8
Table 4. N4433A Electronic Calibration Module
N5230A/C - Option 140 or 240 (Standard Test Set and Standard Power Range), Option 145 or 245
(Configurable Test Set and Extended Power Range), Option 146 or 246 (Configurable
Test Set and Extended Power Range with Internal Second Source)
Applies to the N5230A/C Option 140 or 145 or 146 or 240 or 245 or 246 analyzers, N4433A (3.5mm) electronic calibration module, 85131F
flexible test port cable set, and a full 4-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Description Specification (dB)
10 MHz to
500 MHz
500 MHz to
2 GHz
2 to
8 GHz
8 to
13.5 GHz
13.5 to
20 GHz
Directivity 52 52 47 45 45
Source Match 42 42 39 31 31
Load Match 45 45 41 35 35
Reflection Tracking
±0.060
+0.010/°C
±0.060
+0.010/°C
±0.090
+0.020/°C
±0.180
+0.030/°C
±0.180
+0.030/°C
Transmission Tracking
±0.045
+0.010/°C
±0.039
+0.010/°C
±0.055
+0.020/°C
±0.160
+0.030/°C
±0.160
+0.030/°C