Agilent Technologies InfiniiVision 5000 Series User's Guide Page 211

  • Download
  • Add to my manuals
  • Print
  • Page
    / 428
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 210
Measurements and Math Functions 6
InfiniiVision Oscilloscopes User’s Guide 211
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a percentage of
Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
reference point).
A
UTOMATIC OVERSHOOT MEASUREMENT
Rising edge overshoot =
local Maximum - D Top
Amplitude
----------------------------------------------------------
X 100
Falling edge overshoot =
Base - D local Minimum
Amplitude
-----------------------------------------------------------
X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base
Page view 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 427 428

Comments to this Manuals

No comments