Agilent Technologies E6640A EXM User Manual Page 6

  • Download
  • Add to my manuals
  • Print
  • Page
    / 23
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 5
6
ProductSpecifications
Definitionsandconditions
Specification
Specificationsdescribetheperformanceparameterscoveredbytheproductwarrantyandare
validfrom20to35°Cunlessotherwisenoted.
Typical
Typicaldescribesadditionalproductperformanceinformationthatisnotcoveredbythe
productwarranty.Itisperformancebeyondspecificationsthat95percentoftheunitsexhibit
witha95percentconfidencelevel.Thisdata,showninitalics,doesnotincludemeasurement
uncertainty,andisvalidonlyatroomtemperature(approximately25°C)afteralignmentwithin
thestatedalignmenttimeandtemperaturelimits.
Nominal
Nominalvaluesindicateexpectedperformance,ordescribeproductperformancethatisuseful
intheapplicationoftheproduct,butarenotcoveredbytheproductwarranty.
Thetestsetwillmeetitsspecificationwhen:
• Thetestsetiswithinitscalibrationcycle
• Thetestsethasbeenstoredatanambienttemperaturewithintheallowedoperating
rangeforatleasttwohoursbeforebeingturnedon;ifithaspreviouslybeenstoredata
temperaturerangeinsidetheallowedstoragerange,butoutsidetheallowedoperating
range
• Thetestsethasbeenturnedonforatleast45minutes
• TheRF,IF,andSourceAlignments
1
havebeenrunwithintheprevious7days
• AnALLAlignment
1
hasbeenrun:
• Withintheprevious8hours
• Ifthetemperaturehaschangedmorethan5°Cfromtheprevious“ALL”alignment
1. For more information on using Alignments in a manufacturing environment, please see the EXM user documentation.
Page view 5
1 2 3 4 5 6 7 8 9 10 11 ... 22 23

Comments to this Manuals

No comments