Agilent Technologies, Inc. DisplayPort Test Implementation Group 1 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 DisplayPor
Agilent Technologies, Inc. DisplayPort Test Implementation Group 10 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 100MHz NA 168mUI 100
Agilent Technologies, Inc. DisplayPort Test Implementation Group 11 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Note: The introducti
Agilent Technologies, Inc. DisplayPort Test Implementation Group 12 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 fine-tuned by using
Agilent Technologies, Inc. DisplayPort Test Implementation Group 13 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Verify ISI to be 220
Agilent Technologies, Inc. DisplayPort Test Implementation Group 14 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 5: J-BERT N4
Agilent Technologies, Inc. DisplayPort Test Implementation Group 15 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 9.2 RJ Calibration
Agilent Technologies, Inc. DisplayPort Test Implementation Group 16 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 c. In the oscillos
Agilent Technologies, Inc. DisplayPort Test Implementation Group 17 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 10: HBR2 RJ
Agilent Technologies, Inc. DisplayPort Test Implementation Group 18 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 c. Alternate betwe
Agilent Technologies, Inc. DisplayPort Test Implementation Group 19 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Tables 4-3 and 4-4 o
Agilent Technologies, Inc. DisplayPort Test Implementation Group 2 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 1! MODIFICATION RECORD
Agilent Technologies, Inc. DisplayPort Test Implementation Group 20 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 14: J-BERT N
Agilent Technologies, Inc. DisplayPort Test Implementation Group 21 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 9.4 Eye Height Cali
Agilent Technologies, Inc. DisplayPort Test Implementation Group 22 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 18: HBR2 cro
Agilent Technologies, Inc. DisplayPort Test Implementation Group 23 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 If the device suppor
Agilent Technologies, Inc. DisplayPort Test Implementation Group 24 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 21: J-BERT N
Agilent Technologies, Inc. DisplayPort Test Implementation Group 25 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Note: Capture a mini
Agilent Technologies, Inc. DisplayPort Test Implementation Group 26 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Note: Crosstalk is
Agilent Technologies, Inc. DisplayPort Test Implementation Group 27 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 10 Testing the DUT
Agilent Technologies, Inc. DisplayPort Test Implementation Group 28 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 A. Appendix: Agilen
Agilent Technologies, Inc. DisplayPort Test Implementation Group 29 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 B. Appendix: Agilen
Agilent Technologies, Inc. DisplayPort Test Implementation Group 3 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 TEST STATION CONFIGURA
Agilent Technologies, Inc. DisplayPort Test Implementation Group 30 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 28: Differen
Agilent Technologies, Inc. DisplayPort Test Implementation Group 31 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 29: HBR2 Equ
Agilent Technologies, Inc. DisplayPort Test Implementation Group 32 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 30: HBR2 Dif
Agilent Technologies, Inc. DisplayPort Test Implementation Group 33 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 c. Press Next. d.
Agilent Technologies, Inc. DisplayPort Test Implementation Group 34 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 ii. For HBR: • Clo
Agilent Technologies, Inc. DisplayPort Test Implementation Group 35 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 k. Press Close to
Agilent Technologies, Inc. DisplayPort Test Implementation Group 36 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 new waveform Press C
Agilent Technologies, Inc. DisplayPort Test Implementation Group 37 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 f. Click Next. g.
Agilent Technologies, Inc. DisplayPort Test Implementation Group 38 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 34: HBR2 Dif
Agilent Technologies, Inc. DisplayPort Test Implementation Group 39 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 35: HBR2 Ran
Agilent Technologies, Inc. DisplayPort Test Implementation Group 4 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 1 MODIFICATION RECO
Agilent Technologies, Inc. DisplayPort Test Implementation Group 40 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Eye Height Calibra
Agilent Technologies, Inc. DisplayPort Test Implementation Group 41 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 38: HBR2 wit
Agilent Technologies, Inc. DisplayPort Test Implementation Group 42 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 C. Appendix: RBR IS
Agilent Technologies, Inc. DisplayPort Test Implementation Group 43 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Scope based ISI meas
Agilent Technologies, Inc. DisplayPort Test Implementation Group 44 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 41: RBR ISI
Agilent Technologies, Inc. DisplayPort Test Implementation Group 45 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 42: RBR eye
Agilent Technologies, Inc. DisplayPort Test Implementation Group 46 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 D. Appendix: Test A
Agilent Technologies, Inc. DisplayPort Test Implementation Group 47 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 The station configur
Agilent Technologies, Inc. DisplayPort Test Implementation Group 48 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 DUT Configuration Se
Agilent Technologies, Inc. DisplayPort Test Implementation Group 49 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Automated Sink Calib
Agilent Technologies, Inc. DisplayPort Test Implementation Group 5 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 2 INTRODUCTION Thes
Agilent Technologies, Inc. DisplayPort Test Implementation Group 50 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 50: Connecti
Agilent Technologies, Inc. DisplayPort Test Implementation Group 51 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Automated Sink Testi
Agilent Technologies, Inc. DisplayPort Test Implementation Group 52 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 53: Frequenc
Agilent Technologies, Inc. DisplayPort Test Implementation Group 53 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 E. Appendix: ISI Ad
Agilent Technologies, Inc. DisplayPort Test Implementation Group 54 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 56: Align Er
Agilent Technologies, Inc. DisplayPort Test Implementation Group 55 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 59: Activat
Agilent Technologies, Inc. DisplayPort Test Implementation Group 56 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 F. Appendix: Use of
Agilent Technologies, Inc. DisplayPort Test Implementation Group 6 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 3 REFERENCES The fo
Agilent Technologies, Inc. DisplayPort Test Implementation Group 7 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 5 TABLE OF FIGURES F
Agilent Technologies, Inc. DisplayPort Test Implementation Group 8 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Figure 50: Connection
Agilent Technologies, Inc. DisplayPort Test Implementation Group 9 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2 Sink Jitter Tolerance