Setting Up a Data Pattern Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 43
The Data page now looks as follows:
Reconnect the DUT Enable the outputs:
Generated Signals Use the pulse generator’s TRIGGER OUT to trigger the scope on
the data pattern.
To set up the trigger output:
1 On the Aux page, in the Trigger Output panel:
Select the trigger mode Data.
One trigger pulse occurs at the start of the repetitive data
pattern.
Set the Ampl to 2 V.
Set the Offset to 1 V.
Numeric
Edit Mode
Data pattern
length
Data pattern for
channel 2
Page view 42