Agilent Technologies 85032B/E 50 Specifications Page 130

  • Download
  • Add to my manuals
  • Print
  • Page
    / 269
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 129
Assembly Replacement 8712ET/ES and 8714ET/ES
A4 Source Assembly Replacement RF Network Analyzers
3-12 Service Guide
Installing the Source Assembly
CAUTION Be very careful when reconnecting the semirigid cables (items numbered 1
and 4 in Table 7-2 on page 7-4). They are not interchangeable. Ensure that
one cable is not confused with the other. It is possible to force the wrong cable
onto the wrong connector, resulting in damage to the connectors or the cables.
Tighten to 10 in-lb of torque.
NOTE Insure that SCREW 9 is re-installed and secure when you have completed
reassembling the analyzer
To install the source assembly into the analyzer, reverse the steps of the “Removing the
Source Assembly” procedure.
Page view 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 268 269

Comments to this Manuals

No comments