Technical SpecificationsAgilent TechnologiesPNA Series Network AnalyzersN5230A/COptions 140/145/146/240/245/246(4-Port PNA-L)Manufacturing Part Number
6 NOTE: The following graphs also apply to the “C” model of the analyzer. Transmission Uncertainty (Specifications)
7 Reflection Uncertainty (Specifications)
8 Table 4. N4433A Electronic Calibration Module N5230A/C - Option 140 or 240 (Standard Test Set and Standard Power Range), Option 145 or 245 (Configu
9NOTE: The following graphs also apply to the “C” model of the analyzer. Transmission Uncertainty (Specifications)
10 Reflection Uncertainty (Specifications) This N5230A/C document does not present specifications for the 85052C or 85052D Calibration Kit. Ple
11Table 5. Uncorrected System Performance1 Description Specification Typical Options 140/145/146/240/245/246 Options 140/145/146/240/245/246 Dir
12 Table 6. Test Port Output1 Description Specification Typical Options 140/240 Opt 145/146/245/246 Options 140/240 Opt 145/146/245/246Frequenc
13 Table 6. Test Port Output (Continued) Description Specification Typical Options 140/240 Opt 145/146/245/246 Options 140/240 Opt 145/146/245/
14 Table 7. Test Port Input Description Specification Typical Options 140/145/146/240/245/246 Options 140/145/146/240/245/246 Test Port Noise Fl
15 Table 7. Test Port Input (Continued) Description Specification Typical Options 140/145/146/240/245/246 Options 140/145/146/240/245/246 Test P
Documentation WarrantyTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTU
16 Table 7. Test Port Input (Continued) Description Specification Typical Options 140/240 Options 145/245 Options 146/246 Options 140/240 Option
17 Table 7. Test Port Input (Continued) Description Specification Typical Options 140/240 Options 145/245 Options 146/246 Options 140/240 Options
18 Table 8. Dynamic Accuracy (Specificationa) Accuracy of the test port input power reading relative to the reference input power level. NOTE: The
19 Dynamic Accuracy, 0.500 GHz
20 Dynamic Accuracy, 1- 2 GHz
21 Dynamic Accuracy, 2 - 12.5 GHz
22 Dynamic Accuracy, 12.5 – 13.5 GHz
23 Dynamic Accuracy, 13.5 - 20 GHz a Dynamic accuracy is verified with the following measurements: • Compression over frequency • IF linea
24 Table 9. Test Port Input (Group Delay)a Description Supplemental Information (typ.) Aperture (selectable) (frequency span)/(number of points -1
25General Information Table 10. Miscellaneous Information Description Supplemental Information System IF Bandwidth Range 1 Hz to 600 kHz, nominal C
Contacting AgilentAssistance with test and measurement needs and information on finding a local Agilent office are available on the Web at:www.agilent
26 Table 12. Rear Panel Information Description Supplemental Information Trigger Inputs/Outputs BNC(f), TTL/CMOS compatible 10 MHz Reference In Con
27Note: Option H08 and Option H11 are not available with the N5230A/C network analyzer. Table 13. Analyzer Dimensions and Weight Description Supple
28 Measurement Throughput Summary Table 14 Typical Cycle Timea (ms) for Measurement Completion Description Typical Performance Options 140/145/24
29Table 15. Cycle Time vs. IF Bandwidth, Options 140/145/240/245 Applies to the Preset condition (201 points, correction off) except for the followin
30 Table 16. Cycle Time vs. Number of Points, Options 140/145/240/245 Applies to the Preset condition (correction off) except for the following chang
31Table 17. Data Transfer Time (ms) Description Typical Performance Number of Points 201 401 1601 16,001 SCPI over GPIB (Program executed on ex
32 Specifications: Front-Panel Jumpers ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Model N5230A/C Option 145/146/245/246 Table 18: Measurement Recei
33Table 20: Reference Output (Source Out) @ Max Specified Output Power Description Specification Typical Opt 145/245 Opt 146/246 Maximum Outp
34 Table 22: Coupler Inputs (Port 1 Cplr Thru, Port 2 Cplr Thru, Port 3 Cplr Thru, Port 4 Cplr Thru) Insertion Loss of Coupler Thru Description Spec
35Test Set Block Diagrams N5230A/C Option 140 or 240 Standard Test Set and Standard Power Range
This page intentionally left blank.
36 N5230A/C Option 145 or 245 Configurable Test Set and Extended Power Range Item Description Item Description Item Description a RCVR R IN h
37N5230A/C Option 146 or 246 Configurable Test Set and Extended Power Range
1 Technical Specifications for the N5230A/C Options 140/145/146/240/245/246 (4-Port PNA-L) (N5230-90020, Rev. 2008-11-11) DEFINITIONS ...
2 This page is intentionally blank.
3 This is a complete list of the N5230A/C Options 140, 145, 146, 240, 245, 246 network analyzer technical specifications. • To view or print the .pd
4 Corrected System Performance The specifications in this section apply for measurements made with the N5230A/C analyzer with the following condition
5N5230A/C Corrected System Performance with 3.5mm Connectors1 Option 140 or 145 or 146 or 240 or 245 or 246 1 From 300 kHz to 10 MHz, performance is