116 Programmer’s Guide
5 Command Reference
Set dBc method of OBW
[:SENSe]:OBWidth:XDB <value>
[:SENSe]:OBWidth:XDB?
specify the power level used to determine the emission bandwidth as the
number of dB down from the highest signal point, within the occupied
bandwidth span.
*RST: 26
Range: 0.10 to 100
Example: :OBWidth:XDB 4
Key access: Meas > Occupied BW > dBc
OBW Span
[:SENSe]:OBWidth:FREQuency:SPAN <freq>
[:SENSe]:OBWidth:FREQuency:SPAN?
Use this command to set the span in OBW test.
Example: :OBWidth:FREQuency:SPAN 50 MHz
Key access: Span > Span
OBW Max Hold State
[:SENSe]:OBWidth:MAXHold OFF|ON|0|1
[:SENSe]:OBWidth:MAXHold?
Use this command to turn on/off trace max hold state in OBW test.
Example: :OBWidth:MAXHold 1
Key access: Trace > Max Hold
Query OBW and Centroid Result
:MEASure:OBW?
:FETCh:OBW?
Use this command to query the occupied bandwidth and bandwidth
centroid according to the method you set.