Agilent Technologies B2902 User Manual

Browse online or download User Manual for Measuring, testing & control Agilent Technologies B2902. Keysight Technologies B2900A Series Precision Source

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Precision Source/Measure Unit

Keysight TechnologiesB2900A SeriesPrecision Source/Measure UnitData Sheet

Page 2 - Introduction

Other supplemental characteristicsOutput characteristicsSensing Modes: 2-wire or 4-wire (Remote-sensing) connections

Page 3 - Cost-effective source/

Resistance measurementResistance measurement can be performed under either auto or manual test conditions. Auto resistance measurement is performed in

Page 4 - 

Environment For use in indoor facilitiesTemperature range Operating: 0 °C to +55 °CHumidity range Operat

Page 5 - 

Operation and Functions, continuedGraph viewGraph View displays measurement results on XY graphs (such as I-V and I-t/V-t curves) on up to 2 channels.

Page 6

Source/measurement capabilitiesSweep measurementNumber of steps: 1 to 2500Sweep mode: linear, logarithmic (log) or listSweep direction: single or doub

Page 7 - Pulse width

Program and interface capabilities, continuedLXILAN protocols and adher

Page 8 - 

Power cableUSB cableQuick Reference (English)Product Reference CD-ROM (including PDF manuals, Keysight B2900A Quick I/V Measurement Software and driv

Page 9 - Triggering

Keysight B1500A Semiconductor Device Analyzer www.keysight.com/find/B1500A Keysight B1505A Power Device Analyzer/ Curve Tracer (1500 A/10 kV)www.keysi

Page 10 - High capacitance mode

This information is subject to change without notice.© Keysight Technologies, 2013 - 2014Published in USA, August 3, 20145990-7009ENwww.keysight.com18

Page 11 - System speeds

The Keysight Technologies, Inc. B2900A Series of Precision Source/Measure Units are compact and cost-effective bench-top Source/Measure Units (SMUs) w

Page 12 - Operation and Functions

B2901A B2902A B2911A B2912ANumber of channels 1 2 1 2Max output Voltage 210 V 210 V 210 V 210 VCurrent DC 3.03 A 3.0

Page 13 - Easy-to-use key operations

1. Temperature: 23 °C ±5 °C2. Humidity: 30

Page 14

Range Programming resolution Accuracy (% reading + offset)Noise (peak to peak)0.1 Hz to 10 Hz1Max vo

Page 15 - Digital I/O interface

Source supplemental characteristics Max output power and

Page 16 - Furnished Accessories

Pulse source supplemental characteristics

Page 17 - Ordering Information

Range Measurement resolution Accuracy(% reading + offset)B2901A/B2902A B2911A/B2912A±200 mV

Page 18 - 5990-7009EN

Measurement supplemental characteristics Over range : 102