Agilent 35670A Dynamic Signal AnalyzerTechnical SpecificationsKey SpecificationsFrequency Range 102.4 kHz 1 channel51.2 kHz 2 channel25.6 kHz 4 channe
2Summary of Features on Standard InstrumentThe following features are standard with the Agilent 35670A:Instrument ModesFFT Analysis Histogram/TimeCorr
3Agilent 35670A SpecificationsInstrument specifications apply after 15 minutes warm-up and within 2 hours of thelast self-calibration. When the intern
FFT Dynamic RangeSpurious Free Dynamic Range 90 dB typical (<-80 dBfs)(Includes Spurs, Harmonic Distortion, Intermodulation Distortion, Alias Prod
5InputInput Ranges(full scale)(Auto-Range Capability) +27 dBVrms (31.7 Vpk) to -51 dBVrms(3.99 mVpk) in 2 dB stepsMaximum Input Levels 42 VpkInput I
6Source OutputSource Types Sine, Random Noise, Chrip,Pink Noise, BurstRandom, Burst ChirpAmplitude Range AC: ±5V peak*DC: ±10V** Vacpk+ |Vdc| ≤ 10VAC
7Computed Order Tracking - Option 1D0Maximum Order x Maximum RPM ≤60Online (Real Time) 1 Channel Mode 25,600 Hz 2 Channel Mode 12,800 Hz4 Channel
General SpecificationsSafety Standards CSA Certified for Electronic Test and Measurement Equipment per CSA C22.2, NO. 231 This product is designed for
Comments to this Manuals