Agilent Technologies E1346A Service Manual Page 24

  • Download
  • Add to my manuals
  • Print
  • Page
    / 66
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 23
Ch 90-92
LO Measurements 1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-4
Turn mainframe power ON
1. Measure Channel 90 LO Resistance
Send *RST to multiplexer
Send CLOS (@nn00) to close chan 90 and 00, where nn = card #
(omit leading zeroes in nn ).
Trigger the DMM with TRIG SGL and note reading
Send OPEN (@nn00) to open channel 90 and 00
Subtract measured value of R42 from DMM reading
Enter the result in Table 2-2 for Channel 90 LO
2. Repeat for Channels 91 and 92 LO
Repeat steps 1 and 2 for channels 91 and 92 LO
Use CLOS (@nncc) and OPEN (@nncc), where nn = card # and cc
= channel # (omit leading zeroes in nn ). Use 08 for channel 91 and
16 for channel 92.
Figure 2-4. Ch 90 LO Measurement Connections
2-6 Verification Tests
Page view 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 65 66

Comments to this Manuals

No comments