Agilent Technologies A.18.00 User Manual Page 291

  • Download
  • Add to my manuals
  • Print
  • Page
    / 581
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 290
291
Chapter 5, Advanced Operations
Status Reporting
Accessing the Communicate Status Register Group’s Registers
The following sections show the syntax and give programming examples (using
the Instrument BASIC programming language) for the STATus commands used to
access the Communicate Status Register Group’s registers.
Reading the Condition Register
Syntax
STATus:COMMunicate:CONDition?
Example
OUTPUT 714;"STAT:COMM:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Syntax
STATus:COMMunicate:PTRansition?
STATus:COMMunicate:NTRansition?
Example
OUTPUT 714;"STAT:COMM:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Syntax
STATus:COMMunicate:PTRansition <integer>
STATus:COMMunicate:NTRansition <integer>
Example
OUTPUT 714;"STAT:COMM:PTR 256"