• Cellular and PCS frequency coverage in one box• Analog and TDMA test capabilities• Built-in average power meter with ±7.5% accuracy• Dedicated, one-
2Agilent 8935 Series E6381A TDMA Base Station Test SetSpecifications Specifications describe the instrument’s warranted performance and arevalid over
3OutputRF IN/OUT ConnectorLevel Range: –137 dBm to –40 dBm into 50 ΩLevel Accuracy: ±1.0 dB (level >–127 dBm); if CDPD analyzer is also connected a
4FM Accuracy (1 kHz rate):<10 kHz dev: ±3.5% of setting ±50 Hz,>10 kHz dev: ±3.5% of setting ±500 HzFM Distortion (THD + noise, in a 0.3 to 3 kH
5AM MeasurementFrequency Range: 10 MHz to 1 GHz (usable to 400 kHz), 1.7 GHzto 2.0 GHzDepth: 0 to 95%Input Level Range (levels in PEP):–6 dBm to +52 d
6DC Voltage MeasurementVoltage Range: 100 mV to 42 VAccuracy: ±1.0% of reading ±45 mVDC Offset: ±25 mVSupplemental CharacteristicsResolution: 1 mVDist
7Leakage: At RF generator output levels <–40 dBm, typical radiated leakage is <10 µV induced in a resonant dipole antenna25 mm (1 inch) away fro
Standard User Memory, RAMApproximately 928 Kbytes of RAM are available for nonvolatilesave/recall of settings. This will typically allow >500 sets
Comments to this Manuals