Agilent Technologies
5975 Series
Mass Selective
Detectors
Hardware Installation
Page view 0
1 2 3 4 5 6 ... 59 60

Summary of Contents

Page 1 - Detectors

Agilent Technologies5975 Series Mass Selective DetectorsHardware Installation

Page 2 - Acknowledgements

10 Hardware Installation1 General InformationFigure 1 5975 Series MSD systemALS7890A GCCI gas flow moduleLocal control panel5975 Series MSDMSD power s

Page 3 - Contents

General Information 1Hardware Installation 11Important Safety WarningsBefore moving on, there are several important safety notices that you should alw

Page 4 - 4 Hardware Installation

12 Hardware Installation1 General InformationMany parts are dangerously hotMany parts of the MSD operate at temperatures high enough to cause serious

Page 5 - Hardware Installation 5

General Information 1Hardware Installation 13When using hydrogen gas, check the system for leaks to prevent possible fire and explosion hazards based

Page 6 - 6 Hardware Installation

14 Hardware Installation1 General InformationSafety and Regulatory CertificationsThe 5975 Series Mass Selective Detector conforms to the following saf

Page 7 - General Information

General Information 1Hardware Installation 15Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modific

Page 8 - Introduction

16 Hardware Installation1 General InformationElectromagnetic compatibilityThis device complies with the requirements of CISPR 11. Operation is subject

Page 9 - Hardware Installation 9

General Information 1Hardware Installation 17CleaningTo clean the unit, disconnect the power and wipe down with a damp, lint-free cloth.Recycling the

Page 10 - 1 General Information

18 Hardware Installation1 General InformationCustomer ResponsibilitiesThe following are not included in the standard installation. They are the respon

Page 11 - Important Safety Warnings

General Information 1Hardware Installation 19Before You StartVerify that all the conditions specified in the 5975 Series MSD Site Preparation Guide ha

Page 12 - Hydrogen

2 Hardware InstallationNotices© Agilent Technologies, Inc. 2010No part of this manual may be reproduced in any form or by any means (including elec-tr

Page 13

20 Hardware Installation1 General InformationOther DocumentationAdditional information is contained in the following documentation:• 5975 Series Mass

Page 14 - Information

215975 Series Mass Selective DetectorsHardware InstallationAgilent Technologies2InstallationOverview 22Unpacking and Inspecting the Materials 2

Page 15

22 Hardware Installation2 InstallationOverview Listed below are the steps performed for hardware installation. The actual order of the steps depends u

Page 16 - Electromagnetic compatibility

Installation 2Hardware Installation 23Unpacking and Inspecting the MaterialsShipping containers should not be opened until an Agilent Technologies rep

Page 17 - Cleaning

24 Hardware Installation2 InstallationInstalling the GCMaterials needed:Column cutter (5181-8836) Flat-blade screwdriver GC Series shipping kitPlease

Page 18 - Customer Responsibilities

Installation 2Hardware Installation 25Figure 3 Removing the knockoutsMetal knockoutsGC/MSD interface heater cable (GC end)

Page 19 - Before You Start

26 Hardware Installation2 InstallationConnecting the Helium Lines to the GC1 Connect the 6-foot copper tubing (or a custom length) to the carrier gas

Page 20 - Other Documentation

Installation 2Hardware Installation 27Purge the flow system for 20 minutes. You can continue with the next step while you wait.

Page 21 - Installation

28 Hardware Installation2 InstallationConfiguring the GC1 If there is an ALS, install it now. Follow the instructions provided with the ALS. 2 Install

Page 22 - Overview

Installation 2Hardware Installation 29Conditioning the Column1 Remove the septum covering the free end of the GC column and cut about 1 cm off the end

Page 23

Hardware Installation 3Contents1 General InformationIntroduction 8Important Safety Warnings 11Many internal parts of the MSD carry dangerous

Page 24 - Installing the GC

30 Hardware Installation2 InstallationPreparing the MSD for InstallationMaterials needed:5975 Series MSD Hardware Manual Tygon tubing or hose for pump

Page 25 - Hardware Installation 25

Installation 2Hardware Installation 31Preparing the Vacuum SystemMaterials needed:5975 Series MSD Hardware Manual Diagonal cutters Plastic gloves1 Rem

Page 26 - 2 Installation

32 Hardware Installation2 Installation6 Remove the blank flange from the free end of the hose, and connect the hose to the inlet port of the pump. 7 I

Page 27 - Hardware Installation 27

Installation 2Hardware Installation 339 Install the optional G3397A High vacuum gauge controller (required for CI MSDs). Be sure not to stress the cab

Page 28 - Configuring the GC

34 Hardware Installation2 InstallationPlumbing the Methane Line to the CI MSD Flow ModuleSpecial procedure for CI MSDRefer to the 5975 Series MSD Hard

Page 29 - Conditioning the Column

Installation 2Hardware Installation 357 Let the lines purge into a fume hood for 5 to 10 minutes.8 Connect the other end of the tubing to the Gas A fi

Page 30

36 Hardware Installation2 InstallationConnecting the MSD and the GC1 Verify that the GC power is turned off.2 Connect the two ends of the GC/MSD inter

Page 31 - Preparing the Vacuum System

Installation 2Hardware Installation 37When positioned properly, the end of the GC/MSD interface cover will touch the metal of the oven wall, and the e

Page 32

38 Hardware Installation2 InstallationPress on the side plate of the MSD to achieve a good seal. Verify that the foreline pump and the front fan turn

Page 33 - Hardware Installation 33

Installation 2Hardware Installation 39Installing the ChemStation PC 1 Unpack the PC, monitor, printer, and the modem (if present).2 Position the PC co

Page 34

4 Hardware InstallationConditioning the Column 29Preparing the MSD for Installation 30Preparing the Vacuum System 31Plumbing the Methan

Page 35 - Gas A (methane) line

40 Hardware Installation2 InstallationVerifying EI System PerformanceMaterial needed:1 pg/µL (0.001 ppm) OFN sample (5188-5348)Verify the tune perform

Page 36 - Connecting the MSD and the GC

Installation 2Hardware Installation 41Switching from EI to CI Operating ModeSpecial procedure for CI MSDRefer to the 5975 Series MSD Hardware Manual.1

Page 37 - (loosen completely)

42 Hardware Installation2 InstallationVerifying CI System PerformanceSpecial procedure for CI MSDRefer to the 5975 Series MSD Hardware Manual.Verify P

Page 38

Installation 2Hardware Installation 43Verifying High Mass System PerformanceRefer to the 5975 Series MSD Hardware Manual.To check high mass performanc

Page 39

44 Hardware Installation2 InstallationFamiliarizing the CustomerFollow the “Customer Familiarization Checklist” on page 53.Important points about the

Page 40 - Verify the tune performance

455975 Series Mass Selective DetectorsHardware InstallationAgilent Technologies3ChecklistsInstallation Checklist 46Customer Familiarization Checkl

Page 41 - Special procedure for CI MSD

46 Hardware Installation3 ChecklistsInstallation ChecklistVerifying that site preparation is complete Adequate space is available for the MSD system.

Page 42 - Verify NCI performance

Checklists 3Hardware Installation 47 Shipment damaged or incomplete:GC installation The following steps are general, and may differ slightly dependin

Page 43

48 Hardware Installation3 Checklists GC plumbedConfiguring the GC ALS installed (optional) Merlin Microseal installed (optional)  GC plugged in an

Page 44 - Familiarizing the Customer

Checklists 3Hardware Installation 49 PFDTD vial for CI cal valve filledPreparing the vacuum system Plugs removed from the foreline pump Foreline pu

Page 45 - Checklists

Hardware Installation 5MSD ChemStation software overview 54MSD ChemStation software autotune descriptions 54Methods 55Data acquisition

Page 46 - Installation Checklist

50 Hardware Installation3 Checklists Foreline pump and the front fan operation verified Foreline pump stops gurgling within 60 seconds verified MSD

Page 47 - GC installation

Checklists 3Hardware Installation 51 EI ion source removed CI ion source installed CI interface tip seal installed MSD pumped down and baked out f

Page 48 - MSD installation

52 Hardware Installation3 ChecklistsInstrument Log Book updated MSD serial number recorded on S/N report S/N Report(s) and Tune Report(s) faxed to G

Page 49 - Preparing the vacuum system

Checklists 3Hardware Installation 53Customer Familiarization ChecklistThe purpose of customer familiarization is to demonstrate the steps required to:

Page 50 - Performance verification

54 Hardware Installation3 ChecklistsHardware overview MSD CI MSD (if applicable) Gas Chromatograph ALS (if applicable) Bar code reader (if applic

Page 51

Checklists 3Hardware Installation 55 Manual tune explained View Tunes utility demonstrated and its importance explained PCI Autotune performed (if

Page 52 - Instrument Log Book updated

56 Hardware Installation3 Checklists Library search (searching a spectrum) demonstrated Difference between standalone and online data analysis expla

Page 53

Checklists 3Hardware Installation 57 Differences between the EI and CI source parts described Differences in installation and removal of the EI vers

Page 54 - System operation

58 Hardware Installation3 ChecklistsCompany name:Sales order number:Customer signature/date:Engineer signature/date:

Page 57 - Tune charts

Agilent Technologies© Agilent Technologies, Inc. Printed in USA, March 2010G3170-90002

Page 58 - 3 Checklists

75975 Series Mass Selective DetectorsHardware InstallationAgilent Technologies1General InformationIntroduction 8Important Safety Warnings 11Saf

Page 59

8 Hardware Installation1 General InformationIntroductionThis document provides installation instructions for the Agilent Technologies 5975 Series Mass

Page 60 - Agilent Technologies

General Information 1Hardware Installation 9Table 1 Available high vacuum pumpsModel name Product number Description Ionization modes5975C TAD VL M