Serial Decode/Lister 10
InfiniiVision Oscilloscopes User’s Guide 329
UART/RS232 Serial Decode
1 Select the UART/RS232 serial decode mode:
a Press the [Acquire] key.
b Press the Serial Decode softkey.
c In the Serial Decode Menu, press the Mode softkey.
d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the
UART/RS232 serial decode mode if it is not already selected.
NOTE
Because of oscilloscope hardware resource limitations, you cannot decode UART/RS232
data while CAN, LIN, or FlexRay triggering is selected.
NOTE
For UART/RS232 triggering setup see page 166.
Display
decode
Decode
mode
Signal
setup menu
Return to
previous menu
Frame
counter
reset
Display
decode in
Lister