Agilent Technologies EasyEXPERT User Manual

Browse online or download User Manual for Measuring, testing & control Agilent Technologies EasyEXPERT. Agilent Technologies EasyEXPERT User manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 14
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Introduction
Agilent 4155C and 4156C
Basic functions
• Set measurement and/or stress
conditions
• Control measurement and/or
stress execution
• Perform arithmetic calculations
• Display measured and calculated results
on the LCD display
• Perform graphical analysis
• Store and recall measurement setups, and
measurement and graphical display data
• Dump to printers or plotters
for hardcopy output
• Perform measurement and analysis with
built-in instrument BASIC
• Self test, Auto calibration
Confi guration
The 4155C and 4156C are both furnished
with Desktop EasyEXPERT standard edition.
If you want the Desktop EasyEXPERT Plus
edition, you can request the B1541A-002
when you order a 4155C or 4156C. For more
information about the Desktop EasyEXPERT
Plus, please refer to page 11 of this data
sheet.
Agilent 4155C Semiconductor
Parameter Analyzer
Agilent 4156C Precision Semiconductor
Parameter Analyzer
Data Sheet
4155C 4156C
4xMPSMU 4xHRSMU
2xVMU 2xVMU
2xVSU 2xVSU
Desktop
EasyEXPERT
Desktop
EasyEXPERT
SMU: Source monitor unit
Display resolution: 6 digits at each current
range (0.01 fA display resolution at 10
pA range)
2
HRSMU: High Resolution SMU
(1 fA/2 μV to 100 mA/100 V)
MPSMU: Medium Power SMU
(10 fA/2 μV to 100 mA/100 V)
HPSMU: High Power SMU
(10 fA/2 μV to 1 A/200 V)
VMU: Voltage Monitor Unit
(0.2 μV resolution in differential mode)
VSU: Voltage Source Unit
PGU: Pulse Generator Unit
(1 channel)
GNDU: Ground Unit
2. Accuracy not guaranteed. Minimum
guaranteed resolution is 1 fA at 10 pA range.
1. Minimum number of installable MPSMU
or PGU is two.
41501B (Optional)
GNDU
2xPGU (Option)¹
HPSMU (Option) or
2xMPSMU (Option)¹
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - Parameter Analyzer

IntroductionAgilent 4155C and 4156C Basic functions• Set measurement and/or stress conditions• Control measurement and/or stress execution • Perform a

Page 2 - Hardware

10PGU port signal transfer characteristicsOvershoot: <5% of pulse amplitude (@20 ns leading and trailing time, 50 Ω pulse generator source impedanc

Page 3

11IntroductionAgilent Desktop EasyEXPERT software makes every user a parametric test expert. The Microsoft® Windows®-based interface is familiar, even

Page 4

12Staircase Sweep with Pulsed Bias Measurement ModeThis mode forces swept voltage or current, and measures DC voltage or current. A second channel can

Page 5 - GNDU (Ground Unit)

13System RequirementsThe following are the minimum require-ment for executing Desktop EasyEXPERT.Operating System and Service PackMicrosoft Windows XP

Page 6 - Functions

Remove all doubtOur repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value o

Page 7 - Overlay Graph Comparison

2HardwareSpecifi cation ConditionThe “supplemental” information and “typical” entries in the following speci-fi cations are not warranted, but provide u

Page 8 - Repeating and Automating Test

3Agilent 4156C Semiconductor Parameter AnalyzerMPSMU (Medium Power SMU) Specifi cationsVoltage Range, Resolution, and Accuracy (MPSMU)Voltage Range Set

Page 9

4Agilent 41501B SMU and Pulse Generator ExpanderHPSMU (High Power SMU) Specifi cationsVoltage Range, Resolution, and Accuracy (HPSMU)Voltage Range Set.

Page 10 - Accessory Specifi cations

5PGU (Pulse Generator Unit) Specifi cationsModes: Pulse or constantAmplitude: 0 Vp-p to 40 Vp-pWindow: -40.0 V to +40.0 VMaximum current:±100 mA±200 mA

Page 11 - Functionality

6Capacitance Calculation Accuracy (Supplemental Data)Accuracy is derived from the current range, voltage range, capacitance measurement and leakage cu

Page 12

7Sampling (Time Domain) Measurement CharacteristicsDisplays the time sampled voltage/current data versus time.Max. sampling points: 10,001 (linear)Sam

Page 13 - System Requirements

8Automatic Analysis FunctionOn a graphics plot, the markers and lines can be automatically located using the auto analysis setup. Parameters can be au

Page 14 - Remove all doubt

9TriggerInput: External trigger input starts a sweep or sampling measurement or can be used as a trigger input for continuing an Instrument BASIC pro

Comments to this Manuals

No comments