attenuators Item Desc
141E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees, and Option 016 Receiver Attenuators, and Option
11 Reflection Uncertainty (Specifications)
12 Table 4. 85056A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
13NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
14 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of
15Table 5. 85056D Calibration Kit Standard Configuration and Standard Power Range (E8363B/C AND E8364B/C) Applies to the, E8363B/C AND E8364B/C an
Documentation WarrantyTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTU
16 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
17 Reflection Uncertainty (Specifications)
18 Table 6. 85056D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
19NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E
20 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offs
21E8363B/C AND E8364B/C Corrected System Performance with 2.92mm Connectors Table 7. 85056K Calibration Kit Standard Configuration and Standard Powe
22 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
23 Reflection Uncertainty (Specifications)
24 Table 8. 85056K Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
25 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications * Configurable Test Set,
Contacting AgilentAssistance with test and measurement needs and information on finding a local Agilent office are available on the Web at:www.agilent
26 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offse
27E836xB/C Corrected System Performance with 3.5mm Connectors ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Table 9. 85052B Calibration Kit Standard
28 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
29 Reflection Uncertainty (Specifications)
30 Table 10. 85052B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
31NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
32 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of
33Table 11. 85052C Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85052C (3.5mm) c
34 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
35 Reflection Uncertainty (Specifications)
This page intentionally left blank.
36 Table 12. 85052C Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
37NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
38 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency O
39Table 13. 85052D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85052D (3.5mm) c
40 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
41 Reflection Uncertainty (Specifications)
42 Table 14. 85052D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
43NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
44 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency O
45E836xB/C Corrected System Performance with 7mm Connectors Table 15. 85050B Calibration Kit Standard Configuration and Standard Power Range (E836
1Technical Specifications for the E8362B/C, E8363B/C, E8364B/C (Rev. 2008-10-03) Definitions...
46 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
47 Reflection Uncertainty (Specifications)
48 Table 16. 85050B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
49NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set, E
50 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency
51Table 17. 85050C Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85050C (7mm) cal
52 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
53 Reflection Uncertainty (Specifications)
54 Table 18. 85050C Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
55NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set,
2 Standard Power Range (E836xB/C)... 63 Table 22. 8505
56 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency
57Table 19. 85050D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85050D (7mm) cal
58 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
59 Reflection Uncertainty (Specifications)
60 Table 20. 85050D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
61NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set
62 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequen
63E836xB/C Corrected System Performance with Type-N Connectors Table 21. 85054B Calibration Kit Standard Configuration and Standard Power Range (E
64 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
65 Reflection Uncertainty (Specifications)
3Table 48: Reference Receiver Inputs (Rcvr R1, Rcvr R2) ... 128 Table 49: Reference Outputs
66 Table 22. 85054B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
67NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
68 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency
69Table 23. 85054D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85054D (Type-N)
70 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
71 Reflection Uncertainty (Specifications)
72 Table 24. 85054D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
73NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E
74 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of
75E8363B/C AND E8364B/C Corrected System Performance with WR-28 Connectors Table 25. R11644A Calibration Kit Standard Configuration and Standard Pow
4 Definitions All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument ha
76 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
77 Reflection Uncertainty (Specifications)
78 Table 26. R11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
79NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E
80 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offs
81E8363B/C AND E8364B/C Corrected System Performance with WR-42 Connectors Table 27. K11644A Calibration Kit Standard Configuration and Standard Pow
82 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
83 Reflection Uncertainty (Specifications)
84 Table 28. K11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
85NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
5Table 1. System Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Supplemental InformationD
86 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of
87E836xB/C Corrected System Performance with WR-62 Connectors Table 29. P11644A Calibration Kit Standard Configuration and Standard Power Range (E
88 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
89 Reflection Uncertainty (Specifications)
90 Table 30. P11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &
91NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,
92 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequenc
93E836xB/C Corrected System Performance with WR-90 Connectors Table 31. X11644A Calibration Kit Standard Configuration and Standard Power Range (E
94 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)
95 Reflection Uncertainty (Specifications)