This document describes the performance and features of
Agilent’s 50 and 75 ohm 8712E series RF vector network analyzers:
8712ET transmission/reflection vector network analyzer,
300 kHz to 1.3 GHz
8712ES S-parameter vector network analyzer,
300 kHz to 1.3 GHz
8714ET transmission/reflection vector network analyzer,
300 kHz to 3.0 GHz
8714ES S-parameter vector network analyzer,
300 kHz to 3.0 GHz
For more information about these analyzers, please read the
following documents:
8712E Series Brochure: 5967-6316E
8712E Series Configuration Guide: 5967-6315E
8712ET and 8712ES 300 kHz to 1.3 GHz
8714ET and 8714ES 300 kHz to 3.0 GHz
PORT 1
PORT 2
Agilent Technologies
8712E Series RF Vector
Network Analyzers
Technical Specifications
Discontinued Product Information
— For Support Reference Only —
Information herein, may refer to products/services no longer supported.
We regret any inconvenience caused by obsolete information. For the
latest information on Agilent’s test and measurement products go to:
www.agilent.com/find/products
In the US, call Agilent Technologies at 1-800-829-4444
(any weekday between 8am–5pm in any U.S. time zone)
World-wide Agilent sales office contact information is available at:
www.agilent.com/find/contactus
Page view 0
1 2 3 4 5 6 ... 31 32

Summary of Contents

Page 1 - Agilent Technologies

This document describes the performance and features of Agilent’s 50 and 75 ohm 8712E series RF vector network analyzers:• 8712ET transmission/reflect

Page 2 - Introduction

108712ET/8714ET 85032B/E (Type-N, 50 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationa(in dB)300 kHz 1.3 GHz to 1.3 GHz to 3 GHzDirectivity 50 47So

Page 3 - 8712ES/8714ES

118712ET/8714ET with Attenuator Option 1E1a85032B/E (Type-N, 50 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationb(in dB)300 kHz 1.3 GHz to 1.3 GHz

Page 4

128712ES/8714ES with Option 1ECa85036B/E (Type-N, 75 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationb(in dB)300 kHz 1.3 GHz to 1.3 GHz to 3 GHzDir

Page 5

138712ET/8714ET with Option 1ECa(without Attenuator) 85036B/E (Type-N, 75 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationb(in dB)300 kHz 1.3 GHz

Page 6

148712ET/8714ET with Options 1EC and 1E1a85036B/E (Type-N, 75 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationb(in dB)300 kHz 1.3 GHz to 1.3 GHz to

Page 7

158712ET/ES and 8714ET/ES (Type-N, 50 Ω) 8712ET/8714ET8712ET/8714ETDescription 8712ES/8714ES with Attenuator(without Attenuator)Option 1E1a300 kHz 1

Page 8

168712ET/ES and 8714ET/ES Test Port OutputDescription Specification (in dB) Supplemental InformationFrequencyRange:8712ET/ES 300 kHz to 1.3 GHz8714ET/

Page 9

17Test Port Output, continued8712ET/ES and 8714ET/ES Test Port OutputDescription Specification (in dB) Supplemental InformationOutput Power (continued

Page 10 - 8712ET/8714ET

188712ET/ES and 8714ET/ES Test Port InputDescription Specification Supplemental InformationFrequency Range8712ET/ESNarrowband 300 kHz to 1.3 GHzBroadb

Page 11

8712ET/ES and 8714ET/ES Test Port InputSpecification Typical(in dBm) (in dBm)Description System Bandwidths:Fine Fine Med Wide Wide(15 Hz) (15 Hz) (400

Page 12 - 8712ES/8714ES with Option 1EC

2All specifications and characteristics apply over a 25° C ±5° C range (unlessotherwise stated) and 60 minutes after the instrument has been turned on

Page 13 - (without Attenuator)

208712ET/ES and 8714ET/ES Test Port InputSpecification Typical(in dB) (in dB)Description System Bandwidths:Fine Fine Med Wide Wide(15 Hz) (15 Hz) (400

Page 14

21Dynamic Accuracy (Specification)a,b8712ET/8714ET:Magnitude PhaseDynamic Accuracy (Specification)a,b8712ES/8714ES:Magnitude PhaseTest Port Input, con

Page 15

22Power Accuracy (Characteristic)a8712ET/8714ET: 8712ES/8714ES:Magnitude Magnitude0.1110-50-40-30-20-10010Test Port Power (dBm)Accuracy (dB)75 Ω

Page 16 - Test Port Output

23General Information8712ET/ES and 8714ET/ES General InformationDescription Specification Supplemental InformationDisplay RangeMagnitude 200 dB (at 20

Page 17 - Test Port Output, continued

248712ET/ES and 8714ET/ES General InformationDescription Specification Supplemental InformationRear PanelAuxiliary Input:Connector Female BNCImpedance

Page 18 - Test Port Input

258712ET/ES and 8714ET/ES General InformationMeasurement Speed ConditionsaTypicalCal Number Measurement Number Frequency Cycle Recall Data Measurement

Page 19 - Test Port Input, continued

26General Information, continuedData transfer timesThe tables below show the various data transfer speeds that can be expected using different data fo

Page 20

27Block diagramsSimplified Block Diagram for the 8712ET and 8714ETSimplified Block Diagram for the 8712ES and 8714ESREAR PANELExternal DetectorsInput

Page 21

28MeasurementNumber of display measurementsTwo measurement displays are available, with independentcontrol of display parameters including format type

Page 22

29Product features, continuedHTTP The instrument's built-in web page can be accessed withany standard web browser using http (hypertext transferp

Page 23 - General Information

3System Performance, 2-Port Calibration (7-mm, 50 Ω)8712ES/8714ES 85031B (7-mm, 50 Ω) Cal Kit, User 2-Port CalibrationDescriptionSpecificationa(in dB

Page 24

30Programming with IBASICAs a standard feature, all 8712ET/ES and 8714ET/ES net-work analyzers come with the Instrument BASIC programming language (IB

Page 25

31Key options75 ohms (Option 1EC) Provides 75 ohm system impedance.Step attenuator (Option 1E1) Adds a built-in 60 dB step attenuator totransmission/r

Page 26 - Data transfer times

www.agilent.comAgilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive

Page 27 - Block diagrams

4System Performance, 2-Port Calibration (Type-N, 50 Ω)8712ES/8714ES 85032B/E (Type-N, 50 Ω) Cal Kit, User 2-Port CalibrationDescriptionSpecificationa(

Page 28 - Product features

5System Performance, 2-Port Calibration (3.5 mm, 50 Ω)8712ES/8714ES 85033D (3.5 mm, 50 Ω) Cal Kit, User 2-Port CalibrationDescriptionSpecificationa(in

Page 29 - Product features, continued

6System Performance, 2-Port Calibration (7-16, 50 Ω)8712ES/8714ES85038A (7-16, 50 Ω) Cal Kit, User 2-Port CalibrationDescriptionSpecificationa(in dB)3

Page 30

7System Performance, 2-Port Calibration (Type-N, 75 Ω)8712ES/8714ES with Option 1ECa85036B/E (Type-N, 75 Ω) Cal Kit, User 2-Port CalibrationDescriptio

Page 31

8System Performance, 2-Port Calibration (Type-F, 75 Ω)8712ES/8714ES with Option 1ECa85039B (Type-F, 75 Ω) Cal Kit, User 2-Port CalibrationDescriptionS

Page 32 - Agilent Direct

98712ES/8714ES 85032B/E (Type-N, 50 Ω) Cal Kit, T/R CalibrationDescriptionSpecificationa(in dB)300 kHz 1.3 GHz to 1.3 GHz to 3 GHzDirectivity 50 47Sou