Agilent Technologies N6705 Specifications

Browse online or download Specifications for Multimeters Agilent Technologies N6705. Agilent Technologies N6705 Specifications User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 32
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
1
Agilent DC Power Analyzer
Models: N6705A, N6715A, N6705B, N6715B,
N6731B-36B, N6741B-46B, N6751-54A,
N6761A-62A, N6773A-76A, N6781A-82A
Technical Overview
Ideal for R&D testing and design validation
Sources and measures DC voltage and current into your device under test
Combines 1 to 4 DC power supplies, DMM, oscilloscope, arbitrary waveform
generator, and datalogger in 1 integrated instrument
Saves time – no programming required
Eliminates need to gather and configure multiple instruments
Flexible, modular system: Can mix and match DC source power levels and
measurement performance levels to optimize investment
Connect via GPIB, LAN, or USB
Fully compliant to LXI Class C specification
For Power Solutions in ATE –
See back cover
See insights into power
consumption never seen
before with Agilent’s new
2-quadrant source/measure units
(SMUs) and 14585A control and
analysis software
For details visit
www.agilent.com/find/N6700
www.agilent.com/find/14585
Page view 0
1 2 3 4 5 6 ... 31 32

Summary of Contents

Page 1 - Agilent DC Power Analyzer

1Agilent DC Power Analyzer Models: N6705A, N6715A, N6705B, N6715B, N6731B-36B, N6741B-46B, N6751-54A, N6761A-62A, N6773A-76A, N6781A-82ATechnical Over

Page 2 - Under Time Pressure

ideally suited for semiconductor and passive device testing, or where a precisely controlled output and highly accurate, precise measure-ments are nee

Page 3 - DC Power Supply Outputs

11Output disconnect and polarity reversal relaysModules in the Agilent N6705 can be individually ordered with optional Output Disconnect Relays (optio

Page 4 - Voltmeter/Ammeter

Power management feature allows you to allocate N6705 mainframe powerOften, a DUT requires some high power DC sources and several very low power DC so

Page 5 - Oscilloscope

13N6750 SeriesFor applications where the power supply plays a critical role The Agilent N6750 Series of high-performance, auto ranging DC power module

Page 6 - Datalogger

14Agilent N6751A/N6752A, N6754A and N6761A/N6762A Performance SpecificationsUnless otherwise noted, specifications are warranted over the ambient temp

Page 7 - Arbitrary Waveform Generator

15Agilent N6751A/N6752A, N6754A and N6761A/N6762A Supplemental CharacteristicsSupplemental characteristics are not warranted but are descriptions of p

Page 8 - Additional Features

Agilent N6751A/N6752A, N6754A and N6761A/N6762A Supplemental Characteristics (Continued) N6751A / N6753A NOTE 4 N6754A N6761A / N6752A

Page 9

17Agilent N6731B – N6736B and N6741B – N6746B Performance SpecificationsUnless otherwise noted, specifications are warranted over the ambient temperat

Page 10 - USB port

N6731B/ N6732B/ N6733B/ N6734B/ N6735B/ N6736B/ N6741B N6742B N6743B N6744B N6745B N6746B Programming ranges Voltage 15 mV –

Page 11

19Agilent N6773A – N6776A Performance SpecificationsUnless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55

Page 12

Due to increasing time-to-market pressures, research and development engineers often find themselves on tight schedule to work through device under te

Page 13

Agilent N6773A – N6776A Supplemental CharacteristicsSupplemental characteristics are not warranted but are descriptions of performance determined eith

Page 14 - N6752A N6762A

21Agilent N6705 DC Power Analyzer Supplemental CharacteristicsSupplemental characteristics are not warranted but are descriptions of performance deter

Page 15

22Agilent N6705 DC Power Analyzer Supplemental Characteristics (Continued)Supplemental characteristics are not warranted but are descriptions of perf

Page 16 - (Continued)

23 N6735B N6745B 3 dB max THD at 3 dB 6 dB max THD at 6 dB frequency max frequency frequency max frequency 0.6 Vp-p 100 H

Page 17

24 N6773A 3 dB max THD at 3 dB 6 dB max THD at 6 dB frequency max frequency frequency max frequency 0.2 Vp-p 125 Hz 1.5% 210 H

Page 18

25 N6705A, N6705B Maximum total output power N6705A, N6705B 600 W when operating from 100 – 240 VAC input (= Sum of total module output pow

Page 19

N6705 A, N6705B Regulatory compliance EMC Complies with the European EMC directive for Class A test and measurement products. Complies

Page 20

27Outline Diagram16.9 mm0.663 in425.6 mm16.756 in23.5 mm0.923 in17.7 mm0.697 in177.0 mm6.968 in272.6 mm10.733 inAutoranging CharacteristicVoltageN6754

Page 21

28The DC power analyzer system is available 2 ways:(1) You can order an N6705B mainframe and various modules as separate products. (See steps below.)

Page 22 - (Continued)

29ModulesOrder 1 to 4 modules to be installed in each N6705 DC power analyzer main-frame. (To order modules as part of the N6715 build-to-order system

Page 23

3The Agilent N6705 DC power analyzer is a modular system that is tailorable to meet specific test needs. At the heart of the DC power analyzer is the

Page 24

30N6715B build-to-order systemsTo purchase a DC power analyzer system, order a N6715B. This model number is a build-to-order system that is shipped as

Page 25

31Modules as options to N6715BTo order 1 to 4 modules as options to an N6715B, specify its model number, followed by “–ATO.” For example, to order an

Page 26

Need a power solution for ATE?The new Agilent N6700 Modular Power Analyzer is small, flexible, and fast• Ideal for ATE systems in R&D, design val

Page 27 - Autoranging Characteristic

R&D engineers can invest in high-performance outputs where speed and accuracy are needed, or pur-chase basic performance outputs for simple DC pow

Page 28 - Ordering Information

5Each DC power module in the Agilent N6705 DC power analyzer has a fully integrated digitizer to capture the actual voltage versus time and current ve

Page 29

The Agilent N6705 DC power ana-lyzer can also function as a datalog-ger. Using the measurement capa-bility built into each DC power mod-ule, the N6705

Page 30

7Each DC power output on the Agilent N6705 DC power analyzer can be modulated by the module’s built-in arbitrary waveform genera-tor function. This pe

Page 31

Output sequencingEach DC power module can be individually set to turn on or to turn off with a delay. By adjusting the delay times and then commanding

Page 32 - Remove all doubt

9switching between 2-wire mode (local sensing) and 4-wire mode (remote sensing) is done via an internal relay inside the N6705 DC power analyzer, elim

Comments to this Manuals

No comments