Agilent Technologies InfiniiVision 7000B Series Service Manual

Browse online or download Service Manual for Multimeters Agilent Technologies InfiniiVision 7000B Series. Agilent Technologies InfiniiVision 7000B Series Technical data User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 118
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Agilent Technologies
Agilent InfiniiVision
7000 Series
Oscilloscopes
Service Guide
7000 service guide.book Page 1 Thursday, October 16, 2008 12:18 PM
Page view 0
1 2 3 4 5 6 ... 117 118

Summary of Contents

Page 1 - Oscilloscopes

Agilent Technologies Agilent InfiniiVision 7000 Series Oscilloscopes Service Guide7000 service guide.book Page 1 Thursday, October 16, 2008 12:18 P

Page 2 - Restricted Rights Legend

10 Service Guide FiguresFigure 24. Removal of inverter board 92Figure 25. Removing cable shield 93Figure 26. Removing keyboard cables 93Figure 27. Rem

Page 3 - In This Service Guide

100 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies To remove the power supply shield1 To remove power supply shield, locate and remove

Page 4

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 101 2 Once screws have been removed, carefully remove the power supply cover by lifting

Page 5 - Contents

102 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies To remove the power supply1 Disconnect all cables from power supply board.2 Locate

Page 6 - 6 Service Guide

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 103 To remove the AC input board1 Disconnect all cables on AC input board.2 Disconnect

Page 7 - Service Guide 7

104 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies 6 To remove the switch extender, gently pry open the extender using a flat head scr

Page 8 - 8 Service Guide

105Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 6Replaceable PartsThis chapter describes how to order replaceable a

Page 9 - Service Guide 9

106 7000 Series Oscilloscopes Service Guide 6 Replaceable Parts Ordering Replaceable PartsListed PartsTo order a part in the parts list, quote the Agi

Page 10 - 10 Service Guide

Replaceable Parts 67000 Series Oscilloscopes Service Guide 107 In order for Agilent Technologies to provide these advantages, please send a check or

Page 11 - Service Guide 11

108 7000 Series Oscilloscopes Service Guide 6 Replaceable Parts Exploded ViewsThe following exploded views provide a graphical representation of the o

Page 12 - 12 Service Guide

Replaceable Parts 67000 Series Oscilloscopes Service Guide 109 Figure 2 Exploded View 2 of 27000 service guide.book Page 109 Thursday, October 16,

Page 13 - Agilent Technologies

Service Guide 11 TablesTable 1. List of test equipment 23Table 2. Conventions 25Table 3. Materials required to construct the test connectors 26Table 4

Page 14 - Power Requirements

110 7000 Series Oscilloscopes Service Guide 6 Replaceable Parts Replaceable Parts ListThe information given for each part consists of the following:•

Page 15 - Measurement Category

Replaceable Parts 67000 Series Oscilloscopes Service Guide 111 H1 1400-0611 2 Clamp-FL-CA 1-WD PVCH2 54684-42202 1 Hole plug front panel (2 Channel o

Page 16

112 7000 Series Oscilloscopes Service Guide 6 Replaceable Parts MP1 54692-94302 1 ID Label 2 Ch MSO 500 MHzMP1 54693-94301 1 ID Label 4 Ch DSO 500 MHz

Page 17 - Environmental Conditions

Replaceable Parts 67000 Series Oscilloscopes Service Guide 113 MP17 01680-88001 1 LensMP20 5042-6805 2 Cap - Strap HandleMP21 54695-68711 1 Strap Han

Page 18 - Specifications

114 7000 Series Oscilloscopes Service Guide 6 Replaceable Parts *Optional item.W1 54695-61601 1 LVDS Display CableW2 54695-61602 1 Inverter CableW3 54

Page 19 - Contact us

Service Guide 115 IndexAAC input board removal, 103Agilentcontact information, 19assembliesexchange, 107replacing, 83Bbacklight inverter removal, 99ba

Page 20

116 Service GuideIndex removeAC input board, 103backlight inverter, 99cabinet, 85display assembly, 93front panel, 88handle, 86keyboard, 90power supply

Page 21 - Testing Performance

17000 Series Oscilloscopes Service Guide 117 Safety NoticesThis apparatus has been designed and tested in accordance with IEC Publication 1010, Safet

Page 22 - Overview

118 7000 Series Oscilloscopes Service Guide 1 Safety SymbolsInstruction manual symbol: the product is marked with this symbol when it is necessary for

Page 23 - List of Test Equipment

12 Service Guide Tables7000 service guide.book Page 12 Thursday, October 16, 2008 12:18 PM

Page 24 - 2 Testing Performance

13Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 1Characteristics and Specifications Power Requirements 14Measurement

Page 25 - Conventions

14 7000 Series Oscilloscopes Service Guide 1 Characteristics and Specifications Power RequirementsLine voltage, frequency, and power~Line 100-120 Vac

Page 26

Characteristics and Specifications 17000 Series Oscilloscopes Service Guide 15 Measurement CategoryMeasurement CategoryThe InfiniiVision 7000 Series

Page 27 - Testing Performance 2

16 7000 Series Oscilloscopes Service Guide 1 Characteristics and Specifications Transient Withstand CapabilityCAUTIONMaximum input voltage for analog

Page 28

Characteristics and Specifications 17000 Series Oscilloscopes Service Guide 17 Environmental ConditionsEnvironment Indoor use only.Ambient temperatur

Page 29 - Verifying Test Results

18 7000 Series Oscilloscopes Service Guide 1 Characteristics and Specifications SpecificationsPlease see the InfiniiVision 7000 Series Oscilloscopes D

Page 30

Characteristics and Specifications 17000 Series Oscilloscopes Service Guide 19 Contact usAmericasCanada (877) 894-4414Latin America 305 269 7500Unite

Page 31

7000 Series Oscilloscopes Service Guide Notices© Agilent Technologies, Inc. 2009No part of this manual may be reproduced in any form or by any means (

Page 32 - (on oscilloscope

20 7000 Series Oscilloscopes Service Guide 1 Characteristics and Specifications 7000 service guide.book Page 20 Thursday, October 16, 2008 12:18 PM

Page 33 - *1 LSB = 0.4% of full scale

21Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 2Testing PerformanceOverview 22List of Test Equipment 23To construct

Page 34

22 7000 Series Oscilloscopes Service Guide 2 Testing Performance OverviewTo completely test and troubleshoot MSO models, create and use the test conne

Page 35

Testing Performance 27000 Series Oscilloscopes Service Guide 23 List of Test EquipmentBelow is a list of test equipment and accessories required to p

Page 36

24 7000 Series Oscilloscopes Service Guide 2 Testing Performance Feedthrough 50Ω BNC (f) to BNC (m) Agilent 0960-0301* Required only for testing digit

Page 37

Testing Performance 27000 Series Oscilloscopes Service Guide 25 Fluke MET/CAL ProceduresFluke MET/CAL procedures are available for the 7000 Series os

Page 38 - To verify bandwidth

26 7000 Series Oscilloscopes Service Guide 2 Testing Performance To construct the test connector (for use with MSO models only)Agilent 7000 Series Osc

Page 39

Testing Performance 27000 Series Oscilloscopes Service Guide 27 Figure 1 Constructing the 8-by-2 ConnectorJumper (2)Signal Lead(from scope’s MSO cabl

Page 40

28 7000 Series Oscilloscopes Service Guide 2 Testing Performance To test digital channels (MSO models only)The acquisition system testing provides con

Page 41 - Ω×= 50

Testing Performance 27000 Series Oscilloscopes Service Guide 29 To verify digital channel threshold accuracy (MSO models only)This test verifies the

Page 42 - response(dB) = 20 log

7000 Series Oscilloscopes Service Guide 3 In This Service GuideThis book provides the service information for the Agilent 7000 Series Oscilloscopes.

Page 43

30 7000 Series Oscilloscopes Service Guide 2 Testing Performance 1 Turn on the test equipment and the oscilloscope. Let them warm up for 30 minutes b

Page 44 - To verify time scale accuracy

Testing Performance 27000 Series Oscilloscopes Service Guide 31 Figure 2 Setting Up Equipment for Digital Channel Threshold Accuracy Te s t4 Use a BN

Page 45

32 7000 Series Oscilloscopes Service Guide 2 Testing Performance oscilloscope to set the threshold test settings as shown in Table 5. 8 Do the followi

Page 46

Testing Performance 27000 Series Oscilloscopes Service Guide 33 Before proceeding to the next step, make sure that you have recorded the oscilloscope

Page 47 - To verify trigger sensitivity

34 7000 Series Oscilloscopes Service Guide 2 Testing Performance 1 Set up the oscilloscope. a Adjust the channel 1 position knob to place the baseline

Page 48

Testing Performance 27000 Series Oscilloscopes Service Guide 35 c Press the Acquire key. Then press the Averaging softkey and set #Avgs to 64.Wait a

Page 49 - Oscilloscope

36 7000 Series Oscilloscopes Service Guide 2 Testing Performance Figure 3 Setting up Equipment for Voltage Measurement Accuracy Te s t4 Adjust the out

Page 50

Testing Performance 27000 Series Oscilloscopes Service Guide 37 7 When you are finished checking all of the voltage values, disconnect the oscillosco

Page 51

38 7000 Series Oscilloscopes Service Guide 2 Testing Performance To verify bandwidthThis test checks the bandwidth of the oscilloscope. In this test y

Page 52

Testing Performance 27000 Series Oscilloscopes Service Guide 39 1 Connect the equipment (see Figure 5).a Use the N cable to connect the signal genera

Page 53

4 7000 Series Oscilloscopes Service Guide Digital Channels Because all of the oscilloscopes in the Agilent 7000 Series have analog channels, the anal

Page 54

40 7000 Series Oscilloscopes Service Guide 2 Testing Performance Figure 5 Setting Up Equipment for Bandwidth Verification Test2 Set up the power meter

Page 55

Testing Performance 27000 Series Oscilloscopes Service Guide 41 3 Set up the oscilloscope.a Press the Save/Recall key, then press the Default Setup s

Page 56

42 7000 Series Oscilloscopes Service Guide 2 Testing Performance • 1 GHz Models: 1 GHz• 500 MHz Models: 500 MHz• 350 MHz Models: 350 MHz• 100 MHz Mode

Page 57

Testing Performance 27000 Series Oscilloscopes Service Guide 43 15 The result from step 14 should be between +3.0 dB and -3.0 dB. Record the result i

Page 58

44 7000 Series Oscilloscopes Service Guide 2 Testing Performance To verify time scale accuracyThis test verifies the accuracy of the time scale. In th

Page 59 - Calibrating and Adjusting

Testing Performance 27000 Series Oscilloscopes Service Guide 45 4 Make the measurement.a Set oscilloscope horizontal sweep speed control to 1ms/div.b

Page 60 - 3 Calibrating and Adjusting

46 7000 Series Oscilloscopes Service Guide 2 Testing Performance To verify horizontal delta t accuracyThis test verifies the horizontal ∆t accuracy. I

Page 61 - User Calibration

Testing Performance 27000 Series Oscilloscopes Service Guide 47 4 Press the Quick Meas softkey, set the Source softkey to 1, then press Select and ch

Page 62 - To Channel 2

48 7000 Series Oscilloscopes Service Guide 2 Testing Performance Test limits for:• Internal trigger on all models:• < 10 mV/div: greater of 1 div o

Page 63

Testing Performance 27000 Series Oscilloscopes Service Guide 49 Test Internal Trigger Sensitivity (all models)1 Connect the equipment (see Figure 6).

Page 64 - User Cal Status

Service Guide 5 ContentsFigures 9Tables 111 Characteristics and Specifications 13Power Requirements 14Measurement Category 15Measurement Category 15Me

Page 65 - Troubleshooting

50 7000 Series Oscilloscopes Service Guide 2 Testing Performance 2 Verify the trigger sensitivity at maximum bandwidth. 1 GHz models: 1 GHz500 MHz mod

Page 66 - If there is no trace display

Testing Performance 27000 Series Oscilloscopes Service Guide 51 Test External Trigger Sensitivity (2-channel models)Verify the external trigger sensi

Page 67 - If you cannot see a channel

52 7000 Series Oscilloscopes Service Guide 2 Testing Performance 1 Connect the equipment (see Figure 7).a Use the N cable to connect the signal genera

Page 68

Testing Performance 27000 Series Oscilloscopes Service Guide 53 2 Set up the oscilloscope.a Press the Save/Recall key, then press the Default Setup s

Page 69

54 7000 Series Oscilloscopes Service Guide 2 Testing Performance 4 Verify the trigger sensitivity at maximum frequency for trigger range of 8.0 V.a Pr

Page 70 - To check out the oscilloscope

Testing Performance 27000 Series Oscilloscopes Service Guide 55 1 Connect the equipment (see Figure 8).a Use the N cable to connect the signal genera

Page 71 - AVOID INJURY

56 7000 Series Oscilloscopes Service Guide 2 Testing Performance 2 Set up the oscilloscope.a Press the Save/Recall key, then press the Default Setup s

Page 72

Testing Performance 27000 Series Oscilloscopes Service Guide 57 Agilent 7000 Series Oscilloscopes Performance Test RecordSerial No. _________________

Page 73 - Under compensated

58 7000 Series Oscilloscopes Service Guide 2 Testing Performance Horizontal ∆t AccuracyGenerator Setting Test Limits ResultsPeriod 100 µs 99.8 µs to 1

Page 74 - Troubleshooting Flowchart

59Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 3Calibrating and AdjustingThis chapter explains how to adjust the os

Page 75 - System board drawings

6 Service Guide ContentsWhat to Test 29Verifying Test Results 29To verify voltage measurement accuracy 33To verify bandwidth 38To verify time scale ac

Page 76 - 4 Troubleshooting

60 7000 Series Oscilloscopes Service Guide 3 Calibrating and Adjusting CAUTIONUSE EXTERNAL FAN TO REDUCE TEMPERATURE!When you must operate the oscillo

Page 77

Calibrating and Adjusting 37000 Series Oscilloscopes Service Guide 61 User CalibrationPerform user-calibration:• Each year or after 2000 hours of ope

Page 78

62 7000 Series Oscilloscopes Service Guide 3 Calibrating and Adjusting Use 50Ω RG58AU or equivalent BNC cables when performing User Cal.a For a 2-chan

Page 79 - Signal Normal/Typical Result

Calibrating and Adjusting 37000 Series Oscilloscopes Service Guide 63 BNC(f)-to-BNC(f) (barrel connector) to the tee as shown below.Figure 10 User Ca

Page 80 - To check the fan

64 7000 Series Oscilloscopes Service Guide 3 Calibrating and Adjusting User Cal StatusPressing the User Cal Status softkey displays the following summ

Page 81 - To verify default setup

65Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 4TroubleshootingThis chapter begins with suggestions for solving gen

Page 82

66 7000 Series Oscilloscopes Service Guide 4 Troubleshooting Solving General Problems with the OscilloscopeThis section describes how to solve general

Page 83 - Replacing Assemblies

Troubleshooting 47000 Series Oscilloscopes Service Guide 67 ✔ The sampling speed of the oscilloscope depends on the time/division setting. It may be

Page 84

68 7000 Series Oscilloscopes Service Guide 4 Troubleshooting Performing the checks listed here ensures that the signals from the circuit under test wi

Page 85 - Removing the Handle

Troubleshooting 47000 Series Oscilloscopes Service Guide 69 Troubleshooting the OscilloscopeThe service policy for the Agilent 7000 Series Oscillosco

Page 86 - To remove the Adjustable Legs

ContentsService Guide 7 To run the internal self-tests 81To verify default setup 815Replacing Assemblies83Removing the Handle 85To remove the Adjustab

Page 87 - To remove Cabinet

70 7000 Series Oscilloscopes Service Guide 4 Troubleshooting Equipment required for troubleshootingThe equipment listed in this table is required to t

Page 88

Troubleshooting 47000 Series Oscilloscopes Service Guide 71 3 Connect the power cord to the rear of the oscilloscope, then to a suitable ac voltage s

Page 89 - Replacing Assemblies 5

72 7000 Series Oscilloscopes Service Guide 4 Troubleshooting To verify basic oscilloscope operation1 Press the Save/Recall key on the front panel, the

Page 90

Troubleshooting 47000 Series Oscilloscopes Service Guide 73 To compensate the analog probesYou should compensate your analog probes to match their ch

Page 91

74 7000 Series Oscilloscopes Service Guide 4 Troubleshooting Troubleshooting FlowchartThe following flowchart describes how to troubleshoot the oscill

Page 92

Troubleshooting 47000 Series Oscilloscopes Service Guide 75 System board drawingsRefer to these two drawings to locate test points on the system boar

Page 93

76 7000 Series Oscilloscopes Service Guide 4 Troubleshooting Figure 13 System Board Test Points - Bottom Side66U3202U33007000 service guide.bo

Page 94

Troubleshooting 47000 Series Oscilloscopes Service Guide 77 To check the 7000A Series oscilloscope power supply1 Disconnect the power cord from the o

Page 95

78 7000 Series Oscilloscopes Service Guide 4 Troubleshooting To check the 7000A Series system board1 Remove the handle, tilt legs, and cabinet.2 Check

Page 96

Troubleshooting 47000 Series Oscilloscopes Service Guide 79 6 If all cables are properly connected and none of the previous tests confirm a failure o

Page 97

8 Service Guide Contents7000 service guide.book Page 8 Thursday, October 16, 2008 12:18 PM

Page 98

80 7000 Series Oscilloscopes Service Guide 4 Troubleshooting To check the fanThe fan speed is controlled by a circuit on the system board. 1 If the fa

Page 99 - To remove the Fan assembly

Troubleshooting 47000 Series Oscilloscopes Service Guide 81 To run the internal self-testsSelf Test performs a series of internal procedures to verif

Page 100

82 7000 Series Oscilloscopes Service Guide 4 Troubleshooting • Trigger - Edge trigger, Auto sweep mode, 0 V level, channel 1 source, dc coupling, risi

Page 101

83Agilent TechnologiesAgilent InfiniiVision 7000 Series OscilloscopeService Guide 5Replacing AssembliesThis chapter describes how to remove assemblies

Page 102 - To remove the power supply

84 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies See how the Oscilloscope Parts Fit TogetherAn exploded view of the oscilloscope is i

Page 103 - To remove the AC input board

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 85 Removing the HandleThe Strap Handle must be removed prior to removing the cabinet. T

Page 104 - 5 Replacing Assemblies

86 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies To remove the Adjustable LegsAdjustable legs must be removed prior to removing the c

Page 105 - Replaceable Parts

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 87 To remove CabinetRemoving the cabinet allows access to the rear deck, fan assembly,

Page 106 - Ordering Replaceable Parts

88 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies To remove the Rear Deck AssemblyRemoving the rear deck allows access to the front de

Page 107 - Exchange Assemblies

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 89 4 Disconnect power harness and fan cable from acquisition board. Note cable location

Page 108 - Exploded Views

Service Guide 9 FiguresFigure 1. Constructing the 8-by-2 Connector 27Figure 2. Setting Up Equipment for Digital Channel Threshold Accuracy Test 31Figu

Page 109 - Figure 2 Exploded View 2 of 2

90 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies To remove Acquisition and Inverter assembliesThe following illustrates how to remove

Page 110 - Replaceable Parts List

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 91 3 Using a TORX T10 driver locate and remove 7 mounting screws.Figure 22 TORX T10 mou

Page 111

92 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies 4 Carefully unlock the two alignment pins. Carefully lift acquisition board off fron

Page 112 - 6 Replaceable Parts

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 93 To remove the Front Panel assembly1 Remove cable shield from front deck. Carefully s

Page 113

94 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies 3 Carefully remove knobs from front panel. They may be gently pried loose with a fla

Page 114 - *Optional item

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 95 5 To remove the Softkey board printed circuit board from the bezel, insure cable has

Page 115 - Service Guide 115

96 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies 6 To remove the front panel assembly carefully release front panel clips that secure

Page 116 - 116 Service Guide

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 97 7 To separate the front panel control board from the front panel carefully pull back

Page 117 - Safety Notices

98 7000 Series Oscilloscopes Service Guide 5 Replacing Assemblies 9 Using a TORX T10 remove the four screws that secure the display to the front deck.

Page 118 - Safety Symbols

Replacing Assemblies 57000 Series Oscilloscopes Service Guide 99 To remove the Fan assembly1 Insure that fan power cable has been disconnected from A