174 Programmer’s Guide
5 Command Reference
AM X Scale Coupling State
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:COUPle?
Use this command to turn on/off scale coupling state.
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:COUPle ON
Key access: Shift > Disp >X Scale > Scale Coupling
AM Y Reference Value
:DISPlay:AM:VIEW:WINDow:TRACe:Y[:SCALe]:RVALue <
value>
:DISPlay:AM:VIEW:WINDow:TRACe:Y[:SCALe]:RVALue?
Use this command to set reference value of Y axis. The unit of this
parameter is %.
Range: –150% to 150%
Example: :DISPlay:AM:VIEW:WINDow:TRACe:Y:RVALue 20
Key access: Shift > Disp >Y Scale > Ref Value
AM Y Scale/DIV
:DISPlay:AM:VIEW:WINDow:TRACe:Y[:SCALe]:PDIVi-
sion < double >
:DISPlay:AM:VIEW:WINDow:TRACe:Y[:SCALe]:PDIVi-
sion?
Use this command to set the Scale per division of Y axis. The unit of
this parameter is %.
Range: 0.1% to 100%
Example: :DISPlay:AM:VIEW:WINDow:TRACe:Y:PDIVision 50
Key access: Shift > Disp >Y Scale > Scale/DIV