User Thresholds
1 Press Time .
2 Press the Next Menu softkey until the Define Thresholds softkey is
displayed on the far left side.
3 Press the Define Thresholds softkey.
4 Press the desired Thresholds softkey.
A softkey menu with six softkey choices appears. Five of them are related to
selecting thresholds.
10% 90% Rise time/fall time measurements performed at the 10% (lower)
and 90% (upper) levels. Frequency, period, duty cycle, positive pulse
width, and negative pulse width measurements will be performed at the
50% level.
20% 80% Rise time/fall time measurements performed at the 20% (lower)
and 80% (upper) levels. Frequency, period, duty cycle, positive pulse
width, and negative pulse width measurements will be performed at the
50% level.
Figure 2 5
Operating the Measurement/Storage Module
Setting Thresholds
2–16