Agilent Technologies 437B User's Guide Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 20
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
14
5. Known Problems
The following cases are known problems when
using the E4419A or E4419B dual channel power
meter in 438A language compatibility mode.
1. With the E4419A dual channel power meter in
438A language mode, the data ready bit in the
status register will not be set when in measure-
ment mode B (measurement mode A is not
affected).
Symptom:
10 OUTPUT @Pm; “BE;BP;CS” !Measurement mode B
20 OUTPUT @Pm; “TR1” ! Or TR2
30 IF (BIT (SPOLL(@Pm), 0) ! Wait for data ready
THEN GOTO 30
40 ENTER @Pm;Pwr
Possible work-around:
10 OUTPUT @Pm; “BE;BP;CS”
20 OUTPUT @Pm; “TR1”
30 ENTER @Pm;Pwr
Having triggered the meter, the same effect can be
obtained by addressing the meter to talk and wait-
ing for the response. The result will not be returned
until the measurement cycle has completed.
The data ready bit is intended to allow the con-
troller to do other tasks, while periodically check-
ing the state of this bit. In TR2 mode with 512
averages, it could take 30 seconds before the result
is available.
This bug is corrected on the E4419B dual channel
power meter (firmware revision A.02.03 or higher).
2. With the E4419B dual channel power meter in
438A language mode, measurement error code
36 could be returned for either of the following
reasons:
• E-series sensor error on channel A (sensor
data suspect)
• The “must cal/zero” feature is enabled and
channel B requires calibrating or zeroing.