Table 5. Uncorrected System Performance
1
Description Specification Typical
Options 140/145/146/240/245/246 Options 140/145/146/240/245/246
Directivity
300 KHz to 10 MHz -- -23 dB
10 MHz to 1 GHz -28 dB --
1 GHz to 3 GHz -25 dB --
3 GHz to 5 GHz -20 dB --
5 GHz to 11.5 GHz -17 dB --
11.5 GHz to 13.5 GHz -15 dB --
13.5 GHz to 20 GHz -15 dB --
Source Match
300 KHz to 10 MHz -- -8 dB
10 MHz to 1 GHz -12 dB --
1 GHz to 3 GHz -12 dB --
3 GHz to 5 GHz -12 dB --
5 GHz to 10.5 GHz -12 dB --
10.5 GHz to 11.5 GHz -10 dB --
11.5 GHz to 13.5 GHz -8 dB --
13.5 GHz to 20 GHz -8 dB --
Load Match
300 KHz to 10 MHz -- -9 dB
10 MHz to 1 GHz -20 dB --
1 GHz to 3 GHz -20 dB --
3 GHz to 5 GHz -18 dB --
5 GHz to 11.5 GHz -12 dB --
11.5 GHz to 13.5 GHz -7 dB --
13.5 GHz to 16 GHz -7 dB --
16 GHz to 20 GHz -7.5 dB --
Crosstalk
2
300 KHz to 5 MHz -- -70 dB
5 MHz to 10 MHz -- -100 dB
10 MHz to 45 MHz -- -110 dB
45 MHz to 4 GHz -- -122 dB
4 GHz to 6 GHz -- -123 dB
6 GHz to 10.5 GHz -- -120 dB
10.5 GHz to 13.5 GHz -- -115 dB
13.5 GHz to 15 GHz -- -115 dB
15 GHz to 20 GHz -- -110 dB
1
Specifications apply over environmental temperature of 25 °C ±5 °C, with less than 1°C variation from the calibration temperature.
2
Measurement conditions: normalized to a thru, measured with two shorts, 10 Hz IF bandwidth, averaging factor of 8, alternate mode,
source power set to the lesser of the maximum power-out or the maximum receiver power.