Page view 23
24
General Programming Aspects
Status Model
QUEStionable Status Group
The QUEStionable Status group is used to report warning conditions
amongst the voltage, current, pulse timing and frequency parameters.
Warnings occur when a parameter, although not outside its maximum
limits, could be causing an invalid signal at the output because of the
actual settings and uncertainties of related parameters.
Bit QUEStionable
0 Voltage warning
1 Current warning
2 Time warning
3 Unused, always 0
4 Unused, always 0
5 Frequency warning
6 Unused, always 0
7 Unused, always 0
8 Unused, always 0
9 Unused, always 0
10 Unused, always 0
11 Unused, always 0
12 Unused, always
13 Unused, always 0
14 Unused, always 0
15 Always 0
Page view 23