Agilent Technologies N5230C User Manual Page 21

  • Download
  • Add to my manuals
  • Print
  • Page
    / 41
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 20
17
Table 7.
Test Port Input (Continued)
Description Specification Typical
Options
140/240
Options
145/245
Options
146/246
Options
140/240
Options
145/245
Options 146/246
Stability Phase
Stability is defined as a ratio measurement made at the test port.
300 KHz to 10 MHz -- -- -- +/-0.360°/°C
10 MHz to 45 MHz -- -- -- +/-0.020/°C
45 MHz to 500 MHz -- -- -- +/-0.030°/°C
500 MHz to 2 GHz -- -- -- +/-0.050°/°C +/-0.070°/°C
2 GHz to 4 GHz -- -- -- +/-0.100°/°C +/-0.150°/°C
4 GHz to 8 GHz -- -- -- +/-0.150°/°C +/-0.250°/°C
8 GHz to 13.5 GHz -- -- -- +/-0.300°/°C +/-0.500°/°C
13.5 GHz to 16 GHz -- -- -- +/-0.300°/°C +/-0.500°/°C
16 GHz to 20 GHz -- -- -- +/-0.350°/°C +/-0.650°/°C
Table 7.
Test Port Input (Continued)
Description Specification Typical
Options 140/240 Opt 145/146/245/246 Options 140/240 Opt 145/146/245/246
Reference Level Magnitude
Range +/-200 dB +/-200 dB -- --
Resolution 0.001dB 0.001dB -- --
Reference Level Phase
Range +/-500° +/-500° -- --
Resolution 0.01° 0.01° -- --
Table 7. Test Port Input (Continued)
Typical Performance
Damage Input Level Options 140/240 Opt 145/146/245/246
Test Port 1,2,3, and 4 +27 dBm or +/-16 VDC +27 dBm or +/-16 VDC
Receivers R, A, B, C, D -- +15 dBm or +/-16 VDC
Source out (reference) -- +27 dBm or +/-16 VDC
Source out (test ports) -- +27 dBm or +/-16 VDC
Coupler Thru -- +27 dBm or +/-16 VDC
Coupler Arm -- +15 dBm or +/-0 VDC
Page view 20