Agilent Technologies 1912A Service Manual

Browse online or download Service Manual for Multimeters Agilent Technologies 1912A. Agilent Technologies 1912A Technical data User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 142
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - P-Series Power Meters

Keysight N1911A/1912A P-Series Power MetersService Guide

Page 2

viii N1911A/1912A P-Series Power Meters Service GuideOut position of bi-stable push control.

Page 3

6Repair Guide80 N1911A/1912A P-Series Power Meters Service GuideN1912-61025 Cable clamp34401-86020 Bumper kit34401-45021 HandleKeysight Part NumberDes

Page 4

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 81SundriesKeysight Part NumberDescription VisualN1911-61004N1912-80005Calibrator semi-r

Page 5 - Exclusive Remedies

6Repair Guide82 N1911A/1912A P-Series Power Meters Service GuideTools RequiredKeysight Part NumberDescription VisualN1911-61004N1912-80005• 3 unit of

Page 6 - Equipment Operation

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 83Required Torque Values for FastenersRequired tools and torque values for fasteners ar

Page 7 - General Safety Considerations

6Repair Guide84 N1911A/1912A P-Series Power Meters Service GuideDisassembly InstructionsThe guidelines in this section describe the disassembly of the

Page 8 - Safety Symbols

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 85• Remove the top clamshell (Figure 6-3): Disconnect the mains power connector from t

Page 9

6Repair Guide86 N1911A/1912A P-Series Power Meters Service Guide• Remove the PSU (Figure 6-5): Use the T10 Torx screwdriver bit to remove the 6 screws

Page 10

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 87• With reference to Figure 6-6 and Figure 6-7:• Lift and remove the cable clamp. • Di

Page 11 - Regulatory Information

6Repair Guide88 N1911A/1912A P-Series Power Meters Service Guide• Disconnect the front panel cable (Figure 6-9): Depress both sides of theconnector h

Page 12 - Regulatory Markings

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 89• Remove the front panel (Figure 6-11): Carefully lift and remove the front panel ass

Page 13 - In This Guide

N1911A/1912A P-Series Power Meters Service Guide ixEnvironmental ConditionsThis instrument is designed for indoor use. The table below shows the gener

Page 14

6Repair Guide90 N1911A/1912A P-Series Power Meters Service Guide• Remove DAP/PPMC (Figure 6-12): Use the T8 Torx screwdriver bit to remove the screws

Page 15 - Contents

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 91• Remove the rear panel (Figure 6-14): Use the N1912-61807 special tooling kit to re

Page 16 - 6 Repair Guide

6Repair Guide92 N1911A/1912A P-Series Power Meters Service GuideReassembly InstructionsInstructions Visual• The reassembly process is simply the rever

Page 17

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 93• Sensor RF connections (Figure 6-17): E – Channel A(+), Black cable F – Channel

Page 18

6Repair Guide94 N1911A/1912A P-Series Power Meters Service Guide• PSU cable routing (Figure 6-19): Ensure the PSU cables are positioned such that the

Page 19 - List of Figures

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 95Disassembly vs Part ReplacementDisassembly of Replacement Part InstructionsMain board

Page 20

6Repair Guide96 N1911A/1912A P-Series Power Meters Service GuideDAP assembly, Channel A (Rear connectors option)• Remove handle, bumpers, and top clam

Page 21 - Specifications

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 97Sensor flex assembly (rear connectors option)• Remove handle and front/rear Bumpers•

Page 22

6Repair Guide98 N1911A/1912A P-Series Power Meters Service GuideFront Panel Disassembly InstructionsInstructions VisualIMPORTANT NOTE:• The front pane

Page 23 - Specification Definitions

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 99Step 2:• Use ODU socket to remove N1912-61806 sensor flex assembly• Release the tab h

Page 24 - Conditions

x N1911A/1912A P-Series Power Meters Service GuideRegulatory MarkingsThe CE mark is a registered trademarkof the European Community. This CEmark shows

Page 25 - General Features

6Repair Guide100 N1911A/1912A P-Series Power Meters Service GuideStep 5:• Remove the 4 screws that attach the display to the display support molding,

Page 26

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 101Step 7:• Remove the 3 screws that attach the calibrator assembly to the display supp

Page 27

6Repair Guide102 N1911A/1912A P-Series Power Meters Service GuideStep 10:• Release the metal tabs holding the EMI screen to the front panel sub-frame,

Page 28

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 103Front Panel Reassembly InstructionsInstructions VisualStep 1:• Insert the keymat int

Page 29

6Repair Guide104 N1911A/1912A P-Series Power Meters Service GuideStep 4:• Overlay the EMI screen onto the key flex circuit, ensuring that all of the m

Page 30 - Maximum SWR

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 105Step 6:• Attach the inverter board to the display interface board using the 2 screws

Page 31 - Physical characteristics

6Repair Guide106 N1911A/1912A P-Series Power Meters Service GuideStep 10:• Fit the split washer to the calibrator assemblyStep 11:• Attach the display

Page 32 - 1 mW Power Reference

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 107Step 13:• Attach the front panel sub-frame to the display support moulding, ensuring

Page 33 - Measurement Speed

6Repair Guide108 N1911A/1912A P-Series Power Meters Service GuideAdditional Repair NotesReplacing A Sensor Flex Assembly:• The Sensor flex assembly is

Page 34 - Physical Characteristics

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 109Main Board vs. Rear Panel AssemblyInstructions Visual• Due to a difference in the co

Page 35 - < 5 % < 5 %

N1911A/1912A P-Series Power Meters Service Guide xiIn This Guide ...1Specifications Chapter 1 lists the power meter’s specifications and describes how

Page 36 - Characteristics Peak Flatness

6Repair Guide110 N1911A/1912A P-Series Power Meters Service GuideReplacing the PPMC Assembly• The PPMC assembly is pre- programmed with N1912A firmwar

Page 37 - Noise and drift

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 111Replacing the Calibrator Semi-Rigid/Split FerriteInstructions Visual• Separate the t

Page 38 - 1 Specifications

6Repair Guide112 N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 39 - Performance Tests

N1911A/1912A P-Series Power MetersService Guide7Contacting Keysight TechnologiesIntroduction 114Before Calling Keysight Technologies 115Check the Basi

Page 40

7 Contacting Keysight Technologies114 N1911A/1912A P-Series Power Meters Service GuideIntroductionContacting Keysight TechnologiesThis section details

Page 41 - Complete Equipment List

Contacting Keysight Technologies 7N1911A/1912A P-Series Power Meters Service Guide 115Before Calling Keysight TechnologiesBefore calling Keysight Tech

Page 42 - 2Performance Tests

7 Contacting Keysight Technologies116 N1911A/1912A P-Series Power Meters Service GuideCheck the BasicsProblems can be solved by repeating what was bei

Page 43 - Equipment

Contacting Keysight Technologies 7N1911A/1912A P-Series Power Meters Service Guide 117Instrument Serial NumbersKeysight Technologies makes frequent im

Page 44 - Test Method

7 Contacting Keysight Technologies118 N1911A/1912A P-Series Power Meters Service GuideKeysight Sales and Service OfficesIn any correspondence or telep

Page 45

Contacting Keysight Technologies 7N1911A/1912A P-Series Power Meters Service Guide 119Returning Your Power Meter for ServiceUse the information in thi

Page 46

xii N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 47

7 Contacting Keysight Technologies120 N1911A/1912A P-Series Power Meters Service GuideUseful Web Pages• Main Product Pagewww.keysight.com/find/wideban

Page 48 - Time Base Frequency Accuracy

www.keysight.comContact usTo obtain service, warranty, or technical support assistance, contact us at the following phone numbers:United States:(tel)

Page 49

This information is subject to change without notice.© Keysight Technologies 2006 - 2014Edition 14, November 2014*N1912-90015*N1912-90015www.keysight.

Page 50 - Zero Set (Average Path)

N1911A/1912A P-Series Power Meters Service Guide xiiiContentsNotices iiCertification iiiLimitation of Warranty iiiExclusive Remedies iiiEquipment Ope

Page 51 - Zero Set (Peak Path)

xiv N1911A/1912A P-Series Power Meters Service GuideLinearity (Average Path) 32Linearity (Peak Path) 33Rise/Fall Time (Peak Path) 343 AdjustmentsIn

Page 52 - Linearity (Average Path)

N1911A/1912A P-Series Power Meters Service Guide xvFront Panel Disassembly Instructions 98Front Panel Reassembly Instructions 103Additional Repair No

Page 53 - Linearity (Peak Path)

xvi N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 54 - Rise/Fall Time (Peak Path)

N1911A/1912A P-Series Power Meters Service Guide xviiList of FiguresFigure 2-1 1 mW Power Reference Level Test Setup Connection Diagram 23Figure 2

Page 56

xviii N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 57 - Adjustments

N1911A/1912A P-Series Power MetersService Guide1SpecificationsIntroduction 2Specification Definitions 3General Features 5P-Series Power Meter and Sens

Page 58

1 Specifications2 N1911A/1912A P-Series Power Meters Service GuideIntroductionThis chapter details the power meter’s specifications and supplemental c

Page 59

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 3Specification DefinitionsThere are two types of product specifications:Warranted spe

Page 60 - 3 Adjustments

1 Specifications4 N1911A/1912A P-Series Power Meters Service GuideConditionsThe power meter and sensor will meet its specifications when:• stored for

Page 61 - Theory of Operation

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 5General Features1 Information contained in this document refers to operation with P-

Page 62 - PPMC Assembly

1 Specifications6 N1911A/1912A P-Series Power Meters Service GuideP-Series Power Meter and Sensor Key System Specifications and Characteristics3 Speci

Page 63 - Main Board Assembly

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 7Figure 1-1 Measured rise time percentage error versus signal under test rise timeAlt

Page 64 - DAP Assembly

1 Specifications8 N1911A/1912A P-Series Power Meters Service GuideP-Series Power Meter SpecificationsMeter uncertaintyInstrumentation linearity ± 0.8

Page 65 - Calibrator Assembly

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 96 Internal trigger latency is defined as the delay between the applied RF crossing t

Page 67 - PSU Assembly

1 Specifications10 N1911A/1912A P-Series Power Meters Service GuideP-Series Wideband Power Sensor SpecificationsThe P- Series wideband power sensors a

Page 68 - 4 Theory of Operation

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 11Sensor Calibration Uncertainty9Definition: Uncertainty resulting from non- lineari

Page 69 - Troubleshooting Guide

1 Specifications12 N1911A/1912A P-Series Power Meters Service Guide1 mW Power ReferenceRear panel inputs/outputsNOTEThe 1mW power reference is provide

Page 70

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 13Remote ProgrammingMeasurement SpeedRegulatory InformationInterfaceGPIB interface op

Page 71 - Power-Up Problems

1 Specifications14 N1911A/1912A P-Series Power Meters Service GuidePhysical CharacteristicsEnvironmental ConditionsDimensionsThe following dimensions

Page 72 - Instrument Self-Test

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 15System Specifications and CharacteristicsThe video bandwidth in the meter can be se

Page 73 - Troubleshooting Guide 5

1 Specifications16 N1911A/1912A P-Series Power Meters Service GuideCharacteristics Peak FlatnessThe peak flatness is the flatness of a peak- to- avera

Page 74 - Extended Self-Test

Specifications 1N1911A/1912A P-Series Power Meters Service Guide 17Noise and drift12 Within 1 hour after a zero, at a constant temperature, after 24 h

Page 75 - Performance Test

1 Specifications18 N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 76 - Possible Faults

N1911A/1912A P-Series Power MetersService Guide2Performance TestsIntroduction 20Complete Equipment List 211 mW Power Reference Level Test 23Output Sta

Page 77

ii N1911A/1912A P-Series Power Meters Service GuideNotices© Keysight Technologies 2006 - 2014No part of this manual may be reproduced in any form or b

Page 78 - Additional Diagnostic Tests

2Performance Tests20 N1911A/1912A P-Series Power Meters Service GuideIntroductionThe performance tests described in this chapter test the power meter’

Page 79 - Repair Guide

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 21Complete Equipment ListInstrument Critical Specifications Recommended Keysight M

Page 80

2Performance Tests22 N1911A/1912A P-Series Power Meters Service GuidePulse/Data generator81131A Output modules required81130APower splitter required F

Page 81 - Replaceable Parts

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 231 mW Power Reference Level TestDescriptionThe 1 mW power reference is used for t

Page 82 - Customization Details

2Performance Tests24 N1911A/1912A P-Series Power Meters Service GuideTest Method1 Enter the recorded measurement uncertainty of the E4419B or E4417A 1

Page 83 - Customization Parts

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 25Output Standing Wave Ratio (SWR) TestDescriptionConnector mismatch is the larges

Page 84 - N1912-00026

2Performance Tests26 N1911A/1912A P-Series Power Meters Service GuideFigure 2-2 System Calibration Connection DiagramFigure 2-3 Output SWR Test Setup

Page 85

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 27Test Method1 Obtain the S11 parameter of the calibration system.2 Connect the me

Page 86

2Performance Tests28 N1911A/1912A P-Series Power Meters Service GuideTime Base Frequency AccuracyDescriptionThe accuracy of the 100 MHz sample clock d

Page 87

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 29Test Method1 Enable the path that routes the time base signal to the trigger out

Page 88

N1911A/1912A P-Series Power Meters Service Guide iiiCertificationKeysight Technologies certifies that this product met its published specifications at

Page 89

2Performance Tests30 N1911A/1912A P-Series Power Meters Service GuideZero Set (Average Path)DescriptionZero set is defined as the amount of residual o

Page 90

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 31Zero Set (Peak Path)DescriptionZero set is defined as the amount of residual off

Page 91

2Performance Tests32 N1911A/1912A P-Series Power Meters Service GuideLinearity (Average Path)DescriptionLinearity over the full input voltage range of

Page 92 - PSU (Power Supply Unit)

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 33Linearity (Peak Path)DescriptionLinearity over the full input voltage range of t

Page 93 - Rear Panel Assembly

2Performance Tests34 N1911A/1912A P-Series Power Meters Service GuideRise/Fall Time (Peak Path)DescriptionThe rise and fall time performance of the in

Page 94

Performance Tests 2N1911A/1912A P-Series Power Meters Service Guide 35Test Method1 Capture a train of 10 pulses with very fast rise/fall times.2 Combi

Page 95

2Performance Tests36 N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 96 - Additional Spare Parts

N1911A/1912A P-Series Power MetersService Guide3AdjustmentsIntroduction 38Power Reference Level Adjustment 39This chapter contains checks and adjustme

Page 97

3 Adjustments38 N1911A/1912A P-Series Power Meters Service GuideIntroductionAttempts to correct the power reference level if the performance test has

Page 98 - Improved calibrator assembly

Adjustments 3N1911A/1912A P-Series Power Meters Service Guide 39Power Reference Level AdjustmentEquipment• As per the test equipment list for the powe

Page 99 - Outer Housing Components

iv N1911A/1912A P-Series Power Meters Service GuideEquipment OperationWarnings and CautionsThis guide uses warnings and cautions to denote hazards.Per

Page 100 - Description Visual

3 Adjustments40 N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 101 - Sundries

N1911A/1912A P-Series Power MetersService Guide4Theory of OperationPPMC Assembly 42Main Board Assembly 43DAP Assembly 44Calibrator Assembly 45Front Pa

Page 102 - Tools Required

4 Theory of Operation42 N1911A/1912A P-Series Power Meters Service GuidePPMC AssemblyPurpose• Provides the main processor and memory for the power met

Page 103 - Repair Guide 6

Theory of Operation 4N1911A/1912A P-Series Power Meters Service Guide 43Main Board AssemblyPurpose• Provides the average measurement path(s)• Provides

Page 104 - Disassembly Instructions

4 Theory of Operation44 N1911A/1912A P-Series Power Meters Service GuideDAP AssemblyPurpose• Provides data acquisition and processing for the peak mea

Page 105 - Instructions Visual

Theory of Operation 4N1911A/1912A P-Series Power Meters Service Guide 45Calibrator AssemblyPurpose• Provides a 1 mW (0 dBm) power reference level at 5

Page 106

4 Theory of Operation46 N1911A/1912A P-Series Power Meters Service GuideFront Panel AssemblyPurpose• Provides a keyboard as the manual user interface•

Page 107 - Figure 6-8

Theory of Operation 4N1911A/1912A P-Series Power Meters Service Guide 47PSU AssemblyPurpose:• Provides various DC power suppliesInputs:• 100 Vac~ 240

Page 108 - Figure 6-10

4 Theory of Operation48 N1911A/1912A P-Series Power Meters Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Page 109 - Figure 6-11

N1911A/1912A P-Series Power MetersService Guide5Troubleshooting GuideIntroduction 50Power-Up Problems 51Instrument Self-Test 52Extended Self-Test 54Pe

Page 110 - Figure 6-13

N1911A/1912A P-Series Power Meters Service Guide vGeneral Safety ConsiderationsThe following general safety precautions must be observed during all ph

Page 111 - Figure 6-14

5 Troubleshooting Guide50 N1911A/1912A P-Series Power Meters Service GuideIntroductionThis chapter enables qualified service personnel to diagnose sus

Page 112 - Reassembly Instructions

Troubleshooting Guide 5N1911A/1912A P-Series Power Meters Service Guide 51Power-Up ProblemsBasic External Checks• Check the mains power source is live

Page 113 - Figure 6-18

5 Troubleshooting Guide52 N1911A/1912A P-Series Power Meters Service GuideInstrument Self-TestInstrument Purpose Debug Tips Possible FaultsTest point

Page 114 - Figure 6-19

Troubleshooting Guide 5N1911A/1912A P-Series Power Meters Service Guide 53ChB peak path Verifies that the peak path of channel B is working(Note: This

Page 115

5 Troubleshooting Guide54 N1911A/1912A P-Series Power Meters Service GuideExtended Self-TestInstrument Purpose Debug Tips Possible FaultsKeyboard Veri

Page 116

Troubleshooting Guide 5N1911A/1912A P-Series Power Meters Service Guide 55Performance TestType of Failures Debug Tips Possible Faults1 mW power refere

Page 117

5 Troubleshooting Guide56 N1911A/1912A P-Series Power Meters Service GuidePower Reference Level Adjustment ProblemsPossible Faults• Calibrator assembl

Page 118

Troubleshooting Guide 5N1911A/1912A P-Series Power Meters Service Guide 57Communication Interface FailuresType of Communication Debug Tips Possible Fa

Page 119

5 Troubleshooting Guide58 N1911A/1912A P-Series Power Meters Service GuideAdditional Diagnostic TestsType of Functionality Reason Recommended Test Met

Page 120

N1911A/1912A P-Series Power MetersService Guide6Repair GuideIntroduction 60Replaceable Parts 61 Front Panel Assembly 61 Main Board Assembly 69

Page 121

vi N1911A/1912A P-Series Power Meters Service GuideSafety SymbolsThe following symbol on the instrument and in the documentation indicates precautions

Page 122

6Repair Guide60 N1911A/1912A P-Series Power Meters Service GuideIntroductionThis chapter contains details of some of the higher level components and a

Page 123

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 61Replaceable PartsFront Panel AssemblyMain AssemblyThe standard P- Series power meter

Page 124

6Repair Guide62 N1911A/1912A P-Series Power Meters Service GuideCustomization DetailsThe standard P- Series power meters have the input sensor connect

Page 125

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 63Customization PartsKeysight Part NumberDescription VisualN1912-61806Sensor flex assem

Page 126

6Repair Guide64 N1911A/1912A P-Series Power Meters Service GuideKeysight Part NumberDescription VisualN1912-00025N1912-00026N1912-00027Blank front pan

Page 127

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 65Replaceable PartsKeysight Part NumberDescription VisualN1912-40003 Front panel sub-fr

Page 128 - Additional Repair Notes

6Repair Guide66 N1911A/1912A P-Series Power Meters Service GuideN1912-66502 Key flex circuitNote:• This key flex circuit is used on all variants of th

Page 129 - Figure 6-22

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 672090-0825 DisplayNote:• This display is used on all variants of the front panel assem

Page 130 - Replacing the PPMC Assembly

6Repair Guide68 N1911A/1912A P-Series Power Meters Service Guide0950-4111 Inverter boardNote:• This inverter board is used on all variants of the fron

Page 131 - Figure 6-24

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 69Main Board AssemblyKeysight Part NumberDescription VisualN1911-61801(non-video output

Page 132

N1911A/1912A P-Series Power Meters Service Guide viiPROTECTIVE CONDUCTOR TERMINAL.Frame or chasis TERMINAL.Equipotentiality.On (Supply).Off (Supply).E

Page 133

6Repair Guide70 N1911A/1912A P-Series Power Meters Service GuidePPMC (Processor PCI Mezzanine) AssemblyKeysight Part NumberDescription VisualN1911-665

Page 134 - Introduction

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 71DAP (Digital Acquisition & Processing) AssemblyKeysight Part NumberDescription Vi

Page 135

6Repair Guide72 N1911A/1912A P-Series Power Meters Service GuidePSU (Power Supply Unit)Keysight Part NumberDescription VisualN1912-61808 Power supply

Page 136 - Check the Basics

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 73Rear Panel AssemblyMain AssemblyCustomization DetailsThe standard P- Series power met

Page 137 - Instrument Serial Numbers

6Repair Guide74 N1911A/1912A P-Series Power Meters Service GuideCustomization PartsN1911A (front connectors option)1 unit of sensor flex assembly1 uni

Page 138

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 75N1912-61806 Sensor flex assemblyE4418-61015 Recorder output cable6960-00816960-002469

Page 139

6Repair Guide76 N1911A/1912A P-Series Power Meters Service GuideAdditional Spare PartsKeysight Part NumberDescription VisualN1912-61036 Line moduleN4

Page 140 - Useful Web Pages

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 77Calibrator AssemblyKeysight Part NumberDescription VisualN1911-61001 Calibrator assem

Page 141 - Contact us

6Repair Guide78 N1911A/1912A P-Series Power Meters Service GuideImproved calibrator assemblyThe N1911A/N1912A calibrator assembly is improved with a n

Page 142 - *N1912-90015*

Repair Guide 6N1911A/1912A P-Series Power Meters Service Guide 79Outer Housing ComponentsKeysight Part NumberDescription Visual5041-7717 Clamshell (t

Comments to this Manuals