Agilent TechnologiesAgilent 8494/95/96A/B AttenuatorsOperating and Service Manual
10 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 1 IntroductionProduct OverviewAgilent 8494A/B, 8495A/B, and 8496A/B are 50- ohm co
Introduction 1Agilent 8494/95/96A/B Attenuators Operating and Service Manual 11Instrument OptionsEach instrument is specified with an option number wh
12 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 1 IntroductionSpecificationsFrequency Range and AttenuationAttenuation Accuracy(±d
Introduction 1Agilent 8494/95/96A/B Attenuators Operating and Service Manual 13Maximum SWRMaximum Residual AttenuationAttenuation Repeatability±0.03 d
14 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 1 IntroductionRF Power Handling Capability1 W average, 100 W peak with maximum pul
15Agilent 8494/95/96A/B AttenuatorsOperating and Service ManualAgilent Technologies2Environmental Specifications & Physical DimensionsEnvironmenta
16 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 2 Environmental Specifications & Physical DimensionsEnvironmental Specification
Environmental Specifications & Physical Dimensions 2Agilent 8494/95/96A/B Attenuators Operating and Service Manual 17Physical DimensionsTable 7 sh
18 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 2 Environmental Specifications & Physical DimensionsFigure 1 Dimensions of Agil
19Agilent 8494/95/96A/B AttenuatorsOperating and Service ManualAgilent Technologies3Operating GuidesInstallation 20Initial Inspection 20Mating Connect
2 Agilent 8494/95/96A/B Attenuators Operating and Service ManualNotices© Agilent Technologies, Inc. 2011No part of this manual may be reproduced in an
20 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 3Operating GuidesInstallationInitial Inspection1 Inspect the shipping container for
Operating Guides 3Agilent 8494/95/96A/B Attenuators Operating and Service Manual 21Mating ConnectorsMating RF connectors used with the Option 001 must
22 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 3Operating GuidesOperating InstructionsAfter the instrument is connected, the attenu
Operating Guides 3Agilent 8494/95/96A/B Attenuators Operating and Service Manual 23Operator’s CheckThe operator's check allows the operator to ma
24 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 3Operating GuidesProcedure1 Connect equipment as shown in Figure 2 on page 23 with t
Operating Guides 3Agilent 8494/95/96A/B Attenuators Operating and Service Manual 25Performance TestsThe Agilent 8494/95/96A/B Attenuators can be teste
26 Agilent 8494/95/96A/B Attenuators Operating and Service Manual 3Operating GuidesThis page is intentionally left blank.
Agilent 8494/95/96A/B Attenuators Operating and Service Manual 3CertificationAgilent Technologies certifies that this product met its published specif
4 Agilent 8494/95/96A/B Attenuators Operating and Service ManualRegulatory MarkingsThe CE mark is a registered trademark of the European Community. I
Agilent 8494/95/96A/B Attenuators Operating and Service Manual 5Contacting Agilent For more information, please contact your nearest Agilent office.Or
6 Agilent 8494/95/96A/B Attenuators Operating and Service ManualThis page is intentionally left blank.
Agilent 8494/95/96A/B Attenuators Operating and Service Manual 7Contents1 Introduction 9Product Overview 10Instrument Options 11Specifications 12Freq
8 Agilent 8494/95/96A/B Attenuators Operating and Service ManualThis page is intentionally left blank.
9Agilent 8494/95/96A/B AttenuatorsOperating and Service ManualAgilent Technologies1IntroductionProduct Overview 10Instrument Options 11Specifications