78 L4400 User’s Guide
3 Operating and Programming
Figure 3-12. Channel Delay.
You can set the channel delay to any value between 0 seconds and
60
seconds, with 1 ms resolution. You can select a different delay for
each channel.
You can select a unique delay for every channel on the module.
The channel delay is valid only while scanning. If no channels have
been assigned to the scan list, the specified channel delay is ignored
(no error is generated).
The default channel delay is 0.0 seconds.
A Factory Reset (*RST command) sets the channel delay to 0.0s.
The following command adds a 2- second channel delay to the specified
channels.
ROUT:CHAN:DELAY 2,(@1003,1013)